Design and validation of an open-source modular Microplate Photoirradiation System for high-throughput photobiology experiments

被引:5
作者
Katz, Suzanna [1 ]
Backeris, Peter [2 ]
Merck, Christopher [3 ]
Suprun, Maria [4 ]
D'Souza, Sunita [1 ]
Bishop, David F. [5 ]
Desnick, Robert J. [5 ]
Moore, Kateri [1 ]
Ubarretxena-Belandia, Iban [6 ,7 ]
Lemischka, Ihor R. [1 ,6 ]
机构
[1] Icahn Sch Med Mt Sinai, Black Family Stem Cell Inst, Dept Cell Dev & Regenerat Biol, New York, NY 10029 USA
[2] Icahn Sch Med Mt Sinai, Dept Neurosurg, Sinai Biodesign, New York, NY 10029 USA
[3] Merck Engn LLC, Hamburg, NJ 07495 USA
[4] Icahn Sch Med Mt Sinai, Dept Pediat Allergy & Immunol, New York, NY 10029 USA
[5] Icahn Sch Med Mt Sinai, Dept Genet & Genom Sci, New York, NY 10029 USA
[6] Icahn Sch Med Mt Sinai, Dept Pharmacol Sci, New York, NY 10029 USA
[7] Univ Basque Country, UPV EHU, CSIC, Biofisika Inst, Leioa, Spain
来源
PLOS ONE | 2018年 / 13卷 / 10期
关键词
INDUCED PROTOPORPHYRIN IX; PHOTODYNAMIC THERAPY; 5-AMINOLEVULINIC ACID; METHYL AMINOLEVULINATE; ACTINIC KERATOSES; LIGHT; PORPHYRINS; SKIN; INACTIVATION; EXPRESSION;
D O I
10.1371/journal.pone.0203597
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Research in photobiology is currently limited by a lack of devices capable of delivering precise and tunable irradiation to cells in a high-throughput format. This limits researchers to using expensive commercially available or custom-built light sources which make it difficult to replicate, standardize, optimize, and scale experiments. Here we present an open-source Microplate Photoirradiation System (MPS) developed to enable high-throughput light experiments in standard 96 and 24-well microplates for a variety of applications in photobiology research. This open-source system features 96 independently controlled LEDs (4 LEDs per well in 24-well), Wi-Fi connected control and programmable graphical user interface (GUI) for control and programming, automated calibration GUI, and modular control and LED boards for maximum flexibility. A web-based GUI generates light program files containing irradiation parameters for groups of LEDs. These parameters are then uploaded wirelessly, stored and used on the MPS to run photoirradiation experiments inside any incubator. A rapid and semi-quantitative porphyrin metabolism assay was also developed to validate the system in wild-type fibroblasts. Protoporphyrin IX (PpIX) fluorescence accumulation was induced by incubation with 5-aminolevulinic acid (ALA), a photosensitization method leveraged clinically to destroy malignant cell types in a process termed photodynamic therapy (PDT), and cells were irradiated with 405nm light with varying irradiance, duration and pulsation parameters. Immediately after light treatment with the MPS, subsequent photobleaching was measured in live, adherent cells in both 96-well and a 24-well microplates using a microplate reader. Results demonstrate the utility and reliability of the Microplate Photoirradiation System to irradiate cells with precise irradiance and timing parameters in order to measure PpIx photobleaching kinetics in live adherent cells and perform comparable experiments with both 24 and 96 well microplate formats. The high-throughput capability of the MPS enabled measurement of enough irradiance conditions in a single microplate to fit PpIX fluorescence to a bioexponential decay model of photobleaching, as well as reveal a dependency of photobleaching on duty-cycle-but not frequency-in a pulsed irradiance regimen.
引用
收藏
页数:23
相关论文
共 3 条
  • [1] Effect of nitrogen source and acclimatization on specific growth rates of microalgae determined by a high-throughput in vivo microplate autofluorescence method
    Podevin, Michael
    De Francisci, Davide
    Holdt, Susan L.
    Angelidaki, Irini
    JOURNAL OF APPLIED PHYCOLOGY, 2015, 27 (04) : 1415 - 1423
  • [2] PICT: A low-cost, modular, open-source camera trap system to study plant-insect interactions
    Droissart, Vincent
    Azandi, Laura
    Onguene, Eric Rostand
    Savignac, Marie
    Smith, Thomas B.
    Deblauwe, Vincent
    METHODS IN ECOLOGY AND EVOLUTION, 2021, 12 (08): : 1389 - 1396
  • [3] Validation of a high-throughput fermentation system based on online monitoring of biomass and fluorescence in continuously shaken microtiter plates
    Kensy, Frank
    Zang, Emerson
    Faulhammer, Christian
    Tan, Rung-Kai
    Buechs, Jochen
    MICROBIAL CELL FACTORIES, 2009, 8