Oxygen contamination of the nitrous oxide pipeline supply

被引:7
作者
Lye, A [1 ]
Patrick, R [1 ]
机构
[1] Royal Prince Alfred Hosp, Dept Anaesthet, Camperdown, NSW 2050, Australia
关键词
equipment; failure; nitrous oxide oxygen blender; Midogas; oxygen alarm;
D O I
10.1177/0310057X9802600215
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Midogas(R) (CIG Australia) nitrous oxide/oxygen blenders are commonly used in delivery wards in Australia. In this ease report we describe an incident where a small hole in the diaphragm of the oxygen failure alarm in a Midogas nitrous oxide/oxygen blender led to retrograde flow of oxygen contaminating the nitrous oxide pipeline supply to the operating theatres and the delivery ward. This caused a reduced level of nitrous oxide to the patient in theatre, but there was no adverse outcome, However if the oxygen pipeline pressure had been lower than the nitrous oxide pipeline pressure, the contamination would have been reversed potentially resulting in a hypoxic mixture being delivered to many patients, Suggestions are made as to a method of prevention of this problem.
引用
收藏
页码:207 / 209
页数:3
相关论文
共 34 条
  • [1] Five oxygen nitrous oxide proportioning systems compared
    Feigenwinter, P
    Pohle, TH
    Zbinden, AM
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 1996, 13 (01) : 73 - 80
  • [2] NITROUS OXIDE-OXYGEN ANALGESIA - THE PERFORMANCE OF THE MC MASK DELIVERY SYSTEM
    OOI, R
    JOSHI, P
    SONI, N
    JOURNAL OF THE ROYAL SOCIETY OF MEDICINE, 1992, 85 (09) : 534 - 536
  • [3] NITROUS-OXIDE ADMINISTRATION USING COMMONLY AVAILABLE OXYGEN-THERAPY DEVICES
    JOSHI, P
    OOI, R
    SONI, N
    BRITISH JOURNAL OF ANAESTHESIA, 1992, 68 (06) : 630 - 632
  • [4] Central oxygen pipeline failure
    Mostert, Lelane
    Coetzee, Andre R.
    SOUTHERN AFRICAN JOURNAL OF ANAESTHESIA AND ANALGESIA, 2014, 20 (05) : 214 - 217
  • [5] LIQUID FULL NITROUS-OXIDE CYLINDERS
    MEYER, RM
    FERDERBAR, PJ
    ANESTHESIOLOGY, 1993, 78 (03) : 584 - 586
  • [6] Nitric oxide: Description of a pipeline delivery system
    Whiteley, SM
    Cohen, AT
    Laycock, D
    Arrowsmith, WM
    Henrys, PD
    ANAESTHESIA, 1997, 52 (06) : 561 - 566
  • [7] Anaesthetists' management of oxygen pipeline failure: room for improvement
    Weller, J.
    Merry, A.
    Warman, G.
    Robinson, B.
    ANAESTHESIA, 2007, 62 (02) : 122 - 126
  • [8] Choice of cuff inflation medium during nitrous oxide anaesthesia
    Mitchell, V
    Adams, T
    Calder, I
    ANAESTHESIA, 1999, 54 (01) : 32 - 36
  • [9] OXYGEN-SUPPLY FAILURE REVISITED
    MENDEL, P
    JOURNAL OF CLINICAL MONITORING, 1994, 10 (06): : 405 - 406
  • [10] Pollution with nitrous oxide using laryngeal mask or face mask
    Jenstrup, M
    Fruergaard, KO
    Mortensen, CR