Model eye tool for retinal optical coherence tomography instrument calibration

被引:4
作者
Wang, Hongting [1 ]
Liu, Wenli [1 ]
Hu, Zhixiong [1 ]
Li, Xiuyu [1 ]
Li, Fei [1 ]
Duan, Liangcheng [1 ]
机构
[1] Natl Inst Metrol, Ctr Med Metrol, Beijing 100029, Peoples R China
基金
国家重点研发计划;
关键词
Optical coherence tomography; model eye; microsphere-embedded phantom; RESOLUTION; PHANTOM;
D O I
10.1142/S1793545821500103
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Optical coherence tomography (OCT) has been extensively used as noninvasive tool for biological tissues owing to its three-dimensional imaging ability and high axial resolution. OCT quality assurance is vital in these occasions to keep the reliability and accuracy in medical diagnosis. It is necessary to develop a calibration tool for OCT product manufacture, calibration, and quality control. A practical tool is demanded in the OCT quality control and calibration of OCT. So far, there is no such a practical tool that can test all the key parameters of OCT. We design and fabricate a model eye tool, which has this function. The model eye comprises a doublet lens, a single filament, a piece of glass plate and the microsphere-embedded phantom. The doublets lens is bonded by two pieces of planoconvex lenses in the plane position. The first lens focuses parallel light onto the rear surface of the second lens. The rear surface marked with concentric circles serves as retina to measure the angular field of view (FOV). The small flat surface on the peak of the second lens is used to test signal to noise ratio (SNR). The single filament with 125 mu m diameter is used to check the co-alignment of preview and OCT scan. The empty chamber between the small plane of the second lens and the first surface of glass plate is used to measure the depth scaling of the OCT. The microspheres of 1 mu m diameter distributed uniformly in the phantom, which can test the lateral and the axial resolution of OCT equipment. Experimental results are presented to show the validity of the proposed tool. It is shown that the tool is able to be used in the calibration and quality control of retinal OCT.
引用
收藏
页数:8
相关论文
共 16 条
  • [1] Three-dimensional characterization of optical coherence tomography point spread functions with a nanoparticle-embedded phantom
    Agrawal, Anant
    Pfefer, T. Joshua
    Gilani, Naureen
    Drezek, Rebekah
    [J]. OPTICS LETTERS, 2010, 35 (13) : 2269 - 2271
  • [2] [Anonymous], 2015, 172SC7 ISOTC
  • [3] Retina-simulating phantom for optical coherence tomography
    Baxi, Jigesh
    Calhoun, William
    Sepah, Yasir Jamal
    Hammer, Daniel X.
    Ilev, Ilko
    Pfefer, T. Joshua
    Quan Dong Nguyen
    Agrawal, Anant
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2014, 19 (02)
  • [4] Model eyes with curved multilayer structure for the axial resolution evaluation of an ophthalmic optical coherence tomography device
    Cao, Zhenggang
    Ding, Zengqian
    Hu, Zhixiong
    Qiao, Wen
    Liu, Wenli
    Chen, Xiaojun
    [J]. JOURNAL OF INNOVATIVE OPTICAL HEALTH SCIENCES, 2018, 11 (03)
  • [5] Fabrication and characterization of a multilayered optical tissue model with embedded scattering microspheres in polymeric materials
    Chang, Robert C.
    Johnson, Peter
    Stafford, Christopher M.
    Hwang, Jeeseong
    [J]. BIOMEDICAL OPTICS EXPRESS, 2012, 3 (06): : 1326 - 1339
  • [6] Variations in optical coherence tomography resolution and uniformity: a multi-system performance comparison
    Fouad, Anthony
    Pfefer, T. Joshua
    Chen, Chao-Wei
    Gong, Wei
    Agrawal, Anant
    Tomlins, Peter H.
    Woolliams, Peter D.
    Drezek, Rebekah A.
    Chen, Yu
    [J]. BIOMEDICAL OPTICS EXPRESS, 2014, 5 (07): : 2066 - 2081
  • [7] Segmentation of Intra-Retinal Cysts From Optical Coherence Tomography Images Using a Fully Convolutional Neural Network Model
    Girish, G. N.
    Thakur, Bibhash
    Chowdhury, Sohini Roy
    Kothari, Abhishek R.
    Rajan, Jeny
    [J]. IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2019, 23 (01) : 296 - 304
  • [8] Hu Z., 2015, Acta Optica Sinica, V35
  • [9] [胡志雄 Hu Zhixiong], 2014, [光电工程, Opto-Electronic Engineering], V41, P28
  • [10] OPTICAL COHERENCE TOMOGRAPHY
    HUANG, D
    SWANSON, EA
    LIN, CP
    SCHUMAN, JS
    STINSON, WG
    CHANG, W
    HEE, MR
    FLOTTE, T
    GREGORY, K
    PULIAFITO, CA
    FUJIMOTO, JG
    [J]. SCIENCE, 1991, 254 (5035) : 1178 - 1181