摘要
Purpose: Vascular elastography can visualize the strain distribution in the carotid artery, which could be useful in assessing the propensity of advanced plaques to rupture. In our previous studies, we demonstrated that sparse synthetic aperture (SA) imaging can produce high quality vascular strain elastograms. However, the low output power of SA imaging may hamper its clinical utility. In this study, we hypothesize that multi-element defocused emissions can overcome this limitation and improve the quality of the vascular strain elastograms. Methods: To assess the impact of attenuation on the elastographic performance of SA and (multi-element synthetic aperture) MSA imaging, we conducted experiments using heterogeneous vessel phantoms with ideal (0.1 dB cm(-1) MHz(-1)) and realistic (0.75 dB cm(-1) MHz(-1)) attenuation. Further, we validated the results of the phantom study in vivo, on a healthy male volunteer. All echo imaging was performed at a transmit frequency of 5 MHz, using a commercially available ultrasound scanner (Sonix RP, Ultrasonix Medical Corp., Richmond, BC, Canada). Results: The results from the phantom results demonstrated that plaques were visible in all strain elastograms, but those produced using MSA imaging had less artifacts. MSA imaging improved the elastographic contrast to noise ratio (CNRe) of the vascular elastograms by 14.58 dB relative to SA imaging, and 9.1 dB relative to compounded plane wave (CPW) imaging. Further, the results demonstrated that the elastographic performance of MSA imaging improved with increase in (a) the number of transmit-receive events and (b) the size of the transmit sub-aperture, up to 13 elements. Using larger sub-apertures degraded the elastographic performance. The results from the in vivo study were in good agreement with the phantom results. Specifically, using a defocused multi-element transmit sub-aperture for SA imaging improved the performance of vascular elastography. Conclusions: The results suggested that MSA imaging can produce reliable vascular stain elastograms. Future studies will involve using coded excitations to improve the CNRe and frame-rate of the proposed technique for vascular elastography. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:4068 / 4082
页数:15
相关论文
共 50 条
[21]   Imaging of contact acoustic nonlinearity using synthetic aperture technique [J].
Yun, Dongseok ;
Kim, Jongbeom ;
Jhang, Kyung-Young .
ULTRASONICS, 2013, 53 (07) :1349-1354
[22]   Synthetic Aperture Imaging Beyond Foreground Using Image Matting [J].
Yihua Pan ;
Bin Liu ;
Wenmin Yan ;
Yan Han ;
Yi Zhang .
Sensing and Imaging, 2019, 20
[23]   Synthetic Aperture Imaging Beyond Foreground Using Image Matting [J].
Pan, Yihua ;
Liu, Bin ;
Yan, Wenmin ;
Han, Yan ;
Zhang, Yi .
SENSING AND IMAGING, 2019, 20 (1)
[24]   PULSE WAVE IMAGING IN CAROTID ARTERY STENOSIS HUMAN PATIENTS IN VIVO [J].
Li, Ronny X. ;
Apostolakis, Iason Z. ;
Kemper, Paul ;
McGarry, Matthew D. J. ;
Ip, Ada ;
Connolly, Edward S. ;
McKinsey, James F. ;
Konofagou, Elisa E. .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2019, 45 (02) :353-366
[25]   Clinical evaluation of Synthetic Aperture Sequential Beamforming and Tissue Harmonic Imaging [J].
Brandt, Andreas Hjelm ;
Hemmsen, Martin Christian ;
Hansen, Peter Moller ;
Hansen, Kristoffer Lindskov ;
Lange, Theis ;
Jensen, Jorgen Arendt ;
Nielsen, Michael Bachmann .
2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2014, :1312-1315
[26]   Modified multi-element synthetic transmit aperture method for ultrasound imaging: A tissue phantom study [J].
Tasinkevych, Y. ;
Klimonda, Z. ;
Lewandowski, M. ;
Nowicki, A. ;
Lewin, A. .
ULTRASONICS, 2013, 53 (02) :570-579
[27]   Focus Measure for Synthetic Aperture Imaging Using a Deep Convolutional Network [J].
Pei, Zhao ;
Huang, Li ;
Zhang, Yanning ;
Ma, Miao ;
Peng, Yali ;
Yang, Yee-Hong .
IEEE ACCESS, 2019, 7 :19762-19774
[28]   All-In-Focus Synthetic Aperture Imaging Using Image Matting [J].
Pei, Zhao ;
Chen, Xida ;
Yang, Yee-Hong .
IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS FOR VIDEO TECHNOLOGY, 2018, 28 (02) :288-301
[29]   Synthetic aperture imaging using pixel labeling via energy minimization [J].
Pei, Zhao ;
Zhang, Yanning ;
Chen, Xida ;
Yang, Yee-Hong .
PATTERN RECOGNITION, 2013, 46 (01) :174-187
[30]   Noninvasive real-time imaging of intima thickness after rat carotid artery balloon injury using ultrasound biomicroscopy [J].
Razuvaev, Anton ;
Lund, Kent ;
Roy, Joy ;
Hedin, Ulf ;
Caidahl, Kenneth .
ATHEROSCLEROSIS, 2008, 199 (02) :310-316