Effects of Exposure Parameters and Voxel Size for Cone-Beam Computed Tomography on the Image Matching Accuracy with an Optical Dental Scan Image: An In Vitro Study

被引:6
作者
Mai, Hang-Nga [1 ]
Lee, Du-Hyeong [1 ,2 ]
机构
[1] Kyungpook Natl Univ, Inst Translat Res Dent, 2177 Dalgubeoldae Ro, Daegu 41940, South Korea
[2] Kyungpook Natl Univ, Sch Dent, Dept Prosthodont, 2177 Dalgubeoldae Ro, Daegu 41940, South Korea
基金
新加坡国家研究基金会;
关键词
DIAGNOSTIC ABILITY; ROOT FRACTURES; RECONSTRUCTIONS; OPTIMIZATION; DENTISTRY; QUALITY; CT;
D O I
10.1155/2021/6971828
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
This study is aimed at assessing the effects of exposure parameters and voxel size for cone-beam computed tomography (CBCT) on the image matching accuracy with an optical dental scan image. CBCT and optical scan images of a dry human mandible were obtained. Different CBCT settings were used: tube voltage, 60, 80, and 100 kVp; tube current, 6 and 8 mA; and voxel size, 100, 200, and 300 mu m. Image matching between the CBCT and optical scan images was performed using implant planning software by dental professionals (n=18). The image matching accuracy in each combination of CBCT settings was evaluated by assessing the linear discrepancy between the three-dimensionally reconstructed radiological image and the registered optical scan image using an image analysis software program. The Kruskal-Wallis test and a post hoc Mann-Whitney U test with Bonferroni correction were used to compare the accuracy of image registration between the groups (alpha=0.05). Overall, the image matching accuracy was not significantly different between tube voltage and current settings; however, significantly higher image registration errors were found at the combination of 100 kVp tube voltage/8 mA tube current (F=8.44, P<0.001). Changes in voxel sizes did not significantly interfere with the image registration results. No interaction was found among voltage, current, and voxel size in terms of image registration accuracy (F=2.022, P=0.091). Different exposure parameter settings in tube voltage and tube current did not significantly influence the image matching accuracy between CBCT and optical dental scan images; however, a high radiation dose could be inappropriate. The image matching accuracy was not significantly affected by changing the voxel sizes of CBCT.
引用
收藏
页数:6
相关论文
共 35 条
[1]   Trueness of cone beam computed tomography versus intra-oral scanner derived three-dimensional digital models: An ex vivo study [J].
Al-Rimawi, Ali ;
Shaheen, Eman ;
Albdour, Emad A. ;
Shujaat, Sohaib ;
Politis, Constantinus ;
Jacobs, Reinhilde .
CLINICAL ORAL IMPLANTS RESEARCH, 2019, 30 (06) :498-504
[2]   Image quality assessment of three cone beam CT machines using the SEDENTEXCT CT phantom [J].
Bamba, J. ;
Araki, K. ;
Endo, A. ;
Okano, T. .
DENTOMAXILLOFACIAL RADIOLOGY, 2013, 42 (08)
[3]   Temporomandibular Joint Anatomy Assessed by CBCT Images [J].
Caruso, Silvia ;
Storti, Ennio ;
Nota, Alessandro ;
Ehsani, Shideh ;
Gatto, Roberto .
BIOMED RESEARCH INTERNATIONAL, 2017, 2017
[4]   CONE-BEAM CT FOR RADIOTHERAPY APPLICATIONS [J].
CHO, PS ;
JOHNSON, RH ;
GRIFFIN, TW .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (11) :1863-1883
[5]  
Costa AN, 2018, J BRAS PNEUMOL, V44, P424, DOI [10.4317/jced.54500, 10.1590/s1806-37562017000000209, 10.1590/S1806-37562017000000209]
[6]   Accuracy of linear measurements from cone-beam computed tomography-derived surface models of different voxel sizes [J].
Damstra, Janalt ;
Fourie, Zacharias ;
Slater, James J. R. Huddleston ;
Ren, Yijin .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2010, 137 (01) :16.e1-16.e6
[7]   Optimization of cone beam CT exposure for pre-surgical evaluation of the implant site [J].
Dawood, A. ;
Brown, J. ;
Sauret-Jackson, V. ;
Purkayastha, S. .
DENTOMAXILLOFACIAL RADIOLOGY, 2012, 41 (01) :70-74
[8]   Accuracy of in vitro mandibular volumetric measurements from CBCT of different voxel sizes with different segmentation threshold settings [J].
Dong, Ting ;
Xia, Lunguo ;
Cai, Chenglin ;
Yuan, Lingjun ;
Ye, Niansong ;
Fang, Bing .
BMC ORAL HEALTH, 2019, 19 (01)
[9]   Detection of alveolar bone defects with three different voxel sizes of cone-beam computed tomography: an in vitro study [J].
Dong, Ting ;
Yuan, Lingjun ;
Liu, Lu ;
Qian, Yifeng ;
Xia, Lunguo ;
Ye, Niansong ;
Fang, Bing .
SCIENTIFIC REPORTS, 2019, 9 (1)
[10]   Soft tissue coverage on the segmentation accuracy of the 3D surface-rendered model from cone-beam CT [J].
Dusseldorp, J. K. ;
Stamatakis, H. C. ;
Ren, Y. .
CLINICAL ORAL INVESTIGATIONS, 2017, 21 (03) :921-930