作者
Martisikova, Maria [1 ]
Jaekel, Oliver [1 ,2 ]
机构
[1] German Canc Res Ctr, Dept Med Phys Radiat Oncol, D-69120 Heidelberg, Germany
[2] Univ Klinikum Heidelberg, HIT, D-69120 Heidelberg, Germany
关键词
PROTON; VERIFICATION; DEPENDENCE; PHOTON; SYSTEM; RANGE;
D O I
10.1088/0031-9155/55/13/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For quality assurance measurements in radiotherapy with scanned ion beams, such as beam profile measurements and dosimetrical verification of patient plans, high spatial accuracy is required. We investigated the Gafchromic EBT film, which is the ancestor of the currently available EBT2 film, for this purpose. The excellent spatial resolution and improvements in sensitivity as compared to older types of radiochromic films make it a suitable candidate. In this contribution a study of the dose-response relation for monoenergetic beams (plateau), based on the absolute dose, is presented. We found the film response to be comparable for protons and photons. For carbon ions we observed a decrease in the darkening efficiency of about 30% with respect to photons. This effect is likely to be caused by the ion track structure. We did not find any dependence of the relative efficiency on the applied dose and energy over typical ranges used in ion beam therapy. Furthermore, it is shown that overlaps of ion tracks are not an issue for EBT film dosimetry in medical ion beams. The above findings enable the use of EBT films calibrated in a beam of a particular ion type for dose measurements at any other dose and energy. The uncertainty of the film darkening determination in ion beams is also discussed.
引用
收藏
页码:3741 / 3751
页数:11
相关论文
共 27 条
[1]   Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies [J].
Arjomandy, Bijan ;
Tailor, Ramesh ;
Anand, Aman ;
Sahoo, Narayan ;
Gillin, Michael ;
Prado, Karl ;
Vicic, Milos .
MEDICAL PHYSICS, 2010, 37 (05) :1942-1947
[2]  
B_tter-Jensen L., 2003, Optically Stimulated Luminescence Dosimetry, V1st
[3]   THEORY OF RBE FOR HEAVY ION BOMBARDMENT OF DRY ENZYMES AND VIRUSES [J].
BUTTS, JJ ;
KATZ, R .
RADIATION RESEARCH, 1967, 30 (04) :855-&
[4]   Verification procedure for isocentric alignment of proton beams [J].
Ciangaru, George ;
Yang, James N. ;
Oliver, Patrick J. ;
Bues, Martin ;
Zhu, Mengping ;
Nakagawa, Fumio ;
Chiba, Hitoshi ;
Nakamura, Shin ;
Yoshino, Hirofumi ;
Umezawa, Mosumi ;
Smith, Alfred R. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2007, 8 (04) :65-75
[5]   Precise radiochromic film dosimetry using a flat-bed document scanner [J].
Devic, S ;
Seuntjens, J ;
Sham, E ;
Podgorsak, EB ;
Schmidtlein, CR ;
Kirov, AS ;
Soares, CG .
MEDICAL PHYSICS, 2005, 32 (07) :2245-2253
[6]   Optimizing the dynamic range extension of a radiochromic film dosimetry system [J].
Devic, Slobodan ;
Tomic, Nada ;
Soares, Christopher G. ;
Podgorsak, Ervin B. .
MEDICAL PHYSICS, 2009, 36 (02) :429-437
[7]   Clinical use of EBT model Gafchromic™ film in radiotherapy [J].
Fiandra, Christian ;
Ricardi, Umberto ;
Ragona, Riccardo ;
Anglesio, Silvia ;
Giglioli, Francesca Romana ;
Calamia, Elisa ;
Lucio, Francesco .
MEDICAL PHYSICS, 2006, 33 (11) :4314-4319
[8]   Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance [J].
Fuss, Martina ;
Sturtewagen, Eva ;
De Wagter, Carlos ;
Georg, Dietmar .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) :4211-4225
[9]   MAGNETIC SCANNING SYSTEM FOR HEAVY-ION THERAPY [J].
HABERER, T ;
BECHER, W ;
SCHARDT, D ;
KRAFT, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 330 (1-2) :296-305
[10]   Technical Note: Homogeneity of Gafchromic® EBT2 film [J].
Hartmann, Bernadette ;
Martisikova, Maria ;
Jaekel, Oliver .
MEDICAL PHYSICS, 2010, 37 (04) :1753-1756