Evaluation of scattered radiation from fluoroscopy using small OSL dosimeters

被引:15
作者
Ito, Hajime [1 ]
Kobayashi, Ikuo [1 ,2 ]
Watanabe, Kazutoshi [1 ]
Ochi, Shigehiro [1 ]
Yanagawa, Noriyuki [1 ,3 ]
机构
[1] Eastern Chiba Med Ctr, Dept Radiol, 3-6-2 Okayamadai, Togane, Chiba 2838686, Japan
[2] Nagase Landauer Ltd, C22-1 Shimanasuwa, Tsukuba, Ibaraki 3002655, Japan
[3] Tsukuba Int Univ, Fac Hlth Sci, Dept Radiol Technol, 6-20-1 Manabe, Tsuchiura, Ibaraki 3000051, Japan
关键词
Fluoroscopy; Endoscopic retrograde cholangiopancreatography; Scattered radiation; Dosimetry; Radiation protection; Small OSL dosimeter;
D O I
10.1007/s12194-019-00536-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Recently, there has been a significant amount of interest in studying the importance of radiation doses to the eye lens during endoscopic retrograde cholangiopancreatography (ERCP). A study that focused on measuring the scattered radiation using an ionization chamber survey meter reported that a lead curtain was useful in reducing the scattered radiation. The over-couch X-ray tube system tends to deliver higher doses to the head and neck of the staff involved in the procedure than the under-couch X-ray tube position. In this study, a small optically stimulated luminescence (OSL) dosimeter called the nanoDot was used to evaluate and measure the amount of radiation; this dosimeter was developed by Landauer Ltd. and was specifically designed for point measurements. There are numerous studies that have reported the usefulness of personal OSL dosimeters other than the nanoDot to measure scattered radiation. Here, we evaluated the amount of scattered radiation, along with the degree of reduction achieved with the use of a protective curtain, while employing a personal OSL dosimeter and nanoDot. When the scattered radiation dose was measured using the nanoDot, the maximum recorded value without a protective curtain was 0.363 mGy and that with a protective curtain was 0.026 mGy, both at the height of 100 cm. The maximum reduction rate of scattered radiation while using a protective curtain was approximately 93% and 97% at 100 cm and 150 cm, respectively. The measured values recorded using both personal OSL dosimeters and nanoDot machine were strongly correlated.
引用
收藏
页码:393 / 400
页数:8
相关论文
共 12 条
[1]  
[Anonymous], 2009, MICROSTAR US MAN 1 6, P1
[2]  
Chida K, 2008, JPN SOC RADIOL TECHN, V50, P1009, DOI [10.6009/jjrt.64.1009, DOI 10.6009/JJRT.64.1009]
[3]  
Hayashi H, 2014, MED IMAGING INF SCI, V31, P228
[4]   Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements [J].
Jursinic, Paul A. .
MEDICAL PHYSICS, 2007, 34 (12) :4594-4604
[5]   Angular dependence of the nanoDot OSL dosimeter [J].
Kerns, James R. ;
Kry, Stephen F. ;
Sahoo, Narayan ;
Followill, David S. ;
Ibbott, Geoffrey S. .
MEDICAL PHYSICS, 2011, 38 (07) :3955-3962
[6]  
Kusama T, 1996, IRYO KAGAKUSYA, V1, P98
[7]   A head phantom study for intraocular dose evaluation of 64-slice multidetector CT examination in patients with suspected cranial trauma [J].
Matsubara, Kosuke ;
Koshida, Kichiro ;
Noto, Kimiya ;
Takata, Tadanori ;
Suzuki, Masayuki ;
Shimono, Tetsunori ;
Yamamoto, Tomoyuki ;
Matsui, Osamu .
EUROPEAN JOURNAL OF RADIOLOGY, 2011, 79 (02) :283-287
[8]   Occupational radiation dose to eyes from endoscopic retrograde cholangiopancreatography procedures in light of the revised eye lens dose limit from the International Commission on Radiological Protection [J].
O'Connor, U. ;
Gallagher, A. ;
Malone, L. ;
O'Reilly, G. .
BRITISH JOURNAL OF RADIOLOGY, 2013, 86 (1022)
[9]  
Stewart F. A., 2012, Annals of the ICRP, V41, P37, DOI 10.1016/j.icrp.2012.02.001
[10]  
Take Toshio, 2007, Nihon Hoshasen Gijutsu Gakkai Zasshi, V63, P1246