Assessment of Tumor Blood Flow in Breast Tumors With T1-Dynamic Contrast-Enhanced MR Imaging: Impact of Dose Reduction and the Use of a Prebolus Technique on Diagnostic Efficacy

被引:11
作者
Makkat, Smitha [1 ]
Luypaert, Robert [1 ]
Sourbron, Steven [2 ]
Stadnik, Tadeusz [1 ]
De Mey, Johan [1 ]
机构
[1] Vrije Univ Brussels, Dept Radiol, UZ Brussel, B-1090 Brussels, Belgium
[2] Univ Munich, Inst Clin Radiol, Grosshadern Hosp, Munich, Germany
关键词
blood flow; breast tumors; T1-DCE; prebolus; deconvolution; FAST GRADIENT-ECHO; MYOCARDIAL-PERFUSION; DECONVOLUTION; METABOLISM; CANCERS; DELAY;
D O I
10.1002/jmri.22077
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To prospectively evaluate whether dose reduction and the application of a prebolus technique can effectively alleviate signal saturation effects in T1 dynamic contrast enhanced (T1-DCE) magnetic resonance imaging (MRI) data in breast tumors and lead to increased diagnostic efficacy of the regional tumor blood flow (TBF) values obtained with deconvolution of T1-DCE MRI data. Materials and Methods: After obtaining informed consent, 23 women (32-80 years) with histologically proven breast tumors underwent MR mammography that included a whole-breast T1 DCE sequence. In the slice where the tumor enhanced maximally, a prebolus protocol was applied. One mL of Gd-DTPA solution at 2 mL/s. was injected at the beginning of a dynamic axial single slice inversion-prepared turbo field echo acquisition. At the 400th dynamic, a high dose of either 20 mL (15 patients) or 10 mL (8 patients) of contrast agent was injected at 2 mL/s and a further 400 dynamics were acquired. From the aortic prebolus curve an arterial input function (AIF) was reconstructed by time-shifting and adding the prebolus data. The relative enhancement time course from the tumor region of interest was deconvolved with the reconstructed AIF to generate the impulse response function, the maximum of which yielded the TBF. The institutional ethical committee approved the study. Results: Reducing the contrast dose by a factor of 2 led to an increase in diagnostic contrast for the TBF values of malignant and benign tumors by a factor of slightly more than 2. Addition of the prebolus technique improved this further by 45%. receiver operating characteristic analysis showed a significant increase of diagnostic yield related to the combined use of a prebolus and minimal dose. Conclusion: Using a prebolus approach provides an estimate of the unsaturated AIF, while reduction of the high-dose bolus minimizes possible saturation effects in the tumor time course.
引用
收藏
页码:556 / 561
页数:6
相关论文
共 21 条
[1]   Microcirculation and microvasculature in breast tumors: Pharmacokinetic analysis of dynamic MR image series [J].
Brix, G ;
Kiessling, F ;
Lucht, R ;
Darai, S ;
Wasser, K ;
Delorme, S ;
Griebel, J .
MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (02) :420-429
[2]  
Calamante F, 2000, MAGN RESON MED, V44, P466, DOI 10.1002/1522-2594(200009)44:3<466::AID-MRM18>3.0.CO
[3]  
2-M
[4]   INFLUENCE OF BOLUS VOLUME AND DOSE OF GADOLINIUM CHELATE FOR FIRST-PASS MYOCARDIAL PERFUSION MR-IMAGING STUDIES [J].
CANET, E ;
DOUEK, P ;
JANIER, M ;
BENDID, K ;
AMAYA, J ;
MILLET, P ;
REVEL, D .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1995, 5 (04) :411-415
[5]   Absolute myocardial perfusion in canines measured by using dual-bolus first-pass MR imaging [J].
Christian, TF ;
Rettmann, DW ;
Aletras, AH ;
Liao, SL ;
Taylor, JL ;
Balaban, RS ;
Arai, AE .
RADIOLOGY, 2004, 232 (03) :677-684
[6]   A METHOD OF COMPARING THE AREAS UNDER RECEIVER OPERATING CHARACTERISTIC CURVES DERIVED FROM THE SAME CASES [J].
HANLEY, JA ;
MCNEIL, BJ .
RADIOLOGY, 1983, 148 (03) :839-843
[7]   Inflow effect correction in fast gradient-echo perfusion imaging [J].
Ivancevic, MK ;
Zimine, I ;
Montet, X ;
Hyacinthe, JN ;
Lazeyras, F ;
Foxall, D ;
Vallée, JP .
MAGNETIC RESONANCE IN MEDICINE, 2003, 50 (05) :885-891
[8]   Time delay for arrival of MR contrast agent in collateral-dependent myocardium [J].
Jerosch-Herold, M ;
Hu, XD ;
Murthy, NS ;
Seethamraju, RT .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2004, 23 (07) :881-890
[9]   Magnetic resonance imaging of myocardial contrast enhancement with MS-325 and its relation to myocardial blood flow and the perfusion reserve [J].
Jerosch-Herold, M ;
Hu, XD ;
Murthy, NS ;
Rickers, C ;
Stillman, AE .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 18 (05) :544-554
[10]   Prebolus quantitative MR heart perfusion Imaging [J].
Köstler, H ;
Ritter, C ;
Lipp, M ;
Beer, M ;
Hahn, D ;
Sandstede, J .
MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (02) :296-299