Topology optimization of the permanent magnet type MRI considering the magnetic field homogeneity

被引:8
作者
Lee, Junghoon [2 ]
Yoo, Jeonghoon [1 ]
机构
[1] Yonsei Univ, Sch Mech Engn, Seoul 120749, South Korea
[2] Yonsei Univ, Grad Sch Mech Engn, Seoul 120749, South Korea
关键词
Magnetic resonance imaging; Permanent magnet; Magnetic field; Eddy current; Topology optimization; Density method;
D O I
10.1016/j.jmmm.2009.04.078
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This study is to suggest a concept design of the permanent magnet (PM) type magnetic resonance imaging (MRI) device based on the topology optimization method. The pulse currents in the gradient coils in the MRI device will introduce the effect of eddy currents in ferromagnetic material and it may worsen the quality of imaging. To equalize the magnetic flux in the PM type MRI device for good imaging, the eddy current effect in the ferromagnetic material must be reduced. This study attempts to use the topology optimization scheme for equalizing the magnetic flux in the measuring domain of the PM type MRI device using that the magnetic flux can be calculated directly by a commercial finite element analysis package. The density method is adopted for topology optimization and the sensitivity of the objective function is computed according to the density change of each finite element in the design domain. As a result, optimal shapes of the pole of the PM type MRI device can be obtained. The commercial package, ANSYS, is used for analyzing the magnetic field problem and obtaining the resultant magnetic flux. (C) 2009 Elsevier B. V. All rights reserved.
引用
收藏
页码:1651 / 1654
页数:4