Cross-calibration of a GE iDXA and Prodigy for Total and Regional Body Bone Parameters: The Importance of Using Cross-calibration Equations for Longitudinal Monitoring After a System Upgrade

被引:9
|
作者
Rhodes, Laura A. [1 ]
Cooper, William [1 ]
Oldroyd, Brian [1 ]
Hind, Karen [2 ]
机构
[1] Univ Leeds, Div Med Phys, Leeds LS2 9NL, W Yorkshire, England
[2] Leeds Metropolitan Univ, Carnegie Res Inst, Leeds LS1 3HE, W Yorkshire, England
关键词
Cross-calibration; iDXA; prodigy; regional total body; X-RAY ABSORPTIOMETRY; MINERAL DENSITY; PRECISION;
D O I
10.1016/j.jocd.2013.09.009
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We aimed to determine if cross-calibration equations could be applied to convert GE Lunar Prodigy total and regional bone measurements to the GE iDXA model to support longitudinal monitoring of subjects. The cross-calibration group comprised 63 adults (age 45.1 [12.8] yr; body mass index: 25.6 [3.7] kg/m(2)) and the validation group comprised 25 adults (age 40.5 [11.5] yr; body mass index: 25.7 [3.5] kg/m(2)). The parameters reported were total and regional bone mineral density (BMD), bone mineral content, and bone area. There were significant differences between densitometers for all anatomical regions and reported bone parameters (p < 0.0001); iDXA reported lower BMD than the Prodigy apart from the ribs. Linear regression indicated good agreement for all measurements. Bland-Altman analyses indicated significant bias for all measurements and that cross-calibration equations were required. The derived cross-calibration equations were effective in reducing differences between predicted and measured results for each parameter and at each region apart from leg BMD, where the difference remained significant (0.013 g/cm(2); p < 0.05). Our results indicate that cross-calibration is important to maintain comparability of total body-derived regional bone measurements between the Lunar Prodigy and iDXA.
引用
收藏
页码:496 / 504
页数:9
相关论文
共 18 条
  • [1] Inclusion of Regional Body Composition Parameters Improves Bone Mineral Density Cross-Calibration Between GE Lunar Prodigy and iDXA Densitometers
    Saarelainen, J.
    Hakulinen, M.
    Rikkonen, T.
    Kroger, H.
    Koivumaa-Honkanen, H.
    Honkanen, R.
    Jurvelin, J. S.
    JOURNAL OF CLINICAL DENSITOMETRY, 2017, 20 (01) : 97 - 105
  • [2] Cross-Calibration of iDXA and Prodigy on Spine and Femur Scans in Korean Adults
    Choi, Yong Jun
    Lee, Byung Joo
    Lim, Hyun Chae
    Chung, Yoon-Sok
    JOURNAL OF CLINICAL DENSITOMETRY, 2009, 12 (04) : 450 - 455
  • [3] Trabecular Bone Score (TBS) Cross-Calibration for GE Prodigy and IDXA Scanners
    Leslie, William D.
    Hans, Didier
    JOURNAL OF CLINICAL DENSITOMETRY, 2023, 26 (01) : 56 - 60
  • [4] A Cross-Calibration Study of the GE-Lunar iDXA and Prodigy for the Assessment of Lumbar Spine and Total Hip Bone Parameters via Three Statistical Methods
    Hind, Karen
    Cooper, William
    Oldroyd, Brian
    Davies, Andrew
    Rhodes, Laura
    JOURNAL OF CLINICAL DENSITOMETRY, 2015, 18 (01) : 86 - 92
  • [5] Cross Calibration of the GE Prodigy and iDXA for the Measurement of Total and Regional Body Composition in Adults
    Oldroy, Brian
    Treadgold, Laura
    Hind, Karen
    JOURNAL OF CLINICAL DENSITOMETRY, 2018, 21 (03) : 383 - 393
  • [6] A Cross-Calibration Study of GE Lunar iDXA and GE Lunar DPX Pro for Body Composition Measurements in Children and Adults
    Khadilkar, Anuradha
    Chiplonkar, Shashi
    Sanwalka, Neha
    Khadilkar, Vaman
    Mandlik, Rubina
    Ekbote, Veena
    JOURNAL OF CLINICAL DENSITOMETRY, 2020, 23 (01) : 128 - 137
  • [7] Pediatric in vivo cross-calibration between the GE Lunar Prodigy and DPX-L bone densitometers
    Nicola J. Crabtree
    N. J. Shaw
    C. M. Boivin
    B. Oldroyd
    J. G. Truscott
    Osteoporosis International, 2005, 16 : 2157 - 2167
  • [8] Pediatric in vivo cross-calibration between the GE Lunar Prodigy and DPX-L bone densitometers
    Crabtree, NJ
    Shaw, NJ
    Boivin, CM
    Oldroyd, B
    Truscott, JG
    OSTEOPOROSIS INTERNATIONAL, 2005, 16 (12) : 2157 - 2167
  • [9] iDXA, Prodigy, and DPXL Dual-Energy X-ray Absorptiometry Whole-Body Scans: A Cross-Calibration Study
    Hull, Holly
    He, Qing
    Thornton, John
    Javed, Fahad
    Allen, Lynn
    Wang, Jack
    Pierson, Richard N., Jr.
    Gallagher, Dympna
    JOURNAL OF CLINICAL DENSITOMETRY, 2009, 12 (01) : 95 - 102
  • [10] A cross-calibration study using a novel dual X-ray absorptiometry system for bone mineral density measurements with the European Spine Phantom
    Jiang, Yaojun
    Wu, Yan
    Zhang, Yonggao
    Yang, Xiaopeng
    Gao, Jianbo
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2023, 13 (04) : 2119 - 2127