The relevance of hydroxyapatite and spongious titanium coatings in fixation of cementless stems - An experimental comparative study in rat femur employing histological and microangiographic techniques

被引:17
作者
Pazzaglia, UE [1 ]
Brossa, F
Zatti, G
Chiesa, R
Andrini, L
机构
[1] Univ Brescia, Spedali Civili, Clin Ortoped, I-25100 Brescia, Italy
[2] JRC, Inst Adv Mat, Ispra, Italy
[3] Univ Pavia, Fac Med & Chirurg 2, I-27100 Pavia, Italy
关键词
D O I
10.1007/s004020050247
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Pure titanium rods plasma-spray coated with hydroxyapatite (HA) or porous titanium (Ti) of controlled roughness were implanted bilaterally in the distal femur of Sprague-Dawley rats to compare the extent of bone growth on the two types of coating. The relevance of other factors, like mechanical stability and biological adaptation of the bone to the insertion of a foreign body implant, were investigated in femora which were over-reamed (absence of primary fit) or reamed without insertion of the rod. Continuous tetracycline labeling for the first 30 days and for the last 2 weeks in the 90-day group was performed; histological/histometric, fluorescence and microangiographic studies were carried out on serial sections of the implanted and control femora. In the group of stable implants, HA-coated rods showed 90% integration versus 53% With Ti-coated implants (P < 0.001); in over-reamed implants neither surface bone growth nor endosteal fixation occurred, and both types of rods were surrounded by a thick layer of connective tissue. The study documented early adhesion of osteoblasts and direct deposition of bone matrix on the substrate, while on spongious titanium osteogenesis was observed only in proximity to the surface. Remodeling of the reactive, primary bone to mature, lamellar bone took the form of a capsule surrounding the implants and radial bridges connecting the latter to the endosteal surface. The number, height and thickness of these bridges appeared to be the factors determining implant stability, rather than the extent of the bony capsule on the perimeter of the implant. Integration was a function not only of mechanical conditions and surface geometry, but also of the biological response of the whole bone to changes in the vascularization pattern. The reported phenomena can be seen more easily in experimental models involving small rodents because of their fast bone turnover and revascularization. but it is expected that they take place, even at a lower speed, in clinical situations like cementless stems of total hip replacement.
引用
收藏
页码:279 / 285
页数:7
相关论文
共 34 条
[1]   MINERAL APPOSITION RATES OF HUMAN CANCELLOUS BONE AT THE INTERFACE OF POROUS-COATED IMPLANTS [J].
BLOEBAUM, RD ;
BACHUS, KN ;
MOMBERGER, NG ;
HOFMANN, AA .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1994, 28 (05) :537-544
[2]  
BLOEBAUM RD, 1993, J ARTHROPLASTY, V8, P203, DOI 10.1016/S0883-5403(09)80014-4
[3]  
BOBYN JD, 1980, CLIN ORTHOP RELAT R, P263
[4]  
Brossa F., 1993, Bio-Medical Materials and Engineering, V3, P127
[5]   POSTDEPOSITION TREATMENT EFFECTS ON HYDROXYAPATITE VACUUM PLASMA SPRAY COATINGS [J].
BROSSA, F ;
CIGADA, A ;
CHIESA, R ;
PARACCHINI, L ;
CONSONNI, C .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1994, 5 (12) :855-857
[6]   EFFECT OF MOVEMENT ON BONDING OF POROUS METAL TO BONE [J].
CAMERON, HU ;
PILLIAR, RM ;
MACNAB, I .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1973, 7 (04) :301-311
[7]  
CAMERON HU, 1976, J BIOMED MATER RES, V10, P295
[8]   THE EFFECT OF OPERATIVE FIT AND HYDROXYAPATITE COATING ON THE MECHANICAL AND BIOLOGICAL RESPONSE TO POROUS IMPLANTS [J].
DALTON, JE ;
COOK, SD ;
THOMAS, KA ;
KAY, JF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (01) :97-110
[9]  
DAWKWARDTLILLIE.G, 1970, J BONE JOINT SURG, V52, P1390
[10]  
DAWKWARDTLILLIE.G, 1969, ACTA ORTHOP SCAND, P128