Development of fluorescent materials for diffuse fluorescence tomography standards and phantoms

被引:36
作者
Baeten, John
Niedre, Mark
Dunham, Joshua
Ntziachristos, Vasilis
机构
[1] Harvard Univ, Massachusetts Gen Hosp, Mol Imaging Ctr Mol Imaging Res, Lab Bio Opt, Charlestown, MA 02129 USA
[2] Harvard Univ, MIT, Technol Biomed Opt Summer Inst, Div Hlth Sci, Cambridge, MA USA
[3] Harvard Univ, Massachusetts Gen Hosp, Mol Imaging Ctr Mol Imaging Res, Lab Bio Opt, Boston, MA 02115 USA
关键词
D O I
10.1364/OE.15.008681
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The availability of fluorescence standards is necessary in the development of systems and methods for fluorescence imaging. In this study, two approaches for developing diffuse fluorescence materials to be used as standards or phantoms in diffuse fluorescent tomography applications were investigated. Specifically, silicone rubber and polyester casting resin were used as base materials, and silicone pigments or TiO2/ India Ink were added respectively to vary the optical properties. Characterization of the optical properties achieved was performed using time-resolved methods. Subsequently, different near-infrared fluorochromes were examined for imparting controlled and stable fluorescence properties. It was determined that hydrophobic fluorophores (IR 676 and IR 780 Iodide) suspended in dichloromethane and hydrophilic fluorophores (Cy5.5 and AF 750) suspended in methanol produced diffusive silicone and resin fluorescent materials, respectively. However only the hydrophobic fluorophores embedded within silicone resulted in the construction of a material with the characteristics of a standard, i.e. stability of fluorescence intensity with time and a linear dependence of normalized fluorescence intensity to fluorophore concentration. (c) 2007 Optical Society of America
引用
收藏
页码:8681 / 8694
页数:14
相关论文
共 18 条
[1]   Imaging of fluorescence in highly scattering media [J].
Chang, JH ;
Graber, HL ;
Barbour, RL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1997, 44 (09) :810-822
[2]   Three-dimensional, Bayesian image reconstruction from sparse and noisy data sets: Near-infrared fluorescence tomography [J].
Eppstein, MJ ;
Hawrysz, DJ ;
Godavarty, A ;
Sevick-Muraca, EM .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2002, 99 (15) :9619-9624
[3]   A DESIGN FOR A STABLE AND REPRODUCIBLE PHANTOM FOR USE IN NEAR-INFRARED IMAGING AND SPECTROSCOPY [J].
FIRBANK, M ;
DELPY, DT .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (06) :847-853
[4]   A submillimeter resolution fluorescence molecular imaging system for small animal imaging [J].
Graves, EE ;
Ripoll, J ;
Weissleder, R ;
Ntziachristos, V .
MEDICAL PHYSICS, 2003, 30 (05) :901-911
[5]  
Gurfinkel M, 2003, DIS MARKERS, V19, P107
[6]   Molecular imaging: Looking at problems, seeing solutions [J].
Herschman, HR .
SCIENCE, 2003, 302 (5645) :605-608
[7]   Combined optical and fluorescence imaging for breast cancer detection and diagnosis [J].
Jiang, HB ;
Ramesh, S ;
Bartlett, M .
CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 2000, 28 (3-4) :371-375
[8]   The inverse source problem based on the radiative transfer equation in optical molecular imaging [J].
Klose, AD ;
Ntziachristos, V ;
Hielscher, AH .
JOURNAL OF COMPUTATIONAL PHYSICS, 2005, 202 (01) :323-345
[9]   Fluorescence tomography with simulated data based on the equation of radiative transfer [J].
Klose, AD ;
Hielscher, AH .
OPTICS LETTERS, 2003, 28 (12) :1019-1021
[10]   Preparation and characterization of polyurethane optical phantoms [J].
Moffitt, Theodore ;
Chen, Yin-Chu ;
Prahl, Scott A. .
JOURNAL OF BIOMEDICAL OPTICS, 2006, 11 (04)