Cone-beam CT dose and imaging performance evaluation with a modular, multipurpose phantom

被引:22
作者
Siewerdsen, J. H. [1 ]
Uneri, A. [1 ]
Hernandez, A. M. [2 ]
Burkett, G. W. [2 ]
Boone, J. M. [2 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA
[2] Univ Calif Davis, Dept Radiol, Sacramento, CA 95817 USA
关键词
cone-beam CT; dosimetry; image quality; phantom; technical assessment; BREAST CT; COMPUTED-TOMOGRAPHY; SPATIAL-RESOLUTION; ARM; GUIDANCE; QUALITY; SURGERY; SYSTEMS; MOBILE; MODEL;
D O I
10.1002/mp.13952
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose A modular phantom for dosimetry and imaging performance evaluation in cone-beam computed tomography (CBCT) is reported, providing a tool for quantitative technical assessment that can be adapted to a broad variety of CBCT imaging configurations and clinical applications. Methods The phantom presents a set of modules that can be ordered in various configurations suitable to a particular CBCT system. Modules include slabs containing a uniform medium, low-contrast inserts, line-spread features, and disk features suitable to measurement of image uniformity, noise, noise-power spectrum (NPS), contrast, contrast-to-noise ratio (CNR), Hounsfield (HU) accuracy, linearity, spatial resolution modulation transfer function (MTF), and magnitude of cone-beam artifact. Automated software recognizes the phantom configuration in DICOM images and provides structured reporting of such test measures. In any modular configuration, the phantom permits measurement of air kerma in central and peripheral locations with an air ionization chamber (e.g., Farmer chamber). The utility and adaptability of the phantom were demonstrated across a spectrum of CBCT systems, including scanners for orthopaedic imaging (Carestream OnSight 3D, Rochester NY), breast imaging (Doheny prototype, UC Davis), image-guided surgery (IGS, Medtronic O-arm, Littleton MA), angiography (Siemens Artis Zeego, Forcheim Germany), and image-guided radiation therapy (IGRT, Elekta Synergy XVI, Stockholm Sweden). Results The phantom provided a consistent platform for quantitative assessment of dose and imaging performance compatible with a broad spectrum of CBCT systems. The purpose of the survey was not to obtain head-to-head performance comparison of systems designed for such distinct clinical applications. Rather, the survey demonstrated the suitability of the phantom to a broad spectrum of systems in a manner that provides characterization pertinent to disparate applications and imaging tasks. For example: the orthopaedic CBCT system (pertinent clinical tasks relating to high-resolution bone imaging) was shown to achieve MTF consistent with imaging of high-contrast trabecular bone structures (i.e., the MTF reduced to 10% at spatial frequency, f10 = 1.2 mm(-1)); the breast system (even higher-resolution imaging of microcalcifications) exhibited f10 = 2.2 mm(-1); the IGS system (tasks including both bone and soft-tissue contrast resolution) provided f10 = 0.9 mm(-1) and soft-tissue CNR = 1.64; the angiography system (soft-tissue body interventions) demonstrated CNR = 1.2 in soft tissues approximating liver lesions; and the IGRT system (pertinent tasks emphasizing HU linearity and image uniformity) showed linear response with HU values (R2 = 1), with a cupping artifact (tcup = 5.8%) due to x-ray scatter. Conclusions The phantom provides an adaptable, quantitative basis for CBCT dosimetry and imaging performance evaluation suitable to a broad variety of CBCT systems. The dosimetry and image quality metrics are consistent with up-to-date methods for rigorous, quantitative, physics testing and should be suitable to emerging standards for CBCT quality assurance.
引用
收藏
页码:467 / 479
页数:13
相关论文
共 23 条
  • [1] Imaging doses from the Elekta Synergy X-ray cone beam CT system
    Amer, A.
    Marchant, T.
    Sykes, J.
    Czajka, J.
    Moore, C.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (954) : 476 - 482
  • [2] Scattered data interpolation methods for electronic imaging systems: a survey
    Amidror, I
    [J]. JOURNAL OF ELECTRONIC IMAGING, 2002, 11 (02) : 157 - 176
  • [3] Characteristics of a newly developed dentomaxillofacial X-ray cone beam CT scanner (CB MercuRay™):: system configuration and physical properties
    Araki, K
    Maki, K
    Seki, K
    Sakamaki, K
    Harata, Y
    Sakaino, R
    Okano, T
    Seo, K
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2004, 33 (01) : 51 - 59
  • [4] Investigation of the radiation dose from cone-beam CT for image-guided radiotherapy: A comparison of methodologies
    Buckley, Jarryd G.
    Wilkinson, Dean
    Malaroda, Alessandra
    Metcalfe, Peter
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (01): : 174 - 183
  • [5] Intraoperative cone-beam CT for guidance of head and neck surgery: Assessment of dose and image quality using a C-arm prototype
    Daly, M. J.
    Siewerdsen, J. H.
    Moseley, D. J.
    Jaffray, D. A.
    Irish, J. C.
    [J]. MEDICAL PHYSICS, 2006, 33 (10) : 3767 - 3780
  • [6] Cone beam CT dosimetry: A unified and self-consistent approach including all scan modalities-With or without phantom motion
    Dixon, Robert L.
    Boone, John M.
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 2703 - 2718
  • [7] Dose and image quality for a cone-beam C-arm CT system
    Fahrig, Rebecca
    Dixon, Robert
    Payne, Thomas
    Morin, Richard L.
    Ganguly, Arundhuti
    Strobel, Norbert
    [J]. MEDICAL PHYSICS, 2006, 33 (12) : 4541 - 4550
  • [8] Evolution of spatial resolution in breast CT at UC Davis
    Gazi, Peymon M.
    Yang, Kai
    Burkett, George W., Jr.
    Aminololama-Shakeri, Shadi
    [J]. MEDICAL PHYSICS, 2015, 42 (04) : 1973 - 1981
  • [9] Cone-beam imaging: Applications in ENT
    Hodez, C.
    Griffaton-Taillandier, C.
    Bensimon, I.
    [J]. EUROPEAN ANNALS OF OTORHINOLARYNGOLOGY-HEAD AND NECK DISEASES, 2011, 128 (02) : 65 - 78
  • [10] The current status of cone beam computed tomography imaging in orthodontics
    Kapila, S.
    Conley, R. S.
    Harrell, W. E., Jr.
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2011, 40 (01) : 24 - 34