摘要
Midogas(R) (CIG Australia) nitrous oxide/oxygen blenders are commonly used in delivery wards in Australia. In this ease report we describe an incident where a small hole in the diaphragm of the oxygen failure alarm in a Midogas nitrous oxide/oxygen blender led to retrograde flow of oxygen contaminating the nitrous oxide pipeline supply to the operating theatres and the delivery ward. This caused a reduced level of nitrous oxide to the patient in theatre, but there was no adverse outcome, However if the oxygen pipeline pressure had been lower than the nitrous oxide pipeline pressure, the contamination would have been reversed potentially resulting in a hypoxic mixture being delivered to many patients, Suggestions are made as to a method of prevention of this problem.
引用
收藏
页码:207 / 209
页数:3
相关论文
共 34 条
  • [21] Cp50 of propofol for laryngeal mask airway insertion using predicted concentrations with and without nitrous oxide
    Kodaka, M
    Handa, F
    Kawasaki, J
    Miyao, H
    ANAESTHESIA, 2002, 57 (10) : 956 - 959
  • [22] NITROUS-OXIDE INCREASES CEREBRAL BLOOD-FLOW VELOCITY DURING PHARMACOLOGICALLY INDUCED EEG SILENCE IN HUMANS
    MATTA, BF
    LAM, AM
    JOURNAL OF NEUROSURGICAL ANESTHESIOLOGY, 1995, 7 (02) : 89 - 93
  • [23] Interference of volatile anaesthetics with infrared analysis of carbon dioxide and nitrous oxide tested in the Drager Cicero EM using sevoflurane
    Wilkes, AR
    Mapleson, WW
    BRITISH JOURNAL OF ANAESTHESIA, 1996, 76 (05) : 737 - 739
  • [24] Evaluation of pressure changes in a new design tracheal tube cuff, the Porter Soft Seal, during nitrous oxide anaesthesia
    Al-Shaikh, B
    Jones, M
    Baldwin, F
    BRITISH JOURNAL OF ANAESTHESIA, 1999, 83 (05) : 805 - 806
  • [25] Waste gas exposure to sevoflurane and nitrous oxide during anaesthesia using the oesophageal-tracheal Combitube small adult™
    Hoerauf, KH
    Hartmann, T
    Acimovic, S
    Kopp, A
    Wiesner, G
    Gustorff, B
    Jellinek, H
    Krafft, P
    BRITISH JOURNAL OF ANAESTHESIA, 2001, 86 (01) : 124 - 126
  • [26] Comparison of cuff-pressure changes in silicone and PVC laryngeal masks during nitrous oxide anaesthesia in spontaneously breathing children
    van Zundert, Tom
    Brimacombe, Joseph
    ANAESTHESIOLOGY INTENSIVE THERAPY, 2012, 44 (02) : 63 - 70
  • [27] THE EFFECT OF ALFENTANIL ON CEREBRAL BLOOD-FLOW VELOCITY AND INTRACRANIAL-PRESSURE DURING ISOFLURANE-NITROUS OXIDE ANESTHESIA IN HUMANS
    MAYBERG, TS
    LAM, AM
    ENG, CC
    LAOHAPRASIT, V
    WINN, HR
    ANESTHESIOLOGY, 1993, 78 (02) : 288 - 294
  • [28] An assessment of a method of inflating cuffs with a nitrous oxide gas mixture to prevent an increase in intracuff pressure in five different tracheal tube designs
    Karasawa, F
    Tokunaga, M
    Aramaki, Y
    Shizukuishi, M
    Satoh, T
    ANAESTHESIA, 2001, 56 (02) : 155 - 159
  • [29] Functional and radiographic outcomes of Gartland type II supracondylar humerus fractures managed by closed reduction under nitrous oxide as conscious sedation
    Pierantoni, Silvia
    Alberghina, Flavia
    Cravino, Mattia
    Paonessa, Matteo
    Canavese, Federico
    Andreacchio, Antonio
    JOURNAL OF PEDIATRIC ORTHOPAEDICS-PART B, 2020, 29 (02): : 117 - 125
  • [30] Corrosion Behavior of L415 Natural Gas Pipeline in High pressure Oxygen-enriched and High salt Environment
    Li, Hailing
    Wan, Hongxia
    Liu, Zhiyong
    Du, Cuiwei
    Liu, Yuning
    Li, Xiaogang
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2017, 12 (08): : 6940 - 6951