Peri-Implant Distribution of Polyethylene Debris in Postmortem-Retrieved Knee Arthroplasties: Can Polyethylene Debris Explain Loss of Cement-Bone Interlock in Successful Total Knee Arthroplasties?

被引:6
作者
Cyndari, Karen I. [1 ]
Goodheart, Jacklyn R. [1 ]
Miller, Mark A. [1 ]
Oest, Megan E. [1 ]
Damron, Timothy A. [1 ]
Mann, Kenneth A. [1 ]
机构
[1] SUNY Upstate Med Univ, Dept Orthopaed Surg, 3216 IHP,750 East Adams St, Syracuse, NY 13210 USA
基金
美国国家卫生研究院;
关键词
ultra-high-molecular-weight polyethylene; polyethylene debris; total knee arthroplasty; aseptic loosening; postmortem retrieval; TOTAL JOINT REPLACEMENTS; TOTAL HIP-ARTHROPLASTY; UHMWPE WEAR DEBRIS; IN-VIVO SERVICE; FEMORAL COMPONENTS; FLUID PRESSURE; UNITED-STATES; OSTEOLYSIS; PARTICLES; INTERFACE;
D O I
10.1016/j.arth.2017.01.047
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Loss of mechanical interlock between cement and bone with in vivo service has been recently quantified for functioning, nonrevised, cemented total knee arthroplasties (TKAs). The cause of interlocking trabecular resorption is not known. The goal of this study is to quantify the distribution of PE debris at the cement-bone interface and determine if polyethylene (PE) debris is locally associated with loss of interlock. Methods: Fresh, nonrevised, postmortem-retrieved TKAs (n = 8) were obtained en bloc. Laboratory-prepared constructs (n = 2) served as negative controls. The intact cement-bone interface of each proximal tibia was embedded in Spurr's resin, sectioned, and imaged under polarized light to identify birefringent PE particles. PE wear particle number density was quantified at the cement-bone interface and distal to the interface, and then compared with local loss of cement-bone interlock. Results: The average PE particle number density for postmortem-retrieved TKAs ranged from 8.6 (1.3) to 24.9 (3.1) particles/mm(2) (standard error) but was weakly correlated with years in service. The average particle number density was twice as high as distal (>5mm) to the interface compared to at the interface. The local loss of interlock at the interface was not related to the presence, absence, or particle density of PE. Conclusion: PE debris can migrate extensively along the cement-bone interface of well-fixed tibial components. However, the amount of local bone loss at the cement-bone interface was not correlated with the amount of PE debris at the interface, suggesting that the observed loss of trabecular interlock in these well-fixed TKAs may be due to alternative factors. (C) 2017 Elsevier Inc. All rights reserved.
引用
收藏
页码:2289 / 2300
页数:12
相关论文
共 51 条
[1]   LOCALIZED ENDOSTEAL BONE LYSIS IN RELATION TO THE FEMORAL COMPONENTS OF CEMENTED TOTAL HIP ARTHROPLASTIES [J].
ANTHONY, PP ;
GIE, GA ;
HOWIE, CR ;
LING, RSM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (06) :971-979
[2]   A randomised controlled trial of cemented versus cementless press-fit condylar total knee replacement [J].
Baker, P. N. ;
Khaw, F. M. ;
Kirk, L. M. G. ;
Esler, C. N. A. ;
Gregg, P. J. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (12) :1608-1614
[3]   REGIONAL DISSEMINATION OF WEAR DEBRIS FROM A TOTAL KNEE PROSTHESIS - A CASE-REPORT [J].
BAUER, TW ;
SALTARELLI, M ;
MCMAHON, JT ;
WILDE, AH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1993, 75A (01) :106-111
[4]   Characteristics of highly cross-linked polyethylene wear debris in vivo [J].
Baxter, Ryan M. ;
MacDonald, Daniel W. ;
Kurtz, Steven M. ;
Steinbeck, Marla J. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2013, 101B (03) :467-475
[5]   Distinct immunohistomorphologic changes in periprosthetic hip tissues from historical and highly crosslinked UHMWPE implant retrievals [J].
Baxter, Ryan M. ;
Ianuzzi, Allyson ;
Freeman, Theresa A. ;
Kurtz, Steven M. ;
Steinbeck, Marla J. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2010, 95A (01) :68-78
[6]   Cementing techniques for the tibial component in primary total knee replacement [J].
Cawley, D. T. ;
Kelly, N. ;
McGarry, J. P. ;
Shannon, F. J. .
BONE & JOINT JOURNAL, 2013, 95B (03) :295-300
[7]  
COLLIER JP, 1992, CLIN ORTHOP RELAT R, P97
[8]   Influence of time in-situ and implant type on fixation strength of cemented tibial trays - A post mortem retrieval analysis [J].
de Uhlenbrock, Anne Gebert ;
Pueschel, Vesna ;
Pueschel, Klaus ;
Morlock, Michael M. ;
Bishop, Nicholas E. .
CLINICAL BIOMECHANICS, 2012, 27 (09) :929-935
[9]   A literature review of the association between wear rate and osteolysis in total hip arthroplasty [J].
Dumbleton, JH ;
Manley, MT ;
Edidin, AA .
JOURNAL OF ARTHROPLASTY, 2002, 17 (05) :649-661
[10]   Microplastic pollution in the surface waters of the Laurentian Great Lakes [J].
Eriksen, Marcus ;
Mason, Sherri ;
Wilson, Stiv ;
Box, Carolyn ;
Zellers, Ann ;
Edwards, William ;
Farley, Hannah ;
Amato, Stephen .
MARINE POLLUTION BULLETIN, 2013, 77 (1-2) :177-182