Transition From Manual to Automated Planning and Delivery of Volumetric Modulated Arc Therapy Stereotactic Radiosurgery: Clinical, Dosimetric, and Quality Assurance Results

被引:19
|
作者
Popple, Richard A. [1 ]
Brown, Matthew H. [1 ]
Thomas, Evan M. [1 ]
Willey, Christopher D. [1 ]
Cardan, Rex A. [1 ]
Covington, Elizabeth L. [1 ]
Riley, Kristen O. [2 ]
Markert, James M. [2 ]
Bredel, Markus [1 ]
Fiveash, John B. [1 ]
机构
[1] Univ Alabama Birmingham, Dept Radiat Oncol, Birmingham, AL 35294 USA
[2] Univ Alabama Birmingham, Dept Neurosurg, Birmingham, AL USA
基金
美国国家卫生研究院;
关键词
D O I
10.1016/j.prro.2020.10.013
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Properly planned single isocenter volumetric modulated arc therapy (VMAT) radiosurgery plans exhibit high quality and efficiency. We report here the largest clinical experience to date, to our knowledge, comparing manual planning with a new automated platform designed to standardize and simplify radiosurgery planning and delivery processes. Methods: We treated 693 patients with single isocenter VMAT radiosurgical plans generated by either our conventional manual (mVMAT) or a recently implemented automated (HyperArc (TM)) technique. All plans targeted the gross tumor volume without margin. Radiochromic film was used for patient-specific quality assurance (PSQA). We evaluated local control and toxicity data for a subgroup of 107 patients having 377 metastatic tumors that were treated with HyperArc. Results: The median Radiation Therapy Oncology Group (RTOG) conformity index was 1.14 and was not different between the 2 techniques. The median Paddick gradient index was 5.42 for HyperArc versus 7.09 for mVMAT (P < .001). The median mean brain doses were 4.6% and 5.1% for HyperArc and mVMAT, respectively (P = .04). The PSQA for both techniques met clinical criteria, but 97% of the HyperArc plans satisfied the gamma tolerance limit recommended by the American Association of Physicists in Medicine Task Group No. 218, compared with 94% of the mVMAT plans (P = .02). The median treatment-planning times were not significantly different. The median treatment times were 10.5 and 11.4 minutes for HyperArc and mVMAT, respectively (P < .001). The Kaplan-Meier estimate of local control was 90.1% at 1 year. Conclusions: HyperArc produces high-quality radiosurgical plans that are at least as good as mVMAT plans created by an expert manual planner with easier planning and more efficient delivery workflow. A less experienced planner can produce very high-quality radiosurgical plans even for patients with more than 10 targets. The use of a single-isocenter technique for multiple targets with no PTV margin did not compromise clinical outcomes, and 1-year local control for treated targets remained congruent with historical series. (C) 2020 The Authors. Published by Elsevier Inc. on behalf of American Society for Radiation Oncology.
引用
收藏
页码:E163 / E171
页数:9
相关论文
共 50 条
  • [1] Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy
    Letourneau, Daniel
    Publicover, Julia
    Kozelka, Jakub
    Moseley, Douglas J.
    Jaffray, David A.
    MEDICAL PHYSICS, 2009, 36 (05) : 1813 - 1821
  • [2] Geometric and Dosimetric Quality Assurance in Volumetric Modulated Arc Therapy (VMAT)
    Qian, J.
    Ma, Y.
    Lee, L.
    Xing, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S639 - S639
  • [3] Interinstitutional Dosimetric Comparison Between Stereotactic Radiosurgery and Volumetric Modulated Arc Therapy Radiosurgery for Difficult Intracranial Cases
    Thomas, E. M.
    Chao, S. T.
    Markert, J. M.
    Neyman, G.
    Popple, R. A.
    Barnett, G. H.
    Fiveash, J. B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E693 - E693
  • [4] Dosimetric Effect of Plan Homogeneity On Intracranial Stereotactic Radiosurgery (SRS) Treatment Planning Using Volumetric Modulated Arc Therapy (VMAT)
    Tseng, T.
    Sheu, R.
    Green, S.
    Dharmarajan, K.
    Sharma, S.
    Lo, Y.
    MEDICAL PHYSICS, 2018, 45 (06) : E334 - E334
  • [5] Quality assurance of volumetric modulated arc therapy: Evaluation and comparison of different dosimetric systems
    Masi, L.
    Casamassima, F.
    Doro, R.
    Francescon, P.
    MEDICAL PHYSICS, 2011, 38 (02) : 612 - 621
  • [6] An Automated Approach to High-Quality Single-Isocenter Volumetric Modulated Arc Therapy Stereotactic Radiosurgery Treatment Planning on a Linear Accelerator
    Thomas, E. M.
    Ollila, S. T. T.
    Popple, R. A.
    Fiveash, J. B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E618 - E619
  • [7] Volumetric Modulated Arc Therapy for Delivery of Stereotactic Lung Radiotherapy: A Dosimetric and Treatment Efficiency Analysis
    McGrath, S. D.
    Matuszak, M.
    Yan, D.
    Kestin, L.
    Martinez, A.
    Grills, I.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S683 - S683
  • [8] Comparison of volumetric-modulated arc therapy and dynamic conformal arc treatment planning for cranial stereotactic radiosurgery
    Molinier, Jessica
    Kerr, Christine
    Simeon, Sebastien
    Ailleres, Norbert
    Charissoux, Marie
    Azria, David
    Fenoglietto, Pascal
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (01): : 92 - 101
  • [9] Dosimetric and radiobiological evaluation of stereotactic radiosurgery using volumetric modulated arc therapy and dynamic conformal arc therapy for multiple brain metastases
    Jiang, Chen-Jun
    Ho, Yick-Wing
    Lok, Ka-Hei
    Lu, Yeow-Yuen
    Zhu, Chun-Ran
    Cheng, Harry Chi-Yuen
    SCIENTIFIC REPORTS, 2025, 15 (01):
  • [10] Clinical experience with planning, quality assurance, and delivery of burst-mode modulated arc therapy
    Kainz, Kristofer
    Prah, Douglas
    Ahunbay, Ergun
    Li, X. Allen
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (05): : 47 - 59