Pre-clinical evaluation of fretting-corrosion at stem-head and stem-cement interfaces of hip implants using in vitro and in silico models

被引:0
作者
dos Santos, Vinicius Oliveira [1 ]
Cubillos, Patricia Ortega [1 ]
dos Santos, Claudio Teodoro [2 ]
Fernandes, Wellington Gilbert [2 ]
Monteiro, Mauricio de Jesus [2 ]
Vieira Caminha, Ieda Maria [2 ]
Ocampo More, Ari Digiacomo [1 ,3 ]
de Mello Roesler, Carlos Rodrigo [1 ]
机构
[1] Fed Univ Santa Catarina UFSC, Univ Hosp, Dept Mech Engn, Biomech Engn Lab LEBm, Florianopolis, SC, Brazil
[2] Natl Inst Technol INT, Rio De Janeiro, RJ, Brazil
[3] Univ Fed Santa Catarina, Polydoro Ernani Sao Thiago Univ Hosp, Florianopolis, SC, Brazil
关键词
fretting-corrosion; in silico; in vitro; stem-cement interface; stem-head interface; TAPER JUNCTION; WEAR MECHANISMS; FINITE-ELEMENT; SURFACE FINISH; NECK INTERFACE; FEMORAL STEMS; MICROMOTION; COMPONENTS; DAMAGE;
D O I
10.1002/jbm.b.35110
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Prior to clinical use, the corrosion resistance of new prosthesis system must be verified. The fretting-corrosion mechanisms of total hip arthroplasty (THA) implants generate metal debris and ions that can increase the incidence of adverse tissue reactions. For cemented stems, there are at least two interfaces that can be damaged by fretting-corrosion: stem-head and stem-cement. This investigation aimed to evaluate, through in vitro and in silico analyses, fretting-corrosion at the stem-head and stem-cement interfaces, to determine which surface is most affected in pre-clinical testing and identify the causes associated with the observed behavior. Unimodular stems and femoral heads of three different groups were evaluated, defined according to the head/stem material as group I (SS/SS), group II (CoCr/SS), and group III (CoCr/CoCr). Seven pairs of stems and heads per group were tested: three pairs were subjected to material characterization, three pairs to in vitro fretting-corrosion testing, and one pair to geometric modeling in the in silico analysis. The absolute area of the stem body degraded was more than three times higher compared with the trunnion, for all groups. These results were corroborated by the in silico analysis results, which revealed that the average micromotion at the stem-cement interface (9.65-15.66 mu m) was higher than that at the stem-head interface (0.55-1.08 mu m). In conclusion, the degradation of the stem-cement interface is predominant in the pre-clinical set, indicating the need to consider the fretting-corrosion at the stem-cement interface during pre-clinical implant evaluations.
引用
收藏
页码:2521 / 2532
页数:12
相关论文
共 50 条
  • [41] Variability in stem taper surface topography affects the degree of corrosion and fretting in total hip arthroplasty
    Stockhausen, Kilian Elia
    Riedel, Christoph
    Belinski, Alex Victoria
    Rothe, Dorothea
    Gehrke, Thorsten
    Klebig, Felix
    Gebauer, Matthias
    Amling, Michael
    Citak, Mustafa
    Busse, Bjorn
    [J]. SCIENTIFIC REPORTS, 2021, 11 (01)
  • [42] Potential and frequency effects on fretting corrosion of Ti6Al4V and CoCrMo surfaces
    Swaminathan, Viswanathan
    Gilbert, Jeremy L.
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2013, 101 (09) : 2602 - 2612
  • [43] Theodore W., 2014, Applied Mechanics and Materials, V553, P287, DOI 10.4028/www.scientific.net/AMM.553.287
  • [44] Large-sliding contact elements accurately predict levels of bone-implant micromotion relevant to osseointegration
    Viceconti, M
    Muccini, R
    Bernakiewicz, M
    Baleani, M
    Cristofolini, L
    [J]. JOURNAL OF BIOMECHANICS, 2000, 33 (12) : 1611 - 1618
  • [45] Special modes of corrosion under physiological and simulated physiological conditions
    Virtanen, S.
    Milosev, I.
    Gomez-Barrena, E.
    Trebse, R.
    Salo, J.
    Konttinen, Y. T.
    [J]. ACTA BIOMATERIALIA, 2008, 4 (03) : 468 - 476
  • [46] Sensing Localized Surface Corrosion Damage of CoCrMo Alloys and Modular Tapers of Total Hip Retrievals Using Nearfield Electrochemical Impedance Spectroscopy
    Wiegand, Michael J.
    Shenoy, Aarti A.
    Littlejohn, Sara E.
    Gilbert, Jeremy L.
    [J]. ACS BIOMATERIALS SCIENCE & ENGINEERING, 2020, 6 (03) : 1344 - 1354
  • [47] Willert HG, 1996, CLIN ORTHOP RELAT R, P51
  • [48] Wear characteristics of sliding pairs of zirconia (Y-TZP) for hip endoprostheses
    Willmann, G
    Fruh, HJ
    Pfaff, HG
    [J]. BIOMATERIALS, 1996, 17 (22) : 2157 - 2162
  • [49] On the Formation Mechanism of Column Damage Within Modular Taper Junctions
    Zachariah, Zita
    Balachandran, Shanoob
    Liu, Zhilong
    Pourzal, Robin
    McCarthy, Stephanie M.
    Hall, Deborah J.
    Fischer, Alfons
    Raabe, Dierk
    Herbig, Michael
    [J]. JOURNAL OF ARTHROPLASTY, 2021, 36 (07) : 2603 - +
  • [50] Lessons learned from loosening of the McKee-Farrar metal-on-metal total hip replacement
    Zahiri, CA
    Schmalzried, TP
    Ebramzadeh, E
    Szuszczewicz, ES
    Salib, D
    Kim, C
    Amstutz, HC
    [J]. JOURNAL OF ARTHROPLASTY, 1999, 14 (03) : 326 - 332