Field-testing and validation in a primary care setting of a point-of-care test for C-reactive protein

被引:28
作者
Seamark, DA
Backhouse, SN
Powell, R
机构
[1] Honiton Res Practice, Honiton EX14 2NY, England
[2] Sch Postgrad Med Hlth Sci, Exeter & N Devon NHS Res & Dev Support Unit, Exeter, Devon, England
关键词
D O I
10.1258/000456303763046139
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background Use of point-of-care testing (POCT) in primary care has increased. There is a need for high-quality field evaluation of POCT before deployment can be considered. Method A POCT system for C-reactive protein was evaluated in a routine general practice setting. The standard laboratory method was a dry slide method based in a routine hospital laboratory. Results Scatterplots for both venous and capillary blood POCT system results versus the standard laboratory assay produced correlation coefficients of greater than 0.96. Bland-Altman plots indicated that 95% of venous and capillary POCT results fell within +/- 10 mg/L of the mean value with no clinically significant difference from laboratory results. Conclusions The POCT system performed reliably in a routine general practice setting with satisfactory performance against an accepted laboratory method.
引用
收藏
页码:178 / 180
页数:3
相关论文
共 6 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Systematic review of near patient test evaluations in primary care [J].
Delaney, BC ;
Hyde, CJ ;
McManus, RJ ;
Wilson, S ;
Fitzmaurice, DA ;
Jowett, S ;
Tobias, R ;
Thorpe, GH ;
Hobbs, FDR .
BMJ-BRITISH MEDICAL JOURNAL, 1999, 319 (7213) :824-827
[3]  
HOBBS FD, 1997, HLTH TECHNOL ASSESS, V1
[4]  
Hobbs FDR, 1996, BRIT J GEN PRACT, V46, P395
[5]  
Husebekk A, 1999, C REACTIVE PROTEIN C
[6]  
Lindbaek Morten, 1998, Tidsskrift for den Norske Laegeforening, V118, P1176