2000 Volvo Award Winner in Biomechanical Studies -: Monitoring in vivo implant loads with a telemeterized internal spinal fixation device

被引:107
作者
Rohlmann, A
Graichen, F
Weber, U
Bergmann, G
机构
[1] Free Univ Berlin, Iskar Helene Heim, Biomech Lab, D-14195 Berlin, Germany
[2] Free Univ Berlin, Iskar Helene Heim, Dept Orthopaed Surg, D-14195 Berlin, Germany
关键词
bone healing; implant loading; internal spinal fixators; load measurement; spine; telemetry; temporal course;
D O I
10.1097/00007632-200012010-00004
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Implant loads were measured in 10 patients using telemeterized internal spinal fixation devices. Objective. To determine the postoperative temporal course of implant loads. Summary of Background Data. Little information exists regarding the temporal course of loads on internal spinal fixation devices. Methods. The telemeterized internal spinal fixator allows the measurement of three force components and three moments acting in the fixator. Implant loads were determined in up to 20 measuring sessions for different activities, including walking, standing, sitting; lying in the supine position, and lifting an extended leg while in the supine position. Results. Implant loads often increased shortly after anterior interbody fusion was performed. Several patients retained the same high level even after fusion had taken place. This explains the reason why screw breakage sometimes occurs more than half a year after implantation. The time of fusion could not be pinpointed from the loading curves. Conclusions. The results show that fixators may be highly loaded even after fusion has occurred. A flexion bending moment acts on the implant even with the body in a relaxed lying position. This means that already shortly after the anterior procedure, the shape of the spine is not neutral and unloaded, but slightly deformed, which loads the fixators. Pedicle screw breakage more than half a year after insertion does not prove that anterior interbody fusion has not occurred.
引用
收藏
页码:2981 / 2986
页数:6
相关论文
共 22 条
  • [11] Changes in the loads on an internal spinal fixator after iliac-crest autograft
    Rohlmann, A
    Bergmann, G
    Graichen, F
    Weber, U
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2000, 82B (03): : 445 - 449
  • [12] Telemeterized load measurement using instrumented spinal internal fixators in a patient with degenerative instability
    Rohlmann, A
    Bergmann, G
    Graichen, F
    Mayer, HM
    [J]. SPINE, 1995, 20 (24) : 2683 - 2689
  • [13] A SPINAL FIXATION DEVICE FOR IN-VIVO LOAD MEASUREMENT
    ROHLMANN, A
    BERGMANN, G
    GRAICHEN, F
    [J]. JOURNAL OF BIOMECHANICS, 1994, 27 (07) : 961 - &
  • [14] Influence of muscle forces on loads in internal spinal fixation devices
    Rohlmann, A
    Bergmann, G
    Graichen, F
    Mayer, HM
    [J]. SPINE, 1998, 23 (05) : 537 - 542
  • [15] Comparison of loads on internal spinal fixation devices measured in vitro and in vivo
    Rohlmann, A
    Bergmann, G
    Graichen, F
    Weber, U
    [J]. MEDICAL ENGINEERING & PHYSICS, 1997, 19 (06) : 539 - 546
  • [16] Loads on internal spinal fixators measured in different body positions
    Rohlmann, A
    Bergmann, G
    Graichen, F
    [J]. EUROPEAN SPINE JOURNAL, 1999, 8 (05) : 354 - 359
  • [17] Schlapfer F, 1980, I MECH ENG C, VC131/80, P59
  • [18] White AugaustA., 1990, CLIN BIOMECH, V2e
  • [19] MEASURING METHOD FOR INVIVO MEASUREMENTS IN IMPLANTS FOR FIXATION OF SPINAL FRACTURES
    WILKE, HJ
    DURSELEN, L
    CLAES, L
    WORSDORFER, O
    [J]. BIOMEDIZINISCHE TECHNIK, 1992, 37 (04): : 78 - 85
  • [20] WORSDORFER O, 1981, THESIS U ULM