CORRELATION BETWEEN RANGE OF MOTION AND IMPLANT FRACTURE: A 5 YEAR FOLLOW-UP OF 72 JOINTS IN 18 PATIENTS IN A RANDOMIZED STUDY COMPARING SWANSON AND AVANTA/SUTTER MCP SILICONE PROSTHESIS

被引:13
作者
Tagil, M. [1 ]
Geijer, M.
Malcus, P.
Kopylov, P.
机构
[1] Univ Lund Hosp, Dept Orthoped, Dept Radiol, S-22185 Lund, Sweden
基金
瑞典研究理事会;
关键词
MCP arthroplasty; rheumatoid arthritis; prosthetic fracture; range of motion; ulnar deviation; extension lag; RHEUMATOID-ARTHRITIS; METACARPOPHALANGEAL JOINTS; SUTTER;
D O I
10.1177/1753193409346793
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Eighteen out of 18 rheumatoid patients (at one centre of a two-centre 30 patient study previously reported) with a mean age of 56 years, and 72/72 operated joints were randomized to Avanta/Sutter or Swanson MCP prostheses and followed for 5 years. Both ulnar deviation and extension lag were improved already at 6 weeks and remained improved at 5 years. The Avanta prosthesis had a better range of motion (ROM) than the Swanson. Six of nine patients with Avanta/Sutter implants had at least one implant fracture compared to 1/9 patients with the Swanson implant (P=0.05) but fracture did not change the outcome subjectively. The ROM at 3 months correlated with the occurrence of an implant fracture at 5 years and a greater early ROM may be related to implant fracture. At 5 years patients remained satisfied and the deformities remained corrected.
引用
收藏
页码:743 / 747
页数:5
相关论文
共 12 条
[1]   Effectiveness of rheumatoid hand surgery: Contrasting perceptions of hand surgeons and rheumatologists [J].
Alderman, AK ;
Chung, KC ;
Kim, HM ;
Fox, DA ;
Ubel, PA .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2003, 28A (01) :3-11
[2]   High implant fracture incidence with Sutter silicone metacarpophalangeal joint arthroplasty [J].
Bass, RL ;
Stern, PJ ;
Nairus, JG .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1996, 21A (05) :813-818
[3]   STATISTICS NOTES .12. CALCULATING CORRELATION-COEFFICIENTS WITH REPEATED OBSERVATIONS .1. CORRELATION WITHIN-SUBJECTS [J].
BLAND, JM ;
ALTMAN, DG .
BRITISH MEDICAL JOURNAL, 1995, 310 (6977) :446-446
[4]  
Chung KC, 2000, J RHEUMATOL, V27, P1395
[5]   Metacarpophalangeal joint arthroplasty in rheumatoid arthritis - A long-term assessment [J].
Goldfarb, CA ;
Stern, PJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (10) :1869-1878
[6]   EFFECT SIZES FOR INTERPRETING CHANGES IN HEALTH-STATUS [J].
KAZIS, LE ;
ANDERSON, JJ ;
MEENAN, RF .
MEDICAL CARE, 1989, 27 (03) :S178-S189
[7]   A prospective randomized comparison of Sutter and Swanson silastic spacers [J].
McArthur, PA ;
Milner, RH .
JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME, 1998, 23B (05) :574-577
[8]   Avanta versus Swanson silicone implants in the MCP joint -: A prospective, randomized comparison of 30 patients followed for 2 years [J].
Möller, K ;
Sollerman, C ;
Geher, M ;
Kopylov, P ;
Tägil, M .
JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME, 2005, 30B (01) :8-13
[9]   Comparison of Swanson and Sutter metacarpophalangeal arthroplasties in patients with rheumatoid arthritis: A prospective and randomized trial [J].
Parkkila, T ;
Belt, EA ;
Hakala, M ;
Kautiainen, H ;
Leppilahti, J .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2005, 30A (06) :1276-1281
[10]   Survival and complications are similar after Swanson and Sutter implant replacement of metacarpophalangeal joints in patients with rheumatoid arthritis [J].
Parkkila, TJ ;
Belt, EA ;
Hakala, M ;
Kautiainen, HJ ;
Leppilahti, J .
SCANDINAVIAN JOURNAL OF PLASTIC AND RECONSTRUCTIVE SURGERY AND HAND SURGERY, 2006, 40 (01) :49-53