A Novel Subcutaneous Counterpulsation Device: Acute Hemodynamic Efficacy During Pharmacologically Induced Hypertension, Hypotension, and Heart Failure

被引:24
作者
Bartoli, Carlo R. [1 ,2 ,3 ,4 ,5 ]
Wilson, Gregory C. [2 ]
Giridharan, Guruprasad A. [6 ]
Slaughter, Mark S. [2 ]
Sherwood, Leslie C. [2 ]
Spence, Paul A. [7 ]
Prabhu, Sumanth D. [3 ,4 ,5 ]
Koenig, Steven C. [1 ,6 ]
机构
[1] Univ Louisville, Sch Med, MD PhD Program, Louisville, KY 40292 USA
[2] Univ Louisville, Dept Surg, Cardiovasc Innovat Inst, Louisville, KY 40292 USA
[3] Univ Louisville, Dept Med, Louisville, KY 40292 USA
[4] Univ Louisville, Inst Mol Cardiol, Louisville, KY 40292 USA
[5] Louisville VAMC, Louisville, KY USA
[6] Univ Louisville, Dept Bioengn, Louisville, KY 40292 USA
[7] SCR Inc, Louisville, KY USA
关键词
Heart failure; Counterpulsation; Counterpulsation device; Intra-aortic balloon pump; Cardiac surgery; VENTRICULAR ASSIST DEVICE; INTRAAORTIC BALLOON COUNTERPULSATION; REPERFUSION; THERAPY; MODEL; TERM; PUMP;
D O I
10.1111/j.1525-1594.2010.01009.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The miniaturization of mechanical assist devices and less invasive implantation techniques may lead to earlier intervention in patients with heart failure. As such, we evaluated the effectiveness of a novel, minimally invasive, implantable counterpulsation device (CPD) in augmenting cardiac function during impaired hemodynamics. We compared the efficacy of a 32-mL stroke volume CPD with a standard 40-mL intra-aortic balloon pump (IABP) over a range of clinically relevant pathophysiological conditions. Male calves were instrumented via thoracotomy, the CPD was anastomosed to the left carotid artery, and the IABP was positioned in the descending aorta. Hemodynamic conditions of hypertension, hypotension, and heart failure were pharmacologically simulated and data were recorded during CPD and IABP support (off, 1:2, 1:1 modes) for each condition. In all three pathophysiological conditions, the CPD and IABP produced similar and statistically significant (P < 0.05) increases in coronary artery blood flow normalized to the left ventricular (LV) workload. During hypotension and heart failure conditions, however, the CPD produced significantly greater reductions in LV workload and myocardial oxygen consumption as compared with the IABP. A novel 32-mL CPD connected to a peripheral artery produced equivalent or greater hemodynamic benefits than a standard 40-mL IABP during pharmacologically induced hypertension, hypotension, and heart failure conditions.
引用
收藏
页码:537 / 545
页数:9
相关论文
共 30 条
[1]   The role of bridge to transplantation: should LVAD patients be transplanted? [J].
Birks, EJ ;
Yacoub, MH ;
Banner, NR ;
Khaghani, A .
CURRENT OPINION IN CARDIOLOGY, 2004, 19 (02) :148-153
[2]   Left ventricular assist device and drug therapy for the reversal of heart failure [J].
Birks, Emma J. ;
Tansley, Patrick D. ;
Hardy, James ;
George, Robert S. ;
Bowles, Christopher T. ;
Burke, Margaret ;
Banner, Nicholas R. ;
Khaghani, Asghar ;
Yacoub, Magdi H. .
NEW ENGLAND JOURNAL OF MEDICINE, 2006, 355 (18) :1873-1884
[3]   Multicenter experience: Prevention and management of left ventricular assist device infections [J].
Chinn, R ;
Dembitsky, W ;
Eaton, L ;
Chillcott, S ;
Stahovich, M ;
Rasmusson, B ;
Pagani, F .
ASAIO JOURNAL, 2005, 51 (04) :461-470
[4]   Myocardial expression of the arginine: Glycine amidinotransferase gene is elevated in heart failure and normalized after recovery - Potential implications for local creatine synthesis [J].
Cullen, Martin E. ;
Yuen, Ada H. Y. ;
Felkin, Leanne E. ;
Smolenski, Ryszard T. ;
Hall, Jennifer L. ;
Grindle, Suzanne ;
Miller, Leslie W. ;
Birks, Emma J. ;
Yacoub, Magdi H. ;
Barton, Paul J. R. .
CIRCULATION, 2006, 114 :I16-I20
[5]   Short- and long-term mechanical ventricular assistance towards myocardial recovery [J].
Entwistle, JWC .
SURGICAL CLINICS OF NORTH AMERICA, 2004, 84 (01) :201-+
[6]   Predicted hemodynamic benefits of counterpulsation therapy using a superficial surgical approach [J].
Giridharan, GA ;
Pantalos, GM ;
Litwak, KN ;
Spence, PA ;
Koenig, SC .
ASAIO JOURNAL, 2006, 52 (01) :39-46
[7]   Medical progress - Implantable left ventricular assist devices [J].
Goldstein, DJ ;
Oz, MC ;
Rose, EA .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 339 (21) :1522-1533
[8]  
Hall JL, 2007, EUR HEART J, V28, P613, DOI 10.1093/eurheartj/ehl365
[9]   Circulatory assistance with a permanent implantable IABP: Initial human experience [J].
Jeevanandam, V ;
Jayakar, D ;
Anderson, AS ;
Martin, S ;
Piccione, W ;
Heroux, AL ;
Wynne, J ;
Stephenson, LW ;
Hsu, J ;
Freed, PS ;
Kantrowitz, A .
CIRCULATION, 2002, 106 (13) :I183-I188
[10]  
Kinoshita M, 1988, ASAIO Trans, V34, P208