Computed radiography dose data mining and surveillance as an ongoing quality assurance improvement process

被引:13
作者
Stewart, Brent K.
Kanal, Kalpana M.
Perdue, James R.
Mann, Frederick A.
机构
[1] Univ Washington, Sch Med, Dept Radiol, Seattle, WA 98195 USA
[2] Univ Washington, Harborview Med Ctr, Dept Radiol, Seattle, WA 98104 USA
关键词
ALARA concept; computed radiography; data-mining program; dosimetry; PACS; radiation dose; radiologic physics; S number;
D O I
10.2214/AJR.06.1232
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. A data-mining program extracts computed radiography (CR) sensitivity-number (S-number) information from the PACS at our institution on a monthly basis as an on-going quality assurance (QA) improvement project. These data are compared with the previous month's data and departmental S-number goals. The results are presented at monthly QA meetings. The S-number trends are then used by technologists to modify radiographic technique charts to reach the departmental S-number target range goals. CONCLUSION. This cyclic QA improvement process shows that mining PACS data can be useful in reducing patient radiation dose and interexamination dose variance.
引用
收藏
页码:7 / 11
页数:5
相关论文
共 12 条
[1]   Computed radiography excretory urography: Can the system sensitivity value be used as an image quality indicator? [J].
Andoh, K ;
Odagiri, K ;
Matsumoto, T ;
Yamashita, K ;
Ogino, I ;
Otsuka, T .
JOURNAL OF DIGITAL IMAGING, 1997, 10 (03) :132-136
[2]   Phototimer setup for CR imaging [J].
Christodoulou, EG ;
Goodsitt, MM ;
Chan, HP ;
Hepburn, TW .
MEDICAL PHYSICS, 2000, 27 (12) :2652-2658
[3]   Optimising automatic exposure control in computed radiography and the impact on patient dose [J].
Doyle, P ;
Gentle, D ;
Martin, CJ .
RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) :236-239
[4]   COMPUTED RADIOGRAPHY UTILIZING LASER-STIMULATED LUMINESCENCE - DETECTABILITY OF SIMULATED LOW-CONTRAST RADIOGRAPHIC OBJECTS [J].
HIGASHIDA, Y ;
MORIBE, N ;
HIRATA, Y ;
MORITA, K ;
DOUDANUKI, S ;
SONODA, Y ;
KATSUDA, N ;
HIAI, Y ;
MISUMI, W ;
MATSUMOTO, M ;
YOSHIOKA, S ;
TAKAHASHI, M .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1988, 12 (03) :137-145
[5]  
Kato H, 1985, Neurosurg Rev, V8, P53, DOI 10.1007/BF01744878
[6]  
KATO H, 1994, SPECIFICATION ACCEPT, P731
[7]  
Miyahara J., 1987, Computed Radiography. Ed. by, P7
[8]  
Seibert JA, 2006, MED PHYS MG, P153
[9]  
Seibert JA, 1996, ACAD RADIOL, V3, P313
[10]   COMPUTED RADIOGRAPHY UTILIZING SCANNING LASER STIMULATED LUMINESCENCE [J].
SONODA, M ;
TAKANO, M ;
MIYAHARA, J ;
KATO, H .
RADIOLOGY, 1983, 148 (03) :833-838