Assessment of Corneal Keratometric and Astigmatism Measurements Using Verion System and Other Instruments in Cataract Patient

被引:8
作者
Fu, Yana [1 ,2 ]
Yu, Xinxin [1 ,2 ]
Savini, Giacomo [3 ]
Huang, Jinhai [1 ,2 ]
Lian, Hengli [1 ,2 ]
Song, Benhao [1 ,2 ]
Wang, Qinmei [1 ,2 ]
Zhao, Yune [1 ,2 ]
机构
[1] Wenzhou Med Univ, Sch Ophthalmol, Wenzhou, Peoples R China
[2] Wenzhou Med Univ, Eye Hosp, Wenzhou, Peoples R China
[3] GB Bietti Fdn, IRCCS, Rome, Italy
关键词
Keratometry; corneal astigmatism; Verion Image Guided System; manual keratometry; IOLMaster; Pentacam; INTRAOCULAR-LENS IMPLANTATION; IMAGE-GUIDED SYSTEM; REPEATABILITY; AGREEMENT; SURGERY; COMPARABILITY; BIOMETRY; DEVICES; EYES;
D O I
10.1080/02713683.2018.1488264
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare keratometry and corneal astigmatism measurements obtained by the Verion Image Guided System to those obtained by automated keratometry (IOLMaster), manual keratometry (YZ38, Suzhou 66 visual Polytron Technologies Inc., China), and Scheimpflug imaging (Pentacam HR) in eyes with cataract.Methods: A total of 149 cataractous eyes of 149 patients were examined and categorized as low astigmatism (<1.0D, N=50), moderate astigmatism (1.0-2.0D, N=49), and high astigmatism (>2.0D, N=50). The flattest and the steepest keratometric values (Kf and Ks), the magnitude of astigmatism (Ast), the orientation of the steepest meridian (Axis), and power vectors J(0) and J(45) were compared.Results: Keratometry readings (Kf, Ks, and Km) obtained by the Verion system showed statistically significant differences as compared to the other three devices. The magnitude of astigmatism (Ast) measured by the Verion (1.500.85D) was similar to that measured by the YZ38 (1.45 +/- 0.82D, P=0.110) and slightly lower than that by the IOLMaster (1.56 +/- 0.87D, P=0.014), but much higher than that by the Pentacam (1.36 +/- 0.81D, P=0.000) significantly. However, no statistically significant differences were observed in the J(0) and J(45) vectors in the whole sample, besides moderate and high astigmatism subgroups. A difference in astigmatism magnitude of 0.50D or more was detected between the Verion system and other three devices most commonly in moderate astigmatism eyes (16.3-24.5%). The difference in the location of the steep meridian was greater than 10degrees most frequently in the low astigmatism eyes (26.0-52.0%).Conclusions: Keratometric and astigmatic results obtained with Verion were not completely interchangeable with those obtained with IOLMaster, manual keratometry, and Pentacam. The individual differences in power and orientation of the corneal cylinder should be considered while selecting the appropriate toric intraocular lens.
引用
收藏
页码:1205 / 1214
页数:10
相关论文
共 26 条
[1]   Comparison of keratometric measurements obtained by the Verion Image Guided System with optical biometry and auto-keratorefractometer [J].
Asena, Leyla ;
Gungor, Sirel Gur ;
Akman, Ahmet .
INTERNATIONAL OPHTHALMOLOGY, 2017, 37 (02) :391-399
[2]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]  
Chang Minwook, 2012, Korean J Ophthalmol, V26, P10, DOI 10.3341/kjo.2012.26.1.10
[5]   Biometry and corneal astigmatism in cataract surgery candidates from Southern China [J].
Cui, Ying ;
Meng, Qianli ;
Guo, Haike ;
Zeng, Jin ;
Zhang, Hongyang ;
Zhang, Guanrong ;
Huang, Yu ;
Lan, Jianqing .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (10) :1661-1669
[6]   Comparison of different techniques of anterior chamber depth and keratometric measurements [J].
Elbaz, Uri ;
Barkana, Yaniv ;
Gerber, Yariv ;
Avni, Isaac ;
Zadok, David .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 143 (01) :48-53
[7]   Comparison Between Digital and Manual Marking for Toric Intraocular Lenses A Randomized Trial [J].
Elhofi, Abdel Hamid ;
Helaly, Hany Ahmed .
MEDICINE, 2015, 94 (38)
[8]   Effect of tear osmolarity on repeatability of keratometry for cataract surgery planning [J].
Epitropoulos, Alice T. ;
Matossian, Cynthia ;
Berdy, Gregg J. ;
Malhotra, Ranjan P. ;
Potvin, Richard .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (08) :1672-1677
[9]   Evaluation of Factors Influencing the Remaining Astigmatism After Toric Intraocular Lens Implantation [J].
Hirnschall, Nino ;
Hoffmann, Peter C. ;
Draschl, Petra ;
Maedel, Sophie ;
Findl, Oliver .
JOURNAL OF REFRACTIVE SURGERY, 2014, 30 (06) :394-+
[10]   Results of higher power toric intraocular lens implantation [J].
Hoffmann, Peter C. ;
Auel, Sebastian ;
Huetz, Werner W. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (08) :1411-1418