Three-Dimensional Imaging Properties of Rotation-Free Square and Hexagonal Micro-CT Systems

被引:17
作者
Quan, Enzhuo Michelle [1 ,2 ]
Lalush, David S. [1 ,2 ]
机构
[1] N Carolina State Univ, Dept Biomed Engn, Raleigh, NC 27695 USA
[2] Univ N Carolina Chapel Hill, Raleigh, NC 27695 USA
基金
美国国家卫生研究院;
关键词
Carbon-nanotube; image reconstruction; micro-computed tomography (CT); X-ray computed tomography; X-RAY SOURCE; COMPUTED-TOMOGRAPHY; RECONSTRUCTION; PHANTOM;
D O I
10.1109/TMI.2009.2039799
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We study the 3-D imaging properties of a rotation-free micro-computed tomography (CT) system with square and hexagonal geometries. These systems use linear arrays of carbon-nanotube-based X-ray sources that are individually addressable. The source arrays and area detectors in the square and the hexagonal geometries form the sides of a polygon. The tomographic angular sampling for both geometries requires no motion of the sources, the detectors, or the subject. We demonstrate that the hexagonal geometry has improved angular coverage as compared to the square geometry. The ordered-subset convex iterative algorithm is implemented in both geometries for reconstructions from cone-beam projection data. Simulation studies show that both geometries can be effectively reconstructed with polychromatic or monochromatic source spectra. As a result of the incomplete tomographic sampling of the two geometries, some streaking artifacts appear in the reconstructed images. The hexagonal geometry, in general, produces fewer streaking artifacts than the square geometry. On the other hand, the two geometries perform quite similarly in resolution-noise trade-off, so we conclude that the proposed geometries are comparably effective for the rotation-free micro-CT and the hexagonal geometry is superior in reducing streaking artifacts.
引用
收藏
页码:916 / 923
页数:8
相关论文
共 24 条
[1]   Performance of cooperative time-reversal communication in a mobile wireless environment [J].
Barton, Richard J. ;
Chen, Ji ;
Huang, Kyle ;
Wu, Dagang ;
Wu, Hsiao-Chun .
INTERNATIONAL JOURNAL OF DISTRIBUTED SENSOR NETWORKS, 2007, 3 (01) :59-68
[2]   Dynamic radiography using a carbon-nanotube-based field-emission x-ray source [J].
Cheng, Y ;
Zhang, J ;
Lee, YZ ;
Gao, B ;
Dike, S ;
Lin, W ;
Lu, JP ;
Zhou, O .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (10) :3264-3267
[3]  
Cranley K., 1997, CATALOGUE DIAGNOSTIC
[4]   Theoretical hardness of PtN2 with pyrite structure [J].
Gou, Huiyang ;
Hou, Li ;
Zhang, Jingwu ;
Sun, Guifang ;
Gao, Lihua ;
Gao, Faming .
APPLIED PHYSICS LETTERS, 2006, 89 (14)
[5]  
Han GP, 1999, IEEE NUCL SCI CONF R, P1515, DOI 10.1109/NSSMIC.1999.842846
[6]   Accelerated iterative transmission CT reconstruction using an ordered subsets convex algorithm [J].
Kamphuis, C ;
Beekman, FJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (06) :1101-1105
[7]   Development of microfocused x-ray source by using carbon nanotube field emitter [J].
Kawakita, K ;
Hata, K ;
Sato, H ;
Saito, Y .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 2006, 24 (02) :950-952
[8]  
Lalush D, 2006, IEEE NUCL SCI CONF R, P1670
[9]   Binary encoding of multiplexed images in mixed noise [J].
Lalush, David S. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2008, 27 (09) :1323-1332
[10]  
Lalush DS, 2006, I S BIOMED IMAGING, P1180