Sensitivity and Spatial Resolution of Transvenous Leads in Implantable Cardioverter Defibrillator

被引:12
作者
Requena-Carrion, Jesus [1 ]
Vaisanen, Juho [2 ]
Alonso-Atienza, Felipe [1 ]
Garcia-Alberola, Arcadi [3 ]
Javier Ramos-Lopez, Francisco [1 ]
Luis Rojo-Alvarez, Jose [1 ]
机构
[1] Rey Juan Carlos Univ, Dept Signal Theory & Commun, Madrid 28943, Spain
[2] Tampere Univ Technol, Dept Biomed Engn, FIN-33101 Tampere, Finland
[3] Univ Murcia, Arrhythmia Unit, Hosp Univ Virgen de la Arrixaca, Murcia 30120, Spain
关键词
Bipolar lead; implantable cardioverter defibrillator (ICD); lead field; resolution volume; sensitivity; spatial resolution; transvenous leads; unipolar lead; FINITE-DIFFERENCE METHOD; VENTRICULAR-FIBRILLATION; BIPOLAR; MODEL; ELECTROGRAMS; EEG; DISTRIBUTIONS; RECIPROCITY; POTENTIALS; SYSTEMS;
D O I
10.1109/TBME.2009.2027425
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
It has been previously documented that the main features and sensing performance of electrograms (EGMs) recorded in implantable cardioverter defibrillators (ICDs) depend on lead configuration. Although this dependence has been ascribed to differences in lead sensitivity and spatial resolution, the quantification of these two properties on ICD has not yet been attempted. In this paper, an operative framework to study the spatial resolution of ICD transvenous leads is presented. We propose to quantify the spatial resolution of ICD transvenous leads based on a new characterization called lead resolution volume (ResV). We analyzed the sensitivity distribution and the ResV of two unipolar (tip-can and coil-can) and two bipolar (true or tip-ring and integrated or tip-coil) ICD transvenous lead configurations. A detailed 3-D model of the human thorax based on the visible human man dataset was used to compute the lead sensitivity and computer simulations of simple cardiac dynamics were used to quantify the ResV. Differences in the sensitivity distribution throughout the ventricular myocardium (VM) were observed for each lead configuration. In our computer model of the human thorax, the ResV was found to comprise 7%, 35%, 45%, and 70% of the VM for true bipolar, integrated bipolar, tip-can unipolar, and coil-can unipolar ICD leads, respectively. Furthermore, our analysis shows that the spatial resolution depends on both lead sensitivity and cardiac dynamics, and therefore, it can vary for different heart rhythms.
引用
收藏
页码:2773 / 2781
页数:9
相关论文
共 40 条
[1]  
Ackerman M J, 1991, J Biocommun, V18, P14
[2]  
Arzbaecher R., 1979, IEEE 1979 Frontiers of Engineering in Health Care Conference, P314
[3]   Intrathoracic far-field electrocardiogram allows continuous monitoring of ischemia after total coronary occlusion [J].
Asbach, Stefan ;
Weiss, Ingo ;
Wenzel, Beate ;
Bode, Christoph ;
Zehender, Manfred .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (12) :1334-1340
[4]  
Atienza FA, 2005, REV ESP CARDIOL, V58, P41
[5]  
BELOTT PH, 2007, CLIN CARDIAC PACING, P561
[6]   The diagnostic evolution of the cardiac implantable electronic device: The implantable monitor of ischemia [J].
Bunch, T. Jared ;
Day, John D. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2008, 19 (03) :282-284
[7]   Comparison of bipolar and integrated sensing for redetection of ventricular fibrillation [J].
Cooklin, M ;
Tummala, RV ;
Peters, RW ;
Shorofsky, SR ;
Gold, MR .
AMERICAN HEART JOURNAL, 1999, 138 (01) :133-136
[8]   COMPARISON OF UNIPOLAR AND BIPOLAR ELECTROGRAMS FOR CARDIAC-PACEMAKER SENSING [J].
DECAPRIO, V ;
HURZELER, P ;
FURMAN, S .
CIRCULATION, 1977, 56 (05) :750-755
[9]   ASSESSING THE EFFECT OF UNCERTAINTY IN INTRACAVITARY ELECTRODE POSITION ON ENDOCARDIAL POTENTIAL ESTIMATES [J].
DERFUS, DL ;
PILKINGTON, TC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (07) :676-681
[10]   True bipolar defibrillator leads have increased sensing latency and threshold compared with the integrated bipolar configuration [J].
Frain, Bryan H. ;
Ellison, Kristin E. ;
Michaud, Gregory F. ;
Koo, Charles H. ;
Buxton, Alfred E. ;
Kirk, Malcolm M. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2007, 18 (02) :192-195