Calculation of electric fields induced by body and head motion in high-field MRI

被引:53
|
作者
Liu, F [1 ]
Zhao, HW [1 ]
Crozier, S [1 ]
机构
[1] Univ Queensland, Ctr Magnet Resonance, Sch Informat Technol & Elect Engn, Brisbane, Qld 4072, Australia
关键词
MRI; high magnetic field; induced current; human body model;
D O I
10.1016/S1090-7807(02)00180-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In modern magnetic resonance imaging (MRI), patients are exposed to strong, nonuniform static magnetic fields outside the central imaging region, in which the movement of the body may be able to induce electric currents in tissues which could be possibly harmful. This paper presents theoretical investigations into the spatial distribution of induced electric fields and currents in the patient when moving into the MRI scanner and also for head motion at various positions in the magnet. The numerical calculations are based on an efficient, quasi-static, finite-difference scheme and an anatomically realistic, full-body, male model. 3D field profiles from an actively shielded 4T magnet system are used and the body model projected through the field profile with a range of velocities. The simulation shows that it possible to induce electric fields/currents near the level of physiological significance under some circumstances and provides insight into the spatial characteristics of the induced fields. The results are extrapolated to very high field strengths and tabulated data shows the expected induced currents and fields with both movement velocity and field strength. (C) 2003 Elsevier Science (USA). All rights reserved.
引用
收藏
页码:99 / 107
页数:9
相关论文
共 50 条
  • [1] Numerical evaluation of the fields induced by body motion in or near high-field MRI scanners
    Crozier, S
    Liu, F
    PROGRESS IN BIOPHYSICS & MOLECULAR BIOLOGY, 2005, 87 (2-3): : 267 - 278
  • [2] Numerical evaluation of E-fields induced by body motion near high-field MRI scanner
    Crozier, S
    Liu, F
    PROCEEDINGS OF THE 26TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2004, 26 : 1120 - 1123
  • [3] High-field MRI
    Cunningham, Patricia M.
    Law, Meng
    Schweitzer, Mark E.
    ORTHOPEDIC CLINICS OF NORTH AMERICA, 2006, 37 (03) : 321 - +
  • [4] Efficient high-frequency body coil for high-field MRI
    Vaughan, JT
    Adriany, G
    Snyder, CJ
    Tian, J
    Thiel, T
    Bolinger, L
    Liu, H
    DelaBarre, L
    Ugurbil, K
    MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (04) : 851 - 859
  • [5] Exploring the limits of RF shimming for high-field MRI of the human head
    Mao, Weihua
    Smith, Michael B.
    Collins, Christopher M.
    MAGNETIC RESONANCE IN MEDICINE, 2006, 56 (04) : 918 - 922
  • [6] Advances in High-Field MRI
    Hespel, Adrien-Maxence
    Cole, Robert C.
    VETERINARY CLINICS OF NORTH AMERICA-SMALL ANIMAL PRACTICE, 2018, 48 (01) : 11 - +
  • [7] Prospect of High-Field MRI
    Wada, Hitoshi
    Sekino, Masaki
    Ohsaki, Hiroyuki
    Hisatsune, Tatsuhiro
    Ikehira, Hiroo
    Kiyoshi, Tsukasa
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2010, 20 (03) : 115 - 122
  • [8] The Safety of High-Field MRI?
    Yilmaz, S.
    INTERNATIONAL JOURNAL OF OCCUPATIONAL AND ENVIRONMENTAL MEDICINE, 2014, 5 (02): : 63 - 64
  • [9] High-field electron transport in GaN under crossed electric and magnetic fields
    Kochelap, V. A.
    Korotyeyev, V. V.
    Syngayivska, G. I.
    Varani, L.
    19TH INTERNATIONAL CONFERENCE ON ELECTRON DYNAMICS IN SEMICONDUCTORS, OPTOELECTRONICS AND NANOSTRUCTURES (EDISON' 19), 2015, 647
  • [10] High-field miniband transport in semiconductor superlattices in parallel electric and magnetic fields
    Kleinert, P
    Bryksin, VV
    PHYSICAL REVIEW B, 1997, 56 (24) : 15827 - 15835