A new method for measuring the ballistocardiogram using EMFi sensors in a normal chair

被引:0
作者
Koivistoinen, T [1 ]
Junnila, S [1 ]
Värri, A [1 ]
Kööbi, T [1 ]
机构
[1] Tampere Univ Hosp, Dept Clin Physiol, Tampere, Finland
来源
PROCEEDINGS OF THE 26TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7 | 2004年 / 26卷
关键词
ballistocardiogram; cardiac function; EMFi;
D O I
暂无
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Ballistocardiography is a non-invasive technique for the assessment of cardiac function. We built a measurement setup to measure the ballistocardiogram from a normal chair using EMFi sensors. The ballistocardiogram is recorded from a subject sitting on the chair. The measured signal is amplified by a specially-designed charge amplifier and digitized by a circulation monitor. A PC provides a user interface for the measurement devices, records the data and displays the results. Impedancecardiography and ECG serve as reference measurements for the ballistocardiography. To test the system, one healthy 24-year-old male and one healthy 22-year-old female were measured. It is concluded that the ballistocardiogram waveforms described in the literature can be recognized from the EMFi signal measured from a normal chair.