Small-Field Measurements of 3D Polymer Gel Dosimeters through Optical Computed Tomography

被引:17
|
作者
Shih, Tian-Yu [1 ,2 ,3 ]
Wu, Jay [4 ]
Shih, Cheng-Ting [5 ]
Lee, Yao-Ting [2 ]
Wu, Shin-Hua [3 ]
Yao, Chun-Hsu [1 ,6 ,7 ]
Hsieh, Bor-Tsung [2 ]
机构
[1] China Med Univ, Grad Inst Clin Med Sci, Taichung, Taiwan
[2] Cent Taiwan Univ Sci & Technol, Dept Med Imaging & Radiol Sci, Taichung, Taiwan
[3] Cheng Ching Hosp Chung Kang, Dept Radiol, Taichung, Taiwan
[4] Natl Yang Ming Univ, Dept Biomed Imaging & Radiol Sci, Taipei 112, Taiwan
[5] China Med Univ Hosp, Printing Med Res Ctr 3D, Taichung, Taiwan
[6] China Med Univ, Coll Hlth Care, Dept Biomed Imaging & Radiol Sci, Taichung, Taiwan
[7] Asia Univ, Dept Biomed Informat, Taichung, Taiwan
来源
PLOS ONE | 2016年 / 11卷 / 03期
关键词
RADIATION-DOSE DISTRIBUTIONS; NMR;
D O I
10.1371/journal.pone.0151300
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
With advances in therapeutic instruments and techniques, three-dimensional dose delivery has been widely used in radiotherapy. The verification of dose distribution in a small field becomes critical because of the obvious dose gradient within the field. The study investigates the dose distributions of various field sizes by using NIPAM polymer gel dosimeter. The dosimeter consists of 5% gelatin, 5% monomers, 3% cross linkers, and 5 mM THPC. After irradiation, a 24 to 96 hour delay was applied, and the gel dosimeters were read by a cone beam optical computed tomography (optical CT) scanner. The dose distributions measured by the NIPAM gel dosimeter were compared to the outputs of the treatment planning system using gamma evaluation. For the criteria of 3%/3 mm, the pass rates for 5 x 5, 3 x 3, 2 x 2, 1 x 1, and 0.5 x 0.5 cm(2) were as high as 91.7%, 90.7%, 88.2%, 74.8%, and 37.3%, respectively. For the criteria of 5%/5 mm, the gamma pass rates of the 5 x 5, 3 x 3, and 2 x 2 cm2 fields were over 99%. The NIPAM gel dosimeter provides high chemical stability. With cone-beam optical CT readouts, the NIPAM polymer gel dosimeter has potential for clinical dose verification of small-field irradiation.
引用
收藏
页数:15
相关论文
共 50 条
  • [21] 3D Polymer Gel Dosimeters with iCBCT 3D Reading and polyGeVero-CT Software Package for Quality Assurance in Radiotherapy
    Kozicki, Marek
    Maras, Piotr
    Jaszczak-Kuligowska, Malwina
    MATERIALS, 2024, 17 (06)
  • [22] A method for obtaining three-dimensional measurements of HDR brachytherapy dose distributions using Fricke gel dosimeters and optical computed tomography
    Andre Asena
    Sanna Nilsson
    Shaun T. Smith
    Tanya Kairn
    Scott B. Crowe
    Jamie V. Trapp
    Australasian Physical & Engineering Sciences in Medicine, 2019, 42 : 221 - 226
  • [23] 3D Terahertz Computed Tomography
    Kim, Geun-Ju
    Kim, Jung-Il
    Jeon, Seok-Gy
    Kim, Jaehong
    2012 37TH INTERNATIONAL CONFERENCE ON INFRARED, MILLIMETER, AND TERAHERTZ WAVES (IRMMW-THZ), 2012,
  • [24] Demonstration of 3D computed tomography of chemiluminescence with a restricted field of view
    Liu, Hecong
    Yu, Tao
    Zhang, Man
    Cai, Weiwei
    APPLIED OPTICS, 2017, 56 (25) : 7107 - 7115
  • [25] A method for obtaining three-dimensional measurements of HDR brachytherapy dose distributions using Fricke gel dosimeters and optical computed tomography
    Asena, Andre
    Nilsson, Sanna
    Smith, Shaun T.
    Kairn, Tanya
    Crowe, Scott B.
    Trapp, Jamie V.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2019, 42 (01) : 221 - 226
  • [26] An experimental study of the dose response of polymer gel dosimeters imaged with x-ray computed tomography
    Trapp, JV
    Bäck, SÅJ
    Lepage, M
    Michael, G
    Baldock, C
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (11): : 2939 - 2951
  • [27] Capabilities and limitations of 3D flame measurements based on computed tomography of chemiluminescence
    Li, Xuesong
    Ma, Lin
    COMBUSTION AND FLAME, 2015, 162 (03) : 642 - 651
  • [28] Measurements of Vapor in Confined-space Based on 3D Computed Tomography
    Ling, Chen
    Chen, Haiyan
    Wu, Yue
    Kung Cheng Je Wu Li Hsueh Pao/Journal of Engineering Thermophysics, 2024, 45 (01): : 271 - 278
  • [29] Gamma Knife Perfexion Output Factor Measurements Using PRESAGE 3D Optical Dosimeters
    Klawikowski, S.
    Yang, J.
    Adamovics, J.
    Ibbott, G.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [30] 3D small-field surface imaging based on microscopic fringe projection profilometry:a review
    Wang Yong-hong
    Zhang Qian
    Hu Yin
    Wang Huan-qing
    CHINESE OPTICS, 2021, 14 (03): : 447 - 457