Optical fibre based real-time measurements during an LDR prostate brachytherapy implant simulation: using a 3D printed anthropomorphic phantom

被引:11
作者
Woulfe, P. [1 ,3 ]
Sullivan, F. J. [4 ,5 ]
Byrne, L. [1 ]
Doyle, A. J. [6 ]
Kam, W. [1 ,2 ]
Martyn, M. [3 ]
O'Keeffe, S. [1 ,2 ]
机构
[1] Univ Limerick, Opt Fibre Sensors Res Ctr, Limerick, Ireland
[2] Univ Limerick, Hlth Res Inst, Limerick V94 T9PX, Ireland
[3] Galway Clin, Dept Radiotherapy Phys, Galway, Ireland
[4] Galway Clin, Prostate Canc Inst, Galway, Ireland
[5] Galway Clin, Dept Radiotherapy, Galway, Ireland
[6] Technol Univ Dublin, Sch Phys, FOCAS, Dublin, Ireland
基金
爱尔兰科学基金会;
关键词
I-125; PARAMETERS; DOSIMETRY;
D O I
10.1038/s41598-021-90880-6
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
An optical fibre sensor based on radioluminescence, using the scintillation material terbium doped gadolinium oxysulphide (Gd2O2S:Tb) is evaluated, using a 3D printed anthropomorphic phantom for applications in low dose-rate (LDR) prostate brachytherapy. The scintillation material is embedded in a 700 mu m diameter cavity within a 1 mm plastic optical fibre that is fixed within a brachytherapy needle. The high spatial resolution dosimeter is used to measure the dose contribution from Iodine-125 (I-125) seeds. Initially, the effects of sterilisation on the sensors (1) repeatability, (2) response as a function of angle, and (3) response as a function of distance, are evaluated in a custom polymethyl methacrylate phantom. Results obtained in this study demonstrate that the output response of the sensor, pre- and post-sterilisation are within the acceptable measurement uncertainty ranging from a maximum standard deviation of 4.7% pre and 5.5% post respectively, indicating that the low temperature sterilisation process does not damage the sensor or reduce performance. Subsequently, an LDR brachytherapy plan reconstructed using the VariSeed treatment planning system, in an anthropomorphic 3D printed training phantom, was used to assess the suitability of the sensor for applications in LDR brachytherapy. This phantom was printed based on patient anatomy, with the volume and dimensions of the prostate designed to represent that of the patient. I-125 brachytherapy seeds, with an average activity of 0.410 mCi, were implanted into the prostate phantom under trans-rectal ultrasound guidance; following the same techniques as employed in clinical practice by an experienced radiation oncologist. This work has demonstrated that this sensor is capable of accurately identifying when radioactive I-125 sources are introduced into the prostate via a brachytherapy needle.
引用
收藏
页数:8
相关论文
共 16 条
  • [1] American Brachytherapy Society consensus guidelines for transrectal ultrasound-guided permanent prostate brachytherapy
    Davis, Brian J.
    Horwitz, Eric M.
    Lee, W. Robert
    Crook, Juanita M.
    Stock, Richard G.
    Merrick, Gregory S.
    Butler, Wayne M.
    Grimm, Peter D.
    Stone, Nelson N.
    Potters, Louis
    Zietman, Anthony L.
    Zelefsky, Michael J.
    [J]. BRACHYTHERAPY, 2012, 11 (01) : 6 - 19
  • [2] Doyle AJ, 2019, ESTABLISHING HIGH QU
  • [3] DEVELOPMENT AND PRELIMINARY EVALUATION OF AN ANTHROPOMORPHIC TRANS-RECTAL ULTRASOUND PROSTATE BRACHYTHERAPY TRAINING PHANTOM
    Doyle, Andrea J.
    Sullivan, Frank
    Walsh, John
    King, Deirdre M.
    Cody, Dervil
    Browne, Jacinta E.
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2021, 47 (03) : 833 - 846
  • [4] A COMPARISON OF SOLID PHANTOMS WITH WATER FOR DOSIMETRY OF I125 BRACHYTHERAPY SOURCES
    MEIGOONI, AS
    MELI, JA
    NATH, R
    [J]. MEDICAL PHYSICS, 1988, 15 (05) : 695 - 701
  • [5] Monte Carlo calculations of AAPM Task Group Report No. 43 dosimetry parameters for the 125I I-Seed AgX100 source model
    Mourtada, Firas
    Mikell, Justin
    Ibbott, Geoffrey
    [J]. BRACHYTHERAPY, 2012, 11 (03) : 237 - 244
  • [6] ANISOTROPY FUNCTIONS FOR PD-103, I-125, AND IR-192 INTERSTITIAL BRACHYTHERAPY SOURCES
    NATH, R
    MEIGOONI, AS
    MUENCH, P
    MELILLO, A
    [J]. MEDICAL PHYSICS, 1993, 20 (05) : 1465 - 1473
  • [7] OKeeffe S., 2018, SPIE BIOS, V10488
  • [8] Phantom size in brachytherapy source dosimetric studies
    Pérez-Calatayud, J
    Granero, D
    Ballester, F
    [J]. MEDICAL PHYSICS, 2004, 31 (07) : 2075 - 2081
  • [9] Review of intraoperative imaging and planning techniques in permanent seed prostate brachytherapy
    Polo, Alfredo
    Salembier, Carl
    Venselaar, Jack
    Hoskin, Peter
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 94 (01) : 12 - 23
  • [10] Low dose rate prostate brachytherapy
    Stish, Bradley J.
    Davis, Brian J.
    Mynderse, Lance A.
    McLaren, Robert H.
    Deufel, Christopher L.
    Choo, Richard
    [J]. TRANSLATIONAL ANDROLOGY AND UROLOGY, 2018, 7 (03) : 341 - 356