Performance of an ultra-low elution-volume 96-well plate: drug discovery and development applications

被引:26
作者
Mallet, CR [1 ]
Lu, ZL [1 ]
Fisk, R [1 ]
Mazzeo, JR [1 ]
Neue, UD [1 ]
机构
[1] Waters Corp, Milford, MA 01757 USA
关键词
D O I
10.1002/rcm.888
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Recently, sample preparation has been considered to be the major cause of bottlenecks during high-throughput analysis. With the assistance of robotic liquid handlers and the 96-well plate format, more samples can be prepared for subsequent liquid chromatography/tandem mass spectrometry (LC/MS/MS) analysis. Protein precipitation is still widely used despite potential loss of sensitivity or variable results due to ion suppression. The use of solid-phase extraction (SPE) clearly gives superior results but may not be as cost effective as protein precipitation due to the labor and material costs associated with the process. Here, a novel 96-well SPE plate is described that was designed to minimize the elution volume required for quantitative elution of analytes. The plate is packed with 2 mg of a high-capacity SPE sorbent that allows loading of up to 750 muL of plasma, while the novel design permits elution with as little as 25 muL. Therefore, the plate offers up to a 15-fold increase in sample concentration. The evaporation and reconstitution step that is typically required in SPE is avoided due to the concentrating ability of the plate. Examples of applications in drug discovery/development are shown and results are compared to protein precipitation. Excellent sensitivity and linearity are demonstrated. Copyright (C) 2002 John Wiley Sons, Ltd.
引用
收藏
页码:163 / 170
页数:8
相关论文
共 20 条
  • [1] [Anonymous], RAPID COMMUN MASS SP
  • [2] Use of generic fast gradient liquid chromatography tandem mass spectroscopy in quantitative bioanalysis
    Ayrton, J
    Dear, GJ
    Leavens, WJ
    Mallett, DN
    Plumb, RS
    [J]. JOURNAL OF CHROMATOGRAPHY B, 1998, 709 (02): : 243 - 254
  • [3] BLEVINS DD, 1998, LC GC CURRENT TR MAY, pS16
  • [4] Cabrera K, 2000, HRC-J HIGH RES CHROM, V23, P93
  • [5] Straightforward solid-phase extraction method for the determination of verapamil and its metabolite in plasma in a 96-well extraction plate
    Cheng, YF
    Neue, UD
    Bean, L
    [J]. JOURNAL OF CHROMATOGRAPHY A, 1998, 828 (1-2) : 273 - 281
  • [6] Cheng YF, 2001, RAPID COMMUN MASS SP, V15, P141, DOI 10.1002/1097-0231(20010130)15:2<141::AID-RCM201>3.0.CO
  • [7] 2-I
  • [8] Ermer J, 2000, BIOMED CHROMATOGR, V14, P373, DOI 10.1002/1099-0801(200010)14:6<373::AID-BMC29>3.0.CO
  • [9] 2-S
  • [10] Reduction of signal suppression effects in ESI-MS using a nanosplitting device
    Gangl, ET
    Annan, M
    Spooner, N
    Vouros, P
    [J]. ANALYTICAL CHEMISTRY, 2001, 73 (23) : 5635 - 5644