Development of a reliable microelectrode dissolved oxygen sensor

被引:63
作者
Sosna, Maciej
Denuault, Guy [1 ]
Pascal, Robin W.
Prien, Ralf D.
Mowlem, Matt
机构
[1] Univ Southampton, Sch Chem, Southampton SO17 1BJ, Hants, England
[2] Univ Southampton, Natl Oceanog Ctr, Southampton SO14 3ZH, Hants, England
[3] Baltic Sea Res Inst, D-18119 Rostock, Warnemunde, Germany
基金
英国自然环境研究理事会;
关键词
dissolved oxygen; amperometric oxygen sensor; calibration; microelectrode; microdisc; in situ measurements;
D O I
10.1016/j.snb.2006.08.033
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This article reports the results of a careful experimental and analytical investigation which led to the development of an accurate and reproducible microelectrode dissolved oxygen sensor. Primarily designed for oceanographic applications but also applicable to environmental and water process monitoring, the sensor measures the diffusion controlled current to a bare Pt microdisc electrode for the reduction of oxygen. A successful reconditioning potential waveform is reported which yields a very stable amperometric response over continuous operation, with a maximum deviation of the limiting current under 1.5% over 24h. An automated calibration method developed to accurately characterise the electrodes is described. Excellent linearity is obtained for all electrodes tested and in each case, the number of apparent electrons for the reduction of oxygen is reported. As an alternative to calibration, an analytical treatment which accounts for temperature and salinity effects is given to calculate the dissolved oxygen concentration directly from the limiting current. While the analytical approach yields a concentration relative error circa 11% for a 50 mu m diameter Pt disc, the calibration, has lower errors and yields a detection limit down to 0.9 mu M with the same disc. Although this investigation builds on established principles, this article describes, for the first time, the conditions required to obtain accurate and reproducible measurements and provides an estimate of their precision. Preliminary field trials to measure oxygen depth profiles in the ocean have proved very encouraging [R. Prien, R. Pascal, M. Mowlem, G. Denuault, M. Sosna, Development and first results of a new fast response microelectrode DO-sensor, in Oceans 2005-Europe, Vols. 1 and 2, 2005, pp. 744-747]. (c) 2006 Published by Elsevier B.V.
引用
收藏
页码:344 / 351
页数:8
相关论文
共 48 条
[21]   OXYGEN INDUCED ELECTROREDUCTION OF HYDROGEN PEROXIDE AND REDUCTION OF OXYGEN AT THE ROTATED GOLD WIRE ELECTRODE [J].
KOLTHOFF, IM ;
JORDAN, J .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1952, 74 (19) :4801-4805
[22]   CALIBRATION OF CLARK OXYGEN ELECTRODE FOR USE IN AQUEOUS SOLUTIONS [J].
LEFEVRE, ME .
JOURNAL OF APPLIED PHYSIOLOGY, 1969, 26 (06) :844-&
[23]   In situ deployment of voltammetric, potentiometric. and amperometric microelectrodes from a ROV to determine dissolved O2, Mn, Fe, S(-2), and pH in porewaters [J].
Luther, GW ;
Reimers, CE ;
Nuzzio, DB ;
Lovalvo, D .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1999, 33 (23) :4352-4356
[24]   Simultaneous measurement of O2, Mn, Fe, I-, and S(-II) in marine pore waters with a solid-state voltammetric microelectrode [J].
Luther, GW ;
Brendel, PJ ;
Lewis, BL ;
Sundby, B ;
Lefrancois, L ;
Silverberg, N ;
Nuzzio, DB .
LIMNOLOGY AND OCEANOGRAPHY, 1998, 43 (02) :325-333
[25]   Chemical speciation drives hydrothermal vent ecology [J].
Luther, GW ;
Rozan, TF ;
Taillefert, M ;
Nuzzio, DB ;
Di Meo, C ;
Shank, TM ;
Lutz, RA ;
Cary, SC .
NATURE, 2001, 410 (6830) :813-816
[26]  
Millero, 2006, CHEM OCEANOGRAPHY
[27]  
Montenegro M. I., 1991, MICROELECTRODES THEO
[28]   Towards a microelectrode sensor for the determination of oxygen in waters [J].
Pletcher, D ;
Sotiropoulos, S .
ANALYTICA CHIMICA ACTA, 1996, 322 (1-2) :83-90
[29]   A STUDY OF CATHODIC OXYGEN REDUCTION AT PLATINUM USING MICROELECTRODES [J].
PLETCHER, D ;
SOTIROPOULOS, S .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1993, 356 (1-2) :109-119
[30]  
PRIEN R, 2005, DEV 1 RESULTS NEW FA, V1, P744