Analysis of flow changes in side branches jailed by flow diverters in rabbit models

被引:27
作者
Cebral, Juan R. [1 ]
Raschi, Marcelo [1 ]
Mut, Fernando [1 ]
Ding, Yong-Hond [2 ]
Dai, Daying [2 ]
Kadirvel, Ramanathan [2 ]
Kallmes, David [2 ]
机构
[1] George Mason Univ, Coll Sci, Ctr Computat Fluid Dynam, Fairfax, VA 22030 USA
[2] Mayo Clin, Neuroradiol Res Lab, Rochester, MN 55905 USA
基金
美国国家卫生研究院;
关键词
cerebral aneurysms; flow diverters; perforators; computational fluid dynamics; rabbit models; PIPELINE EMBOLIZATION DEVICE; ENDOVASCULAR TREATMENT; INTRACRANIAL ANEURYSMS; BLOOD-FLOW; CEREBRAL ANEURYSMS; ARTERIAL BRANCHES; IN-VITRO; STENT; HEMODYNAMICS; DIVERSION;
D O I
10.1002/cnm.2640
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Understanding the flow alteration in side branches during flow diversion treatment of cerebral aneurysms is important to prevent ischemic complications and improve device designs. Flow diverters were placed in the aorta of four rabbits crossing the origin of side arteries. Subject-specific computational models were constructed from 3D angiographies and Doppler ultrasounds (DUSs). Flow simulations were run before and after virtually deploying the flow diverters, assuming distal resistances remained unchanged after treatment. All jailed arteries remained patent angiographically 8weeks after treatment. The computational models estimated decreases compared to pretreatment in the mean flow rates between 2% and 20% and in peak flow rates between 5% and 36%. The major changes were observed during systole. Flow patterns did not exhibit recirculation zones before treatment. Implantation of the flow diverters altered the flow structure only locally near the device wires. No major recirculation regions were created or destroyed. Flow diverters seem safe with respect to perforator or side branch occlusion. Relatively small changes in flow rates through jailed arteries are expected, even for moderate to large degrees of coverage of their origins. These results seem consistent with previous clinical experiences where no or very few complications related to perforator occlusion have been reported.Copyright (c) 2014 John Wiley & Sons, Ltd.
引用
收藏
页码:988 / 999
页数:12
相关论文
共 43 条
[1]   Computational fluid dynamics of stented intracranial aneurysms using adaptive embedded unstructured grids [J].
Appanaboyina, S. ;
Mut, F. ;
Lohner, R. ;
Putman, C. A. ;
Cebral, J. R. .
INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN FLUIDS, 2008, 57 (05) :475-493
[2]   Computational modelling of blood flow in side arterial branches after stenting of cerebral aneurysms [J].
Appanaboyina, Sunil ;
Mut, Fernando ;
Loehner, Rainald ;
Scrivano, Esteban ;
Miranda, Carlos ;
Lylyk, Pedro ;
Putman, Christopher ;
Cebral, Juan .
INTERNATIONAL JOURNAL OF COMPUTATIONAL FLUID DYNAMICS, 2008, 22 (10) :669-676
[3]  
Bears J, 1972, DYNAMICS FLUIDS PORO
[4]   Endovascular Treatment of Intracranial Aneurysms With Flow Diverters A Meta-Analysis [J].
Brinjikji, Waleed ;
Murad, Mohammad H. ;
Lanzino, Giuseppe ;
Cloft, Harry J. ;
Kallmes, David F. .
STROKE, 2013, 44 (02) :442-447
[5]   Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective Study [J].
Byrne, James V. ;
Beltechi, Radu ;
Yarnold, Julia A. ;
Birks, Jacqueline ;
Kamran, Mudassar .
PLOS ONE, 2010, 5 (09) :1-8
[6]  
Castro MA, 2006, AM J NEURORADIOL, V27, P1703
[7]   Aneurysm Rupture Following Treatment with Flow-Diverting Stents: Computational Hemodynamics Analysis of Treatment [J].
Cebral, J. R. ;
Mut, F. ;
Raschi, M. ;
Scrivano, E. ;
Ceratto, R. ;
Lylyk, P. ;
Putman, C. M. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2011, 32 (01) :27-33
[8]   Efficient pipeline for image-based patient-specific analysis of cerebral aneurysm hemodynamics: Technique and sensitivity [J].
Cebral, JR ;
Castro, MA ;
Appanaboyina, S ;
Putman, CM ;
Millan, D ;
Frangi, AF .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2005, 24 (04) :457-467
[9]   Comparison of Flow Diversion and Coiling in Large Unruptured Intracranial Saccular Aneurysms [J].
Chalouhi, Nohra ;
Tjoumakaris, Stavropoula ;
Starke, Robert M. ;
Gonzalez, L. Fernando ;
Randazzo, Ciro ;
Hasan, David ;
McMahon, Jeffrey F. ;
Singhal, Saurabh ;
Moukarzel, Lea A. ;
Dumont, Aaron S. ;
Rosenwasser, Robert ;
Jabbour, Pascal .
STROKE, 2013, 44 (08) :2150-2154
[10]  
Chongs W, 2013, AJNR AM J NEURORADIO