Automated fluorescence quantification of extracellular vesicles collected from blood plasma using dielectrophoresis†

被引:29
作者
Gustafson, Kyle T. [1 ,2 ]
Huynh, Katherine T. [1 ,2 ]
Heineck, Daniel [1 ,2 ]
Bueno, Jesus [1 ,2 ]
Modestino, Augusta [1 ,2 ]
Kim, Sejung [1 ,2 ]
Gower, Austin [1 ]
Armstrong, Randall [1 ]
Schutt, Carolyn E. [1 ,2 ]
Ibsen, Stuart D. [1 ,2 ]
机构
[1] Oregon Hlth & Sci Univ, Knight Canc Inst, Canc Early Detect Adv Res Ctr, Portland, OR 97201 USA
[2] Oregon Hlth & Sci Univ, Sch Med, Dept Biomed Engn, Portland, OR 97201 USA
关键词
EXOSOMES; CANCER; BIOMARKERS; PLATFORM; NANOPARTICLES; VIRUS; IMAGE;
D O I
10.1039/d0lc00940g
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Tumor-secreted exosomes and other extracellular vesicles (EVs) in circulation contain valuable biomarkers for early cancer detection and screening. We have previously demonstrated collection of cancer-derived nanoparticles (NPs) directly from whole blood and plasma with a chip-based technique that uses a microelectrode array to generate dielectrophoretic (DEP) forces. This technique enables direct recovery of NPs from whole blood and plasma. The biomarker payloads associated with collected particles can be detected and quantified with immunostaining. Accurately separating the fluorescence intensity of stained biomarkers from background (BG) levels becomes a challenge when analyzing the blood from early-stage cancer patients in which biomarker concentrations are low. To address this challenge, we developed two complementary techniques to standardize the quantification of fluorescently immunolabeled biomarkers collected and concentrated at predictable locations within microfluidic chips. The first technique was an automated algorithm for the quantitative analysis of fluorescence intensity at collection regions within the chip compared to levels at adjacent regions. The algorithm used predictable locations of particle collection within the chip geometry to differentiate regions of collection and BG. We successfully automated the identification and removal of optical artifacts from quantitative calculations. We demonstrated that the automated system performs nearly the same as a human user following a standard protocol for manual artifact removal with Pearson's r-values of 0.999 and 0.998 for two different biomarkers (n = 36 patients). We defined a usable dynamic range of fluorescence intensities corresponding to 1 to 2000 arbitrary units (a.u.). Fluorescence intensities within the dynamic range increased linearly with respect to exposure time and particle concentration. The second technique was the implementation of an internal standard to adjust levels of biomarker fluorescence based on the relative collection efficiency of the chip. Use of the internal standard reduced variability in measured biomarker levels due to differences in chip-to-chip collection efficiency, especially at low biomarker concentrations. The internal standard did not affect linear trends between fluorescence intensity and exposure time. Adjustments using the internal standard improved linear trends between fluorescence intensity and particle concentration. The optical quantification techniques described in this paper can be easily adapted for other lab-on-a-chip platforms that have predefined regions of biomarker or particle collection and that rely on fluorescence detection.
引用
收藏
页码:1318 / 1332
页数:16
相关论文
共 44 条
[1]   The feasibility of using dielectrophoresis for isolation of glioblastoma subpopulations with increased stemness [J].
Alinezhadbalalami, Nastaran ;
Douglas, Temple A. ;
Balani, Nikita ;
Verbridge, Scott S. ;
Davalos, Rafael V. .
ELECTROPHORESIS, 2019, 40 (18-19) :2592-2600
[2]   Quantitative Proteomic Analysis of Serum Exosomes from Patients with Locally Advanced Pancreatic Cancer Undergoing Chemoradiotherapy [J].
An, Mingrui ;
Lohse, Ines ;
Tan, Zhijing ;
Zhu, Jianhui ;
Wu, Jing ;
Kurapati, Himabindu ;
Morgan, Meredith A. ;
Lawrence, Theodore S. ;
Cuneo, Kyle C. ;
Lubman, David M. .
JOURNAL OF PROTEOME RESEARCH, 2017, 16 (04) :1763-1772
[3]   Electrohydrodynamics and dielectrophoresis in microsystems:: scaling laws [J].
Castellanos, A ;
Ramos, A ;
González, A ;
Green, NG ;
Morgan, H .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2003, 36 (20) :2584-2597
[4]   Isolation and identification of Listeria monocytogenes utilizing DC insulator-based dielectrophoresis [J].
Crowther, Claire V. ;
Hilton, Shannon Huey ;
Kemp, LaKeta ;
Hayes, Mark A. .
ANALYTICA CHIMICA ACTA, 2019, 1068 :41-51
[5]  
Ding J, 2016, ANALYST, V141, P1997, DOI [10.1039/c5an02430g, 10.1039/C5AN02430G]
[6]   A feasibility study for enrichment of highly aggressive cancer subpopulations by their biophysical properties via dielectrophoresis enhanced with synergistic fluid flow [J].
Douglas, Temple Anne ;
Cemazar, Jaka ;
Balani, Nikita ;
Sweeney, Daniel C. ;
Schmelz, Eva M. ;
Davalos, Rafael V. .
ELECTROPHORESIS, 2017, 38 (11) :1507-1514
[7]   Exosomes as Biomarker Treasure Chests for Prostate Cancer [J].
Duijvesz, Diederick ;
Luider, Theo ;
Bangma, Chris H. ;
Jenster, Guido .
EUROPEAN UROLOGY, 2011, 59 (05) :823-831
[8]   Dielectrophoretic characterization and separation of monocytes and macrophages using 3D carbon-electrodes [J].
Elitas, Meltem ;
Yildizhan, Yagmur ;
Islam, Monsur ;
Martinez-Duarte, Rodrigo ;
Ozkazanc, Didem .
ELECTROPHORESIS, 2019, 40 (02) :315-321
[9]   Necrosis, a well-orchestrated form of cell demise: Signalling cascades, important mediators and concomitant immune response [J].
Festjens, Nele ;
Vanden Berghe, Tom ;
Vandenabeele, Peter .
BIOCHIMICA ET BIOPHYSICA ACTA-BIOENERGETICS, 2006, 1757 (9-10) :1371-1387
[10]   A prefilled, ready-to-use electrophoresis based lab-on-a-chip device for monitoring lithium in blood [J].
Floris, Arjan ;
Staal, Steven ;
Lenk, Stefan ;
Staijen, Erik ;
Kohlheyer, Dietrich ;
Eijkel, Jan ;
van den Berg, Albert .
LAB ON A CHIP, 2010, 10 (14) :1799-1806