Free-breathing high-pitch 80 kVp dual-source computed tomography of the pediatric chest: Image quality, presence of motion artifacts and radiation dose

被引:24
作者
Bodelle, Boris [1 ]
Fischbach, Constanze [1 ]
Booz, Christian [1 ]
Yel, Ibrahim [1 ]
Frellesen, Claudia [1 ]
Beeres, Martin [1 ]
Vogl, Thomas J. [1 ]
Scholtz, Jan-Erik [1 ]
机构
[1] Goethe Univ Frankfurt, Dept Diagnost & Intervent Radiol, Theodor Stern Kai 7, D-60590 Frankfurt, Germany
关键词
High pitch CT; Third generation dual-source CT; Free breathing; Pediatric CT; Chest CT; FILTERED BACK-PROJECTION; AFFIRMED ITERATIVE RECONSTRUCTION; ADVERSE EVENTS; CT; SEDATION; CHILDREN; CANCER; RISKS;
D O I
10.1016/j.ejrad.2017.01.027
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: To investigate image quality, presence of motion artifacts and effects on radiation dose of 80 kVp high-pitch dual-source CT (DSCT) in combination with an advanced modeled iterative reconstruction algorithm (ADMIRE) of the pediatric chest compared to single-source CT (SSCT). Materials and methods: The study was approved by the institutional review board. Eighty-seven consecutive pediatric patients (mean age 9.1 +/- 4.9 years) received either free-breathing high-pitch (pitch 3.2) chest 192-slice DSCT (group 1, n = 31) or standard-pitch (pitch 1.2) 128-slice SSCT (group 2, n = 56) with breathing-instructions by random assignment. Tube settings were similar in both groups with 80 kVp and 74 ref. mAs. Images were reconstructed using FBP for both groups. Additionally, ADMIRE was used in group 1. Effective thorax diameter, image noise, and signal-to-noise ratio (SNR) of the pectoralis major muscle and the thoracic aorta were calculated. Motion artifacts were measured as doubling boarders of the diaphragm and the heart. Images were rated by two blinded readers for overall image quality and presence of motion artifacts on 5-point-scales. Size specific dose estimates (SSDE, mGy) and effective dose (ED, mSv) were calculated. Results: Age and effective thorax diameter showed no statistically significant differences in both groups. Image noise and SNR were comparable (p > 0.64) for SSCT and DSCT with ADMIRE, while DSCT with FBP showed inferior results (p < 0.01). Motion artifacts were reduced significantly (p = 0.001) with DSCT. DSCT with ADMIRE showed the highest overall IQ (p < 0.0001). Radiation dose was lower for DSCT compared to SSCT (median SSDE: 0.82 mGy vs. 0.92 mGy, p < 0.02; median ED: 0.4 mSv vs. 0.48 mSv, p = 0.02). Conclusions: High-pitch 80 kVp chest DSCT in combination with ADMIRE reduces motion artifacts and increases image quality while lowering radiation exposure in free-breathing pediatric patients without sedation. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:208 / 214
页数:7
相关论文
共 29 条
[1]  
American Association of Physicists in Medicine, 2011, AAPM REPORT 204 SIZE
[2]   Low-dose CT of the lung: potential value of iterative reconstructions [J].
Baumueller, Stephan ;
Winklehner, Anna ;
Karlo, Christoph ;
Goetti, Robert ;
Flohr, Thomas ;
Russi, Erich W. ;
Frauenfelder, Thomas ;
Alkadhi, Hatem .
EUROPEAN RADIOLOGY, 2012, 22 (12) :2597-2606
[3]   Computed Tomography of the Lung in the High-Pitch Mode Is Breath Holding Still Required? [J].
Baumueller, Stephan ;
Alkadhi, Hatem ;
Stolzmann, Paul ;
Frauenfelder, Thomas ;
Goetti, Robert ;
Schertler, Thomas ;
Plass, Andre ;
Falk, Volkmar ;
Feuchtner, Gudrun ;
Scheffel, Hans ;
Desbiolles, Lotus ;
Leschka, Sebastian .
INVESTIGATIVE RADIOLOGY, 2011, 46 (04) :240-245
[4]  
Bongartz G., 1999, REPORT EUR 16262
[5]   Estimated risks of radiation-induced fatal cancer from pediatric CT [J].
Brenner, DJ ;
Elliston, CD ;
Hall, EJ ;
Berdon, WE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :289-296
[6]   A COEFFICIENT OF AGREEMENT FOR NOMINAL SCALES [J].
COHEN, J .
EDUCATIONAL AND PSYCHOLOGICAL MEASUREMENT, 1960, 20 (01) :37-46
[7]   Chest wall trauma [J].
Collins, J .
JOURNAL OF THORACIC IMAGING, 2000, 15 (02) :112-119
[8]   Incidence and nature of adverse events during pediatric sedation/anesthesia for procedures outside the operating room: Report from the pediatric sedation research consortium [J].
Cravero, Joseph P. ;
Blike, George T. ;
Beach, Michael ;
Gallagher, Susan M. ;
Hertzog, James H. ;
Havidich, Jeana E. ;
Gelman, Barry .
PEDIATRICS, 2006, 118 (03) :1087-1096
[9]   Multisection CT Protocols: Sex- and Age-specific Conversion Factors Used to Determine Effective Dose from Dose-Length Product [J].
Deak, Paul D. ;
Smal, Yulia ;
Kalender, Willi A. .
RADIOLOGY, 2010, 257 (01) :158-166
[10]  
Hagelstein C, 2015, Z MED PHYS