The Performance of a 2-Dimensional Shear-Wave Elastography Technique for Predicting Different Stages of Liver Fibrosis Using Transient Elastography as the Control Method

被引:9
作者
Bende, Felix [1 ]
Sporea, Ioan [1 ]
Sirli, Roxana [1 ]
Nistorescu, Silviu [1 ]
Fofiu, Renata [1 ]
Baldea, Victor [1 ]
Popescu, Alina [1 ]
机构
[1] Victor Babes Univ Med & Pharm Timisoara, Dept Gastroenterol, Victor, NY, Romania
关键词
liver elastography; transient elastography; 2D SWE; cutoff values; LOGIQ S8; FORCE IMPULSE ELASTOGRAPHY; CHRONIC HEPATITIS-C; NONINVASIVE ASSESSMENT; STIFFNESS MEASUREMENT; XL PROBES; ARFI; FEASIBILITY; ACCURACY; BIOPSY;
D O I
10.1097/RUQ.0000000000000527
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim was to evaluate the performance of 2-dimensional (2D) shear-wave elastography from general electric (2D SWE-GE), implemented on the new LOGIQ S8 system, for the noninvasive assessment of liver fibrosis, and to identify liver stiffness (LS) cutoff values for predicting different stages of fibrosis using transient elastography (TE) as the control method. We included 179 consecutive subjects, with or without chronic hepatopathies, in whom LS was evaluated in the same session using 2 elastographic techniques: TE (FibroScan, EchoSens) and 2D SWE-GE (LOGIQ S8; GE Healthcare, Chalfont St Giles, United Kingdom). Reliable LS measurements were defined for TE: the median value of 10 measurements with a success rate of 60% or greater and an interquartile range/median ratio (IQR/M) < 0.30; for 2D SWE-GE: the median value of 10 measurements acquired in a homogenous area and IQR/M < 0.30. To discriminate between fibrosis stages by TE, we used the following cutoffs: F2-7; F3-9.5 and F4-12 kPa. Reliable LS measurements were obtained in 97.2% subjects by 2D SWE-GE and in 98.3% by TE (P = 0.72), so that 171 subjects were included for the final analysis. A good correlation was found between the LS values obtained by the 2 methods (r = 0.72, P < 0.0001). The best 2D SWE-GE cutoff value for F >= 2 was 6.9 kPa (areas under receiver operating characteristic [AUROC], 0.93; sensitivity, 85.8%; specificity, 90.2%), for F of 3 or greater, it was 8.2 kPa (AUROC, 0.93; sensitivity, 87.5%; specificity, 86.8%) and for F value of 4, it was 9.3 kPa (AUROC, 0.91; sensitivity, 85.7%; specificity, 81.2%). In conclusion, the best 2D SWE-GE (S8) cutoff values for predicting F2, F >= 3 and F = 4 were 6.9, 8.2, and 9.3 kPa.
引用
收藏
页码:97 / 104
页数:8
相关论文
共 39 条
[1]   Accuracy of Fibroscan, Compared With Histology, in Analysis of Liver Fibrosis in Patients With Hepatitis B or C: A United States Multicenter Study [J].
Afdhal, Nezam H. ;
Bacon, Bruce R. ;
Patel, Keyur ;
Lawitz, Eric J. ;
Gordon, Stuart C. ;
Nelson, David R. ;
Challies, Tracy L. ;
Nasser, Imad ;
Garg, Jyotsna ;
Wei, Lee-Jen ;
McHutchison, John G. .
CLINICAL GASTROENTEROLOGY AND HEPATOLOGY, 2015, 13 (04) :772-+
[2]   ASSESSING LIVER STIFFNESS BY 2-D SHEAR WAVE ELASTOGRAPHY IN A HEALTHY COHORT [J].
Bende, Felix ;
Mulabecirovic, Anesa ;
Sporea, Ioan ;
Popescu, Alina ;
Sirli, Roxana ;
Gilja, Odd Helge ;
Vesterhus, Mette ;
Havre, Roald Flesland .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2018, 44 (02) :332-341
[3]   Performance of 2D-SWE.GE for predicting different stages of liver fibrosis, using Transient Elastography as the reference method [J].
Bende, Felix ;
IoanSporea ;
Sirli, Roxana ;
Popescu, Alina ;
Mare, Ruxandra ;
Miutescu, Bogdan ;
Lupusoru, Raluca ;
Moga, Tudor ;
Pienar, Corina .
MEDICAL ULTRASONOGRAPHY, 2017, 19 (02) :143-149
[4]   The role of viscosity in the impulse diffraction field of elastic waves induced by the acoustic radiation force [J].
Bercoff, J ;
Tanter, M ;
Muller, M ;
Fink, M .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2004, 51 (11) :1523-1536
[5]   Determination of reliability criteria for liver stiffness evaluation by transient elastography [J].
Boursier, Jerome ;
Zarski, Jean-Pierre ;
de Ledinghen, Victor ;
Rousselet, Marie-Christine ;
Sturm, Nathalie ;
Lebail, Brigitte ;
Fouchard-Hubert, Isabelle ;
Gallois, Yves ;
Oberti, Frederic ;
Bertrais, Sandrine ;
Cales, Paul .
HEPATOLOGY, 2013, 57 (03) :1182-1191
[6]   Liver Stiffness in Nonalcoholic Fatty Liver Disease: A Comparison of Supersonic Shear Imaging, FibroScan, and ARFI With Liver Biopsy [J].
Cassinotto, Christophe ;
Boursier, Jerome ;
de Ledinghen, Victor ;
Lebigot, Jerome ;
Lapuyade, Bruno ;
Cales, Paul ;
Hiriart, Jean-Baptiste ;
Michalak, Sophie ;
Le Bail, Brigitte ;
Cartier, Victoire ;
Mouries, Amaury ;
Oberti, Frederic ;
Fouchard-Hubert, Isabelle ;
Vergniol, Julien ;
Aube, Christophe .
HEPATOLOGY, 2016, 63 (06) :1817-1827
[7]   Non-invasive assessment of liver fibrosis with impulse elastography: Comparison of Supersonic Shear Imaging with ARFI and FibroScan [J].
Cassinotto, Christophe ;
Lapuyade, Bruno ;
Mouries, Amaury ;
Hiriart, Jean-Baptiste ;
Vergniol, Julien ;
Gaye, Delphine ;
Castain, Claire ;
Le Bail, Brigitte ;
Chermak, Faiza ;
Foucher, Juliette ;
Laurent, Francois ;
Montaudon, Michel ;
De Ledinghen, Victor .
JOURNAL OF HEPATOLOGY, 2014, 61 (03) :550-557
[8]   Pitfalls of Liver Stiffness Measurement: A 5-Year Prospective Study of 13,369 Examinations [J].
Castera, Laurent ;
Foucher, Juliette ;
Bernard, Pierre-Henri ;
Carvalho, Francoise ;
Allaix, Daniele ;
Merrouche, Wassil ;
Couzigou, Patrice ;
de Ledinghen, Victor .
HEPATOLOGY, 2010, 51 (03) :828-835
[9]   Performance of Transient Elastography for the Staging of Liver Fibrosis in Patients with Chronic Hepatitis B: A Meta-Analysis [J].
Chon, Young Eun ;
Choi, Eun Hee ;
Song, Ki Jun ;
Park, Jun Yong ;
Kim, Do Young ;
Han, Kwang-Hyub ;
Chon, Chae Yoon ;
Ahn, Sang Hoon ;
Kim, Seung Up .
PLOS ONE, 2012, 7 (09)
[10]  
Dietrich CF, 2017, ULTRASCHALL MED, V38, pE52, DOI [10.1055/a-0641-0076, 10.1055/s-0043-103952]