Peripheral Nerve Diffusion Tensor Imaging Interreader and Test-retest Reliability as Quantified by the Standard Error of Measurement

被引:8
|
作者
Preisner, Fabian [1 ]
Baeumer, Philipp [1 ,2 ]
Wehrstein, Michaela [3 ]
Friedmann-Bette, Birgit [3 ]
Hackbusch, Matthes [4 ]
Heiland, Sabine [1 ]
Bendszus, Martin [1 ]
Kronlage, Moritz [1 ]
机构
[1] Heidelberg Univ Hosp, Dept Neuroradiol, Neuenheimer Feld 400, D-69120 Heidelberg, Germany
[2] Ctr Radiol Dialog, Vinzenz von Paul Str 8, D-84503 Altotting, Germany
[3] Heidelberg Univ Hosp, Dept Sports Med Internal Med 7, Med Clin, Neuenheimer Feld 410, D-69120 Heidelberg, Germany
[4] Heidelberg Univ, Inst Med Biometry & Informat, Neuenheimer Feld 130-3, D-69120 Heidelberg, Germany
关键词
Diffusion tensor imaging; Magnetic resonance imaging; Peripheral nervous system; Reference values; Reproducibility of results; INTRACLASS CORRELATION-COEFFICIENTS; MAGNETIC-RESONANCE NEUROGRAPHY; MEDIAN NERVE; FIBER TRACTOGRAPHY; MR NEUROGRAPHY; FRACTIONAL ANISOTROPY; DIAGNOSTIC-ACCURACY; TUNNEL; NEUROPATHY; AGREEMENT;
D O I
10.1007/s00062-019-00859-0
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Purpose Diffusion tensor imaging (DTI) is increasingly being used in magnetic resonance neurography (MRN). The purpose of this study was to determine the interreader and test-retest reliability of peripheral nerve DTI in MRN with focus on the sciatic nerve. Methods In this prospective study 27 healthy volunteers each underwent 3 scans of a short DTI protocol on separate days consisting of a T2-weighted turbo spin-echo and single-shot DTI sequence of the sciatic nerve of the dominant leg. The DTI parameters fractional anisotropy (FA), mean diffusivity (MD), axial diffusivity (AD), and radial diffusivity (RD) were obtained after manual nerve segmentation by two independent readers. Intraclass correlation coefficients (ICC), standard error of measurement (SEM), and Bland-Altman plots were calculated as measures for both interreader and test-retest agreement for all readout parameters. Results The mean +/- standard deviation was 0.507 +/- 0.05 for FA, 1308.5 +/- 162.4x 10(-6)mm(2)/s for MD, 905.6 +/- 145.4x10(-6)mm(2)/s for RD and 2114.1 +/- 219.2x 10(-6)mm(2)/s for AD. The SEM for FA was 0.02 for interreader and test-retest agreement, the SEM for MD, RD, and AD ranged between 46.2x 10(-6)mm(2)/s (RD) and 70.1x 10(-6)mm(2)/s (AD) for interreader reliability and between 45.9x 10(-6)mm(2)/s (RD) and 70.1x 10(-6)mm(2)/s (AD) for test-retest reliability. The ICC for interreader reliability of DTI parameters ranged between 0.81 and 0.92 and ICC for test-retest reliability between 0.76 and 0.91. Conclusion Peripheral nerve DTI of the sciatic nerve is reliable and reproducible. The measures presented here may serve as first orientation values of measurement accuracy when interpreting parameters of sciatic nerve DTI.
引用
收藏
页码:679 / 689
页数:11
相关论文
共 50 条
  • [1] Peripheral Nerve Diffusion Tensor ImagingInterreader and Test-retest Reliability as Quantified by the Standard Error of Measurement
    Fabian Preisner
    Philipp Bäumer
    Michaela Wehrstein
    Birgit Friedmann-Bette
    Matthes Hackbusch
    Sabine Heiland
    Martin Bendszus
    Moritz Kronlage
    Clinical Neuroradiology, 2020, 30 : 679 - 689
  • [2] Test-retest reliability of diffusion tensor imaging of the liver at 3.0 T
    Girometti, Rossano
    Maieron, Marta
    Lissandrello, Giovanni
    Bazzocchi, Massimo
    Zuiani, Chiara
    RADIOLOGIA MEDICA, 2015, 120 (06): : 489 - 497
  • [3] Test-retest reliability of Diffusion Tensor Imaging metrics in neonates
    Merisaari, Harri
    Tuulari, Jetro J.
    Karlsson, Linnea
    Scheinin, Noora M.
    Parkkola, Riitta
    Saunavaara, Jani
    Lahdesmaki, Tuire
    Lehtola, Satu J.
    Keskinen, Maria
    Lewis, John D.
    Evans, Alan C.
    Karlsson, Hasse
    NEUROIMAGE, 2019, 197 : 598 - 607
  • [4] Peripheral nerve diffusion tensor imaging (DTI): normal values and demographic determinants in a cohort of 60 healthy individuals
    Kronlage, Moritz
    Schwehr, Veronique
    Schwarz, Daniel
    Godel, Tim
    Uhlmann, Lorenz
    Heiland, Sabine
    Bendszus, Martin
    Baeumer, Philipp
    EUROPEAN RADIOLOGY, 2018, 28 (05) : 1801 - 1808
  • [5] Test-retest reliability of high spatial resolution diffusion tensor and diffusion kurtosis imaging
    Shahim, Pashtun
    Holleran, Laurena
    Kim, Joong H.
    Brody, David L.
    SCIENTIFIC REPORTS, 2017, 7
  • [6] Test–retest reliability of diffusion tensor imaging of the liver at 3.0 T
    Rossano Girometti
    Marta Maieron
    Giovanni Lissandrello
    Massimo Bazzocchi
    Chiara Zuiani
    La radiologia medica, 2015, 120 : 489 - 497
  • [7] Test-retest reliability and repeatability of renal diffusion tensor MRI in healthy subjects
    Cutajar, Marica
    Clayden, Jonathan D.
    Clark, Christopher A.
    Gordon, Isky
    EUROPEAN JOURNAL OF RADIOLOGY, 2011, 80 (03) : E263 - E268
  • [8] Test-Retest reliability and measurement error of the uncontrolled manifold analysis: A step towards the clinical translation
    Piscitelli, Daniele
    Buttram, Adrien
    Gibson, Stephanie
    Hager, Joel
    Thomas, Ben
    Solnik, Stanislaw
    JOURNAL OF BIOMECHANICS, 2024, 162
  • [9] Peripheral nerve diffusion tensor imaging: Overview, pitfalls, and future directions
    Jeon, Tina
    Fung, Maggie M.
    Koch, Kevin M.
    Tan, Ek T.
    Sneag, Darryl B.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2018, 47 (05) : 1171 - 1189
  • [10] Stimulated echo acquisition mode (STEAM) diffusion tensor imaging with different diffusion encoding times in the supraspinatus muscle: Test-retest reliability and comparison to spin echo diffusion tensor imaging
    Marth, Adrian Alexander
    Sommer, Stefan
    Feiweier, Thorsten
    Sutter, Reto
    Nanz, Daniel
    von Deuster, Constantin
    NMR IN BIOMEDICINE, 2025, 38 (01)