Single-fraction spine SBRT end-to-end testing on TomoTherapy, Vero, TrueBeam, and CyberKnife treatment platforms using a novel anthropomorphic phantom

被引:36
作者
Gallo, John J. [1 ]
Kaufman, Isaac [2 ]
Powell, Rachel [3 ]
Pandya, Shalini [4 ]
Somnay, Archana [5 ]
Bossenberger, Todd [6 ,7 ]
Ramirez, Ezequiel [8 ]
Reynolds, Robert [8 ]
Solberg, Timothy [9 ]
Burmeister, Jay [6 ,7 ]
机构
[1] Eastern Maine Med Ctr, Brewer, ME USA
[2] McLaren Northern Michigan Hosp, Dept Radiat Oncol, Petoskey, MI USA
[3] St John Providence Hlth Syst, Webber Canc Ctr, Warren, MI USA
[4] Oakwood Healthcare Syst, Dept Radiat Oncol, Dearborn, MI USA
[5] Huron Valley Sinai Hosp, Dept Radiat Oncol, Commerce Township, MI USA
[6] Wayne State Univ, Sch Med, Dept Radiat Oncol, Detroit, MI USA
[7] Karmanos Canc Ctr, Gershenson ROC, Detroit, MI USA
[8] Univ Texas SW Med Ctr Dallas, Dept Radiat Oncol, Dallas, TX 75390 USA
[9] Univ Penn, Sch Med, Dept Radiat Oncol, Philadelphia, PA 19104 USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2015年 / 16卷 / 01期
关键词
stereotactic body radiotherapy; SBRT; single-fraction spine; SBRT end-to-end testing; RADIOCHROMIC FILM DOSIMETRY; BODY RADIATION-THERAPY; STEREOTACTIC RADIOSURGERY; CLINICAL-EXPERIENCE; EBT3; FILMS; RADIOTHERAPY; CALIBRATION; PHOTON; TUMORS; PAIN;
D O I
10.1120/jacmp.v16i1.5120
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Spine SBRT involves the delivery of very high doses of radiation to targets adjacent to the spinal cord and is most commonly delivered in a single fraction. Highly conformal planning and accurate delivery of such plans is imperative for successful treatment without catastrophic adverse effects. End-to-end testing is an important practice for evaluating the entire treatment process from simulation through treatment delivery. We performed end-to-end testing for a set of representative spine targets planned and delivered using four different treatment planning systems (TPSs) and delivery systems to evaluate the various capabilities of each. An anthropomorphic E2E SBRT phantom was simulated and treated on each system to evaluate agreement between measured and calculated doses. The phantom accepts ion chambers in the thoracic region and radiochromic film in the lumbar region. Four representative targets were developed within each region (thoracic and lumbar) to represent different presentations of spinal metastases and planned according to RTOG 0631 constraints. Plans were created using the TomoTherapy TPS for delivery using the Hi . Art system, the iPlan TPS for delivery using the Vero system, the Eclipse TPS for delivery using the TrueBeam system in both flattened and flattening filter free (FFF), and the MultiPlan TPS for delivery using the CyberKnife system. Delivered doses were measured using a 0.007 cm(3) ion chamber in the thoracic region and EBT3 GAFCHROMIC film in the lumbar region. Films were scanned and analyzed using an Epson Expression 10000XL flatbed scanner in conjunction with FilmQAPro2013. All treatment platforms met all dose constraints required by RTOG 0631. Ion chamber measurements in the thoracic targets delivered an overall average difference of 1.5%. Specifically, measurements agreed with the TPS to within 2.2%, 3.2%, 1.4%, 3.1%, and 3.0% for all three measureable cases on TomoTherapy, Vero, TrueBeam (FFF), TrueBeam (flattened), and CyberKnife, respectively. Film measurements for the lumbar targets resulted in average global gamma index passing rates of 100% at 3%/3 mm, 96.9% at 2%/2 mm, and 61.8% at 1%/1 mm, with a 10% minimum threshold for all plans on all platforms. Local gamma analysis was also performed with similar results. While gamma passing rates were consistently accurate across all platforms through 2%/2 mm, treatment beam-on delivery times varied greatly between each platform with TrueBeam FFF being shortest, averaging 4.4 min, TrueBeam using flattened beam at 9.5 min, TomoTherapy at 30.5 min, Vero at 19 min, and CyberKnife at 46.0 min. In spite of the complexity of the representative targets and their proximity to the spinal cord, all treatment platforms were able to create plans meeting all RTOG 0631 dose constraints and produced exceptional agreement between calculated and measured doses. However, there were differences in the plan characteristics and significant differences in the beam-on delivery time between platforms. Thus, clinical judgment is required for each particular case to determine most appropriate treatment planning/delivery platform.
引用
收藏
页码:170 / 182
页数:13
相关论文
共 29 条
  • [1] Ashland Advanced Materials, GAFCHR MD V3 PROD GU
  • [2] Ashland Advanced Materials, 2009, GAFCHR EBT3 SCAN HAN
  • [3] Spinal radiosurgery: technology and clinical outcomes
    Avanzo, M.
    Romanelli, P.
    [J]. NEUROSURGICAL REVIEW, 2009, 32 (01) : 1 - 12
  • [4] The biological effectiveness of intermittent irradiation as a function of overall treatment time: Development of correction factors for linac-based stereotactic radiotherapy
    Benedict, SH
    Lin, PS
    Zwicker, RD
    Huang, DT
    SchmidtUllrich, RKA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 37 (04): : 765 - 769
  • [5] Stereotactic body radiation therapy: The report of AAPM Task Group 101
    Benedict, Stanley H.
    Yenice, Kamil M.
    Followill, David
    Galvin, James M.
    Hinson, William
    Kavanagh, Brian
    Keall, Paul
    Lovelock, Michael
    Meeks, Sanford
    Papiez, Lech
    Purdie, Thomas
    Sadagopan, Ramaswamy
    Schell, Michael C.
    Salter, Bill
    Schlesinger, David J.
    Shiu, Almon S.
    Solberg, Timothy
    Song, Danny Y.
    Stieber, Volker
    Timmerman, Robert
    Tome, Wolfgang A.
    Verellen, Dirk
    Wang, Lu
    Yin, Fang-Fang
    [J]. MEDICAL PHYSICS, 2010, 37 (08) : 4078 - 4101
  • [6] Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification
    Borca, Valeria Casanova
    Pasquino, Massimo
    Russo, Giuliana
    Grosso, Pierangelo
    Cante, Domenico
    Sciacero, Piera
    Girelli, Giuseppe
    La Porta, Maria Rosa
    Tofani, Santi
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02): : 158 - 171
  • [7] Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams
    Brown, Thomas A. D.
    Hogstrom, Kenneth R.
    Alvarez, Diane
    Matthews, Kenneth L., II
    Ham, Kyungmin
    Dugas, Joseph P.
    [J]. MEDICAL PHYSICS, 2012, 39 (12) : 7412 - 7417
  • [8] Calibrating page sized Gafchromic EBT3 films
    Crijns, W.
    Maes, F.
    van der Heide, U. A.
    Van den Heuvel, F.
    [J]. MEDICAL PHYSICS, 2013, 40 (01)
  • [9] CyberKnife stereotactic radiosurgical treatment of spinal tumors for pain control and quality of life
    Degen, JW
    Gagnon, GJ
    Voyadzis, JM
    McRae, DA
    Lunsden, M
    Dieterich, S
    Molzahn, I
    Henderson, FC
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2005, 2 (05) : 540 - 549
  • [10] TREATMENT OF SPINAL TUMORS USING CYBERKNIFE FRACTIONATED STEREOTACTIC RADIOSURGERY: PAIN AND QUALITY-OF-LIFE ASSESSMENT AFTER TREATMENT IN 200 PATIENTS
    Gagnon, Gregory J.
    Nasr, Nadim M.
    Liao, Jay J.
    Molzahn, Inge
    Marsh, David
    McRae, Donald
    Henderson, Fraser C., Sr.
    [J]. NEUROSURGERY, 2009, 64 (02) : 297 - 306