Effect of Tube Voltage on Image Quality in 64-Section Multidetector 3D CT Angiography: Evaluation with a Vascular Phantom with Superimposed Bone Skull Structures

被引:16
作者
Murakami, Y. [1 ]
Kakeda, S.
Kamada, K.
Ohnari, N.
Nishimura, J.
Ogawa, M.
Otsubo, K. [2 ]
Morishita, Y. [2 ]
Korogi, Y.
机构
[1] Univ Occupat & Environm Hlth, Dept Radiol, Sch Med, Yahatanishi Ku, Kitakyushu, Fukuoka 8078555, Japan
[2] Toshiba Co Ltd, Med Syst Div, Tokyo, Japan
关键词
DOSE REDUCTION; SUBARACHNOID HEMORRHAGE; INTRACRANIAL ANEURYSMS; COMPUTED-TOMOGRAPHY; 80; KVP; KILOVOLTAGE; SETTINGS; CIRCLE; WILLIS; HEAD;
D O I
10.3174/ajnr.A1871
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: Optimal tube voltage and tube current settings are not well established. The purpose of our study was to investigate the image quality on 3D CT angiograms of the brain at various kilovoltage settings by evaluating the depiction of simulated intracranial lesions by using a vascular phantom. MATERIALS AND METHODS: An anthropomorphic vascular phantom with superimposed bone skull structures was designed to simulate various intracranial aneurysms with aneurysmal blebs. We performed CT angiography by using a 64-detector row CT scanner for various effective tube currents with 4 tube voltages of 80, 100, 120, and 135 kV(p). Simulated aneurysm enhancement and image noise were quantified; SNR and CNR were calculated. The depiction of the simulated aneurysms and blebs on 3D CT angiograms obtained with the volume-rendering technique was subjectively assessed. The effective dose was calculated on the basis of a CTDIw. The results of several protocols were compared by using the Student t test. RESULTS: At identical doses levels (CTDIw), the mean SNR and CNR at 100 kV(p) were significantly higher than those at 80, 120, and 135 kV(p); and the mean qualitative image score at 100 kV(p) was significantly superior to those at 80 and 135 kV(p). CONCLUSIONS: Our phantom study suggests that the tube voltage of 100 kV(p) is desirable for cerebral 3D CT angiograms, and the higher or lower kilovoltage settings may result in the degradation in diagnostic image quality.
引用
收藏
页码:620 / 625
页数:6
相关论文
共 18 条
  • [1] CT angiography: In vitro comparison of five reconstruction methods
    Addis, KA
    Hopper, KD
    Iyriboz, TA
    Liu, Y
    Wise, SW
    Kasales, CJ
    Blebea, JS
    Mauger, DT
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (05) : 1171 - 1176
  • [2] Improved vascular opacification in cerebral computed tomography angiography with 80 kVp
    Bahner, ML
    Bengel, A
    Brix, G
    Zuna, I
    Kauczor, HU
    Delorme, S
    [J]. INVESTIGATIVE RADIOLOGY, 2005, 40 (04) : 229 - 234
  • [3] Radiation exposure in multi-slice versus single-slice spiral CT: results of a nationwide survey
    Brix, G
    Nagel, HD
    Stamm, G
    Veit, R
    Lechel, U
    Griebel, J
    Galanski, M
    [J]. EUROPEAN RADIOLOGY, 2003, 13 (08) : 1979 - 1991
  • [4] QUANTITATIVE THEORY OF HOUNSFIELD UNIT AND ITS APPLICATION TO DUAL ENERGY SCANNING
    BROOKS, RA
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1977, 1 (04) : 487 - 493
  • [5] Cohnen M, 2000, AM J NEURORADIOL, V21, P1654
  • [6] Multi-detector row CT angiography of the brain at various kilovoltage settings
    Ertl-Wagner, BB
    Hoffmann, RT
    Bruning, R
    Herrmann, K
    Snyder, B
    Blume, JD
    Reiser, MF
    [J]. RADIOLOGY, 2004, 231 (02) : 528 - 535
  • [7] Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions
    Greess, H
    Wolf, H
    Baum, U
    Lell, M
    Pirkl, M
    Kalender, W
    Bautz, WA
    [J]. EUROPEAN RADIOLOGY, 2000, 10 (02) : 391 - 394
  • [8] Pancreaticoduodenal arcades and dorsal pancreatic artery: Comparison of CT angiography with three-dimensional volume rendering, maximum intensity projection, and shaded-surface display
    Hong, KC
    Freeny, PC
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 172 (04) : 925 - 931
  • [9] Intracranial aneurysms: Detection with three-dimensional CT angiography with volume rendering - Comparison with conventional angiographic and surgical findings
    Korogi, Y
    Takahashi, M
    Katada, K
    Ogura, Y
    Hasuo, K
    Ochi, M
    Utsunomiya, H
    Abe, T
    Imakita, S
    [J]. RADIOLOGY, 1999, 211 (02) : 497 - 506
  • [10] Sixty-four-row multisection CT angiography for detection and evaluation of ruptured intracranial aneurysms: Interobserver and intertechnique reproducibility
    Lubicz, B.
    Levivier, M.
    Francois, O.
    Thoma, P.
    Sadeghi, N.
    Collignon, L.
    Baleriaux, D.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2007, 28 (10) : 1949 - 1955