Free-breathing high-pitch 80 kVp dual-source computed tomography of the pediatric chest: Image quality, presence of motion artifacts and radiation dose

被引:22
|
作者
Bodelle, Boris [1 ]
Fischbach, Constanze [1 ]
Booz, Christian [1 ]
Yel, Ibrahim [1 ]
Frellesen, Claudia [1 ]
Beeres, Martin [1 ]
Vogl, Thomas J. [1 ]
Scholtz, Jan-Erik [1 ]
机构
[1] Goethe Univ Frankfurt, Dept Diagnost & Intervent Radiol, Theodor Stern Kai 7, D-60590 Frankfurt, Germany
关键词
High pitch CT; Third generation dual-source CT; Free breathing; Pediatric CT; Chest CT; FILTERED BACK-PROJECTION; AFFIRMED ITERATIVE RECONSTRUCTION; ADVERSE EVENTS; CT; SEDATION; CHILDREN; CANCER; RISKS;
D O I
10.1016/j.ejrad.2017.01.027
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: To investigate image quality, presence of motion artifacts and effects on radiation dose of 80 kVp high-pitch dual-source CT (DSCT) in combination with an advanced modeled iterative reconstruction algorithm (ADMIRE) of the pediatric chest compared to single-source CT (SSCT). Materials and methods: The study was approved by the institutional review board. Eighty-seven consecutive pediatric patients (mean age 9.1 +/- 4.9 years) received either free-breathing high-pitch (pitch 3.2) chest 192-slice DSCT (group 1, n = 31) or standard-pitch (pitch 1.2) 128-slice SSCT (group 2, n = 56) with breathing-instructions by random assignment. Tube settings were similar in both groups with 80 kVp and 74 ref. mAs. Images were reconstructed using FBP for both groups. Additionally, ADMIRE was used in group 1. Effective thorax diameter, image noise, and signal-to-noise ratio (SNR) of the pectoralis major muscle and the thoracic aorta were calculated. Motion artifacts were measured as doubling boarders of the diaphragm and the heart. Images were rated by two blinded readers for overall image quality and presence of motion artifacts on 5-point-scales. Size specific dose estimates (SSDE, mGy) and effective dose (ED, mSv) were calculated. Results: Age and effective thorax diameter showed no statistically significant differences in both groups. Image noise and SNR were comparable (p > 0.64) for SSCT and DSCT with ADMIRE, while DSCT with FBP showed inferior results (p < 0.01). Motion artifacts were reduced significantly (p = 0.001) with DSCT. DSCT with ADMIRE showed the highest overall IQ (p < 0.0001). Radiation dose was lower for DSCT compared to SSCT (median SSDE: 0.82 mGy vs. 0.92 mGy, p < 0.02; median ED: 0.4 mSv vs. 0.48 mSv, p = 0.02). Conclusions: High-pitch 80 kVp chest DSCT in combination with ADMIRE reduces motion artifacts and increases image quality while lowering radiation exposure in free-breathing pediatric patients without sedation. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:208 / 214
页数:7
相关论文
共 50 条
  • [41] The ultrafast, high-pitch turbo FLASH mode of third-generation dual-source CT: Effect of different pitch and corresponding SFOV on image quality in a phantom study
    Zhou, Yang
    Hu, Lei
    Du, Silin
    Jin, Rui
    Li, Wangjia
    Lv, Fajin
    Zhang, Zhiwei
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (12): : 158 - 167
  • [42] Adenosine Stress High-Pitch 128-Slice Dual-Source Myocardial Computed Tomography Perfusion for Imaging of Reversible Myocardial Ischemia Comparison With Magnetic Resonance Imaging
    Feuchtner, Gudrun
    Goetti, Robert
    Plass, Andre
    Wieser, Monika
    Scheffel, Hans
    Wyss, Christophe
    Stolzmann, Paul
    Donati, Olivio
    Schnabl, Johannes
    Falk, Volkmar
    Alkadhi, Hatem
    Leschka, Sebastian
    Cury, Ricardo C.
    CIRCULATION-CARDIOVASCULAR IMAGING, 2011, 4 (05) : 540 - 549
  • [43] Diagnostic accuracy and effective radiation dose of high pitch dual source multidetector computed tomography in evaluation of coronary artery bypass graft patency
    Koplay, Mustafa
    Guneyli, Serkan
    Akbayrak, Hakan
    Demir, Kenan
    Sivri, Mesut
    Avci, Ahmet
    Erdogan, Hasan
    Paksoy, Yahya
    WIENER KLINISCHE WOCHENSCHRIFT, 2016, 128 (13-14) : 488 - 494
  • [44] Image quality and radiation dose of dual-source CT cardiac angiography using prospective ECG-triggering technique in pediatric patients with congenital heart disease
    Liu, Ying
    Li, Jian
    Zhao, Hongliang
    Jia, Yan
    Ren, Jing
    Xu, Jian
    Hao, Yuewen
    Zheng, Minwen
    JOURNAL OF CARDIOTHORACIC SURGERY, 2016, 11
  • [45] Comparison of Radiation Dose and Image Quality Between Split-Filter Twin Beam Dual-Energy Images and Single-Energy Images in Single-Source Contrast-Enhanced Chest Computed Tomography
    Zhou, Li-Na
    Zhao, Shi-Jun
    Wang, Ruo-Bing
    Wang, Ya-Wen
    Yang, Shou-Xin
    Wu, Ning
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2021, 45 (06) : 888 - 893
  • [46] Low-Concentration Contrast Medium for 128-Slice Dual-Source CT Coronary Angiography at a Very Low Radiation Dose Using Prospectively ECG-Triggered High-Pitch Spiral Acquisition
    Zheng, Minwen
    Wu, Yongjie
    Wei, Mengqi
    Liu, Ying
    Zhao, Hongliang
    Li, Jian
    ACADEMIC RADIOLOGY, 2015, 22 (02) : 195 - 202
  • [47] Prospective evaluation of ultra-low-dose contrast-enhanced 100-kV abdominal computed tomography with tin filter: effect on radiation dose reduction and image quality with a third-generation dual-source CT system
    Leyendecker, Pierre
    Faucher, Vanina
    Labani, Aissam
    Noblet, Vincent
    Lefebvre, Francois
    Magotteaux, Paul
    Ohana, Mickael
    Roy, Catherine
    EUROPEAN RADIOLOGY, 2019, 29 (04) : 2107 - 2116
  • [48] Initial Results of a Single-Source Dual-Energy Computed Tomography Technique Using a Split-Filter Assessment of Image Quality, Radiation Dose, and Accuracy of Dual-Energy Applications in an In Vitro and In Vivo Study
    Euler, Andre
    Parakh, Anushri
    Falkowski, Anna L.
    Manneck, Sebastian
    Dashti, David
    Krauss, Bernhard
    Szucs-Farkas, Zsolt
    Schindera, Sebastian T.
    INVESTIGATIVE RADIOLOGY, 2016, 51 (08) : 491 - 498
  • [49] Image Quality and Radiation Dose of Contrast-Enhanced Chest-CT Acquired on a Clinical Photon-Counting Detector CT vs. Second-Generation Dual-Source CT in an Oncologic Cohort: Preliminary Results
    Hagen, Florian
    Walder, Lukas
    Fritz, Jan
    Gutjahr, Ralf
    Schmidt, Bernhard
    Faby, Sebastian
    Bamberg, Fabian
    Schoenberg, Stefan
    Nikolaou, Konstantin
    Horger, Marius
    TOMOGRAPHY, 2022, 8 (03) : 1466 - 1476
  • [50] Comparison of 100-Kilovoltage Tin Filtration With Advanced Modeled Iterative Reconstruction Protocol to an Automated Kilovoltage Selection With Filtered Back Projection Protocol on Radiation Dose and Image Quality in Pediatric Noncontrast-Enhanced Chest Computed Tomography
    Siegel, Marilyn J.
    Raptis, Demetrios
    Bhalla, Sanjeev
    Ramirez-Giraldo, Juan Carlos
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2022, 46 (01) : 64 - 70