Expired EBT3 Films' Sensitivity for the Measurement of X-ray and UV Radiation: An Optical Analysis

被引:2
|
作者
Abu Mhanna, Hamad Yahia [1 ]
Omar, Ahmad Fairuz [1 ]
Radzi, Yasmin Md [1 ]
Akhdar, Hanan Fawaz [2 ]
Al Ewaidat, Haytham [3 ]
机构
[1] Univ Sains Malaysia, Sch Phys, Gelugor 11800, Penang, Malaysia
[2] Imam Mohammad Ibn Saud Islamic Univ IMSIU, Phys Dept, Coll Sci, Riyadh 13318, Saudi Arabia
[3] Jordan Univ Sci & Technol, Fac Appl Med Sci, Dept Allied Med Sci Radiol Technol, Irbid 22110, Jordan
关键词
colorimetry; EBT3; expiry date; visible spectroscopy; RADIOCHROMIC FILM; GAFCHROMIC EBT3; DOSIMETRY; CALIBRATION; PROTOCOL;
D O I
10.3390/ma15082903
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The aim of this study is to compare the optical responses of external beam therapy 3 (EBT3) films exposed to X-rays and solar ultraviolet rays (SUV-rays), as a dose control technique in the clinical sector for various radiation types, energies, and absorbed doses up to 4 Gy. In this study, EBT3 films with three different expiry dates were prepared and cut into pieces of size 2 by 2 cm(2). The first group was exposed to 90 kVp X-rays, while the second group was exposed to the SUV-rays at noon. The analysis was performed using a visible Jaz spectrometer and an EPSON Perfection V370 Photo scanner to obtain the absorbance, the net reflective optical density (ROD) and the red-green-blue (RGB) values of the samples. The results have shown that spectroscopic measurements of the exposed expired EBT3 films with these radiation sources are able to produce primary peaks and secondary peaks at lambda = 641.74 nm and lambda = 585.98 nm for X-rays, and at lambda = 637.93 nm and lambda = 584.45 nm for SUV-rays, respectively. According to these findings, compared to 2021 films that expired shortly before the trial start date; 2018 films responded better to the absorbed dose than 2016 films when exposed to both X-ray and SUV-rays. In terms of energy dependence, the expired EBT3 2018 had the largest net ROD value. Using L*a*b* indices extracted from the RGB data, and despite that EBT3 films have expiry dates according to the manufacturer; all the films exhibited a substantial colour change, indicating that these films are still usable for clinical and research purposes.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Optical Response of Expired EBT3 Film for Absorbed Dose Measurement in X-ray and Electron Beam Range
    Musa, Maisarah
    Abu Mhanna, Hamad Yahia
    Omar, Ahmad Fairuz
    Radzi, Yasmin Md
    Akhdar, Hanan Fawaz
    Ishak, Nor Hafizah
    Al Ewaidat, Haytham
    APPLIED SCIENCES-BASEL, 2022, 12 (08):
  • [2] EBT3 Films in Low Solar Ultraviolet and X-Ray Dose Measurement: A Comparative Analysis
    Alsaee, Saleh K.
    Omar, Ahmad Fairuz
    Ahmed, Naser M.
    Alsadig, Ahmed
    Sulieman, A.
    Alzimami, Khalid
    DOSE-RESPONSE, 2019, 17 (02):
  • [3] Spectral analysis of the EBT3 radiochromic film irradiated with 6 MV X-ray radiation
    Leon-Marroquin, Elsa Y.
    Camacho-Lopez, Miguel A.
    Garcia-Garduno, Olivia A.
    Alfredo Herrera-Gonzalez, J.
    Eduardo Villarreal-Barajas, J.
    Gutierrez-Fuentes, Ruben
    Contreras-Bulnes, Rosalia
    RADIATION MEASUREMENTS, 2016, 89 : 82 - 88
  • [4] Responses of Gafchromic EBT3 films with polypropylene barriers to UV radiation
    Yu, K. N.
    Chun, S. L.
    Chan, P. M.
    RESULTS IN PHYSICS, 2017, 7 : 1976 - 1977
  • [5] Changes of the optical characteristics of radiochromic films in the transition from EBT3 to EBT-XD films
    Schoenfeld, Andreas A.
    Wieker, Soeren
    Harder, Dietrich
    Poppe, Bjoern
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (14) : 5426 - 5442
  • [6] Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams
    Brown, Thomas A. D.
    Hogstrom, Kenneth R.
    Alvarez, Diane
    Matthews, Kenneth L., II
    Ham, Kyungmin
    Dugas, Joseph P.
    MEDICAL PHYSICS, 2012, 39 (12) : 7412 - 7417
  • [7] Characterization of Gafchromic EBT3 films for ultraviolet radiation dosimetry
    Aydarous, Abdulkadir
    Al-Omary, Enaam Ali
    El Ghazaly, Mahmoud
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2014, 169 (03): : 249 - 255
  • [8] Comparison of three film analysis softwares using EBT2 and EBT3 films in radiotherapy
    Pocza, Tamas
    Zongor, Zsuzsanna
    Melles-Bencsik, Barbara
    Zita Tatai-Szabo, Dora
    Major, Tibor
    Pesznyak, Csilla
    RADIOLOGY AND ONCOLOGY, 2020, 54 (04) : 505 - 512
  • [9] Comparing three UV wavelengths for pre-exposing Gafchromic EBT2 and EBT3 films
    Katsuda, Toshizo
    Gotanda, Rumi
    Gotanda, Tatsuhiro
    Akagawa, Takuya
    Tanki, Nobuyoshi
    Kuwano, Tadao
    Yabunaka, Kouichi
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (06): : 449 - 457
  • [10] A study on the use of Gafchromic™ EBT3 film for output factor measurements in kilovoltage X-ray beams
    Gill, Simran
    Hill, Robin
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2013, 36 (04) : 465 - 471