In vivo dosimetry of high-dose-rate brachytherapy:: Study on 61 head-and-neck cancer patients using radiophotoluminescence glass dosimeter

被引:20
|
作者
Nose, T
Koizumi, M
Yoshida, K
Nishiyama, K
Sasaki, J
Ohnishi, T
Peiffert, D
机构
[1] Toyonaka City Hosp, Dept Radiol, Toyonaka, Osaka 5608565, Japan
[2] Osaka Med Ctr, Dept Radiat Oncol, Osaka, Japan
[3] Osaka Natl Hosp, Dept Radiol, Osaka, Japan
[4] Ctr Alexis Vautrin, Dept Brachytherapy, Nancy, France
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2005年 / 61卷 / 03期
关键词
in vivo dosimetry; radiophotoluminescence glass dosimeter; brachytherapy; high-dose-rate;
D O I
10.1016/j.ijrobp.2004.10.031
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The largest in vivo dosimetry study for interstitial brachytherapy yet examined was performed using new radiophotoluminescence glass dosimeters (RPLGDs). Based on the results, a dose prescription technique achieving high reproducibility and eliminating large hyperdose sleeves was studied. Methods and Materials: For 61 head-and-neck cancer patients who underwent high-dose-rate interstitial brachytherapy, new RPLGDs were used for an in vivo study. The Paris System was used for implant. An arbitrary isodose surface was selected for dose prescription. Locations of 83 dosimeters were categorized as on target (n = 52) or on nontarget organ (n = 31) and were also scaled according to % basal dose isodose surface (% BDIS). Compatibility (measured dose/calculated dose) was analyzed according to location. The hyperdose sleeve was assessed in terms of prescription surface expressed in % BDIS. Results: The spread of compatibilities was larger for on nontarget organ (1.06+/-0.32) than for on target (0.87 +/- 0.17, p = 0.01). Within on target RPLGDs, compatibility on < 95 % BDIS (0.95+/- 0.10) was better than on greater than or equal to95 % BDIS (0.84+/-0.18, p = 0.02). The number of patients with diameter of hyperdose sleeve greater than or equal to10 mm was increased with a dose prescription to <77% BDIS (p = 0.046). For nontarget organs, the maximal positive deviation was 84% of the calculated dose. Conclusions: Dose prescription is recommended to >77% and <95% BDIS for reproducibility and elimination of excessive hyperdose sleeve. For organs at risk, radioprotection should be considered even when calculated dose seems sufficiently low. Further development of planning software is necessary to prevent overestimation. (C) 2005 Elsevier Inc.
引用
收藏
页码:945 / 953
页数:9
相关论文
共 50 条
  • [1] In vivo dosimetry of high-dose-rate interstitial brachytherapy in the pelvic region:: Use of a radiophotoluminescence glass dosimeter for measurement of 1004 points in 66 patients with pelvic malignancy
    Nose, Takayuki
    Koizumi, Masahiko
    Yoshida, Ken
    Nishiyama, Kinji
    Sasaki, Junichi
    Hnishi, Takeshi
    Kozuka, Takuyo
    Gomi, Kotaro
    Oguchi, Masahiko
    Sumida, Iori
    Takahashi, Yutaka
    Ito, Akira
    Yamashita, Takashi
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (02): : 626 - 633
  • [2] The American Brachytherapy Society recommendations for high-dose-rate brachytherapy for head-and-neck carcinoma
    Nag, S
    Cano, ER
    Demanes, DJ
    Puthawala, AA
    Vikram, B
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 50 (05): : 1190 - 1198
  • [3] THE POTENTIAL USES OF HIGH-DOSE-RATE BRACHYTHERAPY IN PATIENTS WITH HEAD AND NECK-CANCER
    DONATH, D
    VUONG, T
    SHENOUDA, G
    MACDONALD, B
    TABAH, R
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 1995, 252 (06) : 321 - 324
  • [4] HIGH-DOSE-RATE INTRAOPERATIVE RADIATION THERAPY FOR RECURRENT HEAD-AND-NECK CANCER
    Perry, David J.
    Chan, Kelvin
    Wolden, Suzanne
    Zelefsky, Michael J.
    Chiu, Johnny
    Cohen, Gilad
    Zaider, Marco
    Kraus, Dennis
    Shah, Jatin
    Lee, Nancy
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (04): : 1140 - 1146
  • [5] Interstitial high-dose-rate brachytherapy combined with cervical dissection on head and neck cancer
    Pellizzon, ACA
    Novaes, PERD
    Maia, MAC
    Ferrigno, R
    Fogarolli, R
    Salvajoli, JV
    Kowalski, LP
    HEAD AND NECK-JOURNAL FOR THE SCIENCES AND SPECIALTIES OF THE HEAD AND NECK, 2005, 27 (12): : 1035 - 1041
  • [6] Preliminary report of pulsed dose rate Brachytherapy in head-and-neck cancer
    Ziemlewski, Adam
    Zienkiewicz, Jozef
    Serkies, Krystyna
    Badzio, Andrzej
    STRAHLENTHERAPIE UND ONKOLOGIE, 2007, 183 (09) : 512 - 516
  • [7] Pilot study of intraoperative high dose rate brachytherapy for head and neck cancer
    Nag, S
    Schuller, D
    Pak, V
    Young, D
    Grecula, J
    Bauer, C
    Samsami, N
    RADIOTHERAPY AND ONCOLOGY, 1996, 41 (02) : 125 - 130
  • [8] In vivo dosimetry in high dose rate brachytherapy for cervical cancer treatments
    Gonzalez-Azcorra, S. A.
    Mota-Garcia, A.
    Poitevin-Chacon, M. A.
    Santamaria-Torruco, B. J.
    Rodriguez-Ponce, M.
    Herrera-Martinez, F. P.
    Gamboa, de Buen, I
    Ruiz-Trejo, C.
    Buenfil, A. E.
    MEDICAL PHYSICS, 2008, 1032 : 149 - +
  • [9] A phamtom study: In vivo rectal dosimetry of high dose rate brachytherapy in cervical cancer
    Moonkum, Nutthapong
    Turathong, Suriyaporn
    Pinitpatcharalert, Attapol
    Seechompoo, Janrudi
    Suetong, Chantiya
    Danthanavat, Nuttapong
    Tochaikul, Gunjanaporn
    APPLIED RADIATION AND ISOTOPES, 2023, 192
  • [10] Characterization of microMOSFET detectors for in vivo dosimetry in high-dose-rate brachytherapy with 192Ir
    Ruiz-Arrebola, Samuel
    Fabregat-Borras, Rosa
    Rodriguez, Eduardo
    Fernandez-Montes, Manuel
    Perez-Macho, Mercedes
    Ferri, Maria
    Garcia, Ana
    Cardenal, Juan
    Pacheco, Maria T.
    Anchuelo, Javier
    Tornero-Lopez, Ana M.
    Prada, Pedro J.
    Guirado, Damian
    MEDICAL PHYSICS, 2020, 47 (05) : 2242 - 2253