Calibration of low-temperature ac susceptometers with a copper cylinder standard

被引:21
|
作者
Chen, D. -X. [1 ]
Skumryev, V.
机构
[1] Univ Autonoma Barcelona, ICREA, E-08193 Barcelona, Spain
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2010年 / 81卷 / 02期
关键词
calibration; copper; eddy currents; magnetic susceptibility; magnetic variables measurement; CRITICAL-CURRENT-DENSITY; HIGH-TC SUPERCONDUCTORS; CRITICAL-STATE SUSCEPTIBILITY; DEMAGNETIZING FACTORS; MAGNETIC-SUSCEPTIBILITY; CONDUCTING CYLINDERS; RECTANGULAR PRISMS; BARS; BEHAVIOR; GRAINS;
D O I
10.1063/1.3309779
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A high-quality low-temperature ac susceptometer is calibrated by comparing the measured ac susceptibility of a copper cylinder with its eddy-current ac susceptibility accurately calculated. Different from conventional calibration techniques that compare the measured results with the known property of a standard sample at certain fixed temperature T, field amplitude H-m, and frequency f, to get a magnitude correction factor, here, the electromagnetic properties of the copper cylinder are unknown and are determined during the calibration of the ac susceptometer in the entire T, H-m, and f range. It is shown that the maximum magnitude error and the maximum phase error of the susceptometer are less than 0.7% and 0.3 degrees, respectively, in the region T=5-300 K and f=111-1111 Hz at H-m=800 A/m, after a magnitude correction by a constant factor as done in a conventional calibration. However, the magnitude and phase errors can reach 2% and 4.3 degrees at 10 000 and 11 Hz, respectively. Since the errors are reproducible, a large portion of them may be further corrected after a calibration, the procedure for which is given. Conceptual discussions concerning the error sources, comparison with other calibration methods, and applications of ac susceptibility techniques are presented.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] ON THE LAYOUT OF LOW-TEMPERATURE BUSHINGS FOR HIGH AC VOLTAGE
    GERHOLD, J
    CRYOGENICS, 1981, 21 (07) : 426 - 430
  • [32] LOW-TEMPERATURE AC CONDUCTIVITY OF VANADIUM DIOXIDE CRYSTALS
    MANSINGH, A
    SINHA, AK
    DHEER, PN
    SAYER, M
    PHILOSOPHICAL MAGAZINE B-PHYSICS OF CONDENSED MATTER STATISTICAL MECHANICS ELECTRONIC OPTICAL AND MAGNETIC PROPERTIES, 1984, 50 (05): : 621 - 634
  • [33] Measuring Systems for Thermometer Calibration in Low-Temperature Range
    A. Szmyrka-Grzebyk
    L. Lipiński
    H. Manuszkiewicz
    A. Kowal
    A. Grykałowska
    D. Jancewicz
    International Journal of Thermophysics, 2011, 32 : 2466 - 2476
  • [34] Low-temperature radiator for verification and calibration of radiation thermometers
    Zuev, A. V.
    Rozhkov, V. A.
    MEASUREMENT TECHNIQUES, 2013, 56 (07) : 805 - 808
  • [35] Measuring Systems for Thermometer Calibration in Low-Temperature Range
    Szmyrka-Grzebyk, A.
    Lipinski, L.
    Manuszkiewicz, H.
    Kowal, A.
    Grykalowska, A.
    Jancewicz, D.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2011, 32 (11-12) : 2466 - 2476
  • [36] Comparison of Blackbody Sources for Low-Temperature IR Calibration
    Ljungblad, S.
    Holmsten, M.
    Josefson, L. E.
    Klason, P.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2015, 36 (12) : 3310 - 3319
  • [37] Low-temperature radiator for verification and calibration of radiation thermometers
    A. V. Zuev
    V. A. Rozhkov
    Measurement Techniques, 2013, 56 : 805 - 808
  • [38] Comparison of Blackbody Sources for Low-Temperature IR Calibration
    S. Ljungblad
    M. Holmsten
    L. E. Josefson
    P. Klason
    International Journal of Thermophysics, 2015, 36 : 3310 - 3319
  • [39] AUTOMATICALLY CONTROLLED LOW-TEMPERATURE SENSOR CALIBRATION FACILITY
    DEMATTIO, H
    RUSCH, KD
    SUSSER, M
    CRYOGENICS, 1994, 34 : 409 - 412
  • [40] LOW-TEMPERATURE STUDIES OF MUON LOCALIZATION IN COPPER
    HARTMANN, O
    NORLIN, LO
    YAOUANC, A
    LEHERICY, J
    KARLSSON, E
    NIINIKOSKI, TO
    HYPERFINE INTERACTIONS, 1981, 8 (4-6): : 533 - 537