High Resolution, Full Field-of-View, Whole Body Photon-Counting Detector CT: System Assessment and Initial Experience

被引:6
作者
Rajendran, Kishore [1 ]
Marsh, Jeffrey [1 ]
Petersilka, Martin [2 ]
Henning, Andre [2 ]
Shanblatt, Elisabeth [3 ]
Schmidt, Bernhard [2 ]
Flohr, Thomas [2 ]
Fletcher, Joel [1 ]
McCollough, Cynthia [1 ]
Leng, Shuai [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[2] Siemens Healthcare GmbH, Forchheim, Germany
[3] Siemens Med Solut USA Inc, Malvern, PA USA
来源
MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING | 2021年 / 11595卷
基金
美国国家卫生研究院;
关键词
Photon-counting detectors; X-ray computed tomography; ultra-high spatial resolution; multi-energy CT; image quality; COMPUTED-TOMOGRAPHY;
D O I
10.1117/12.2581944
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Computed tomography (CT) using photon-counting detectors (PCD) offers dose-efficient ultra-high-resolution imaging, high iodine contrast-to-noise ratio, multi-energy and material decomposition capabilities. We have previously demonstrated the potential benefits of PCD-CT using phantoms, cadavers, and human studies on a prototype PCD-CT system. This system, however, had several limitations in terms of scan field-of-view (FOV) and longitudinal coverage. Recently, a full FOV (50 cm) PCD-CT system with wider longitudinal coverage and higher spatial resolution (0.15 mm detector pixels) has been installed in our lab capable of human scanning at clinical dose and dose rate. In this work, we share our initial experience of the new PCD-CT system and compare its performance with a state-of-the-art 3rd generation dual-source CT scanner. Basic image quality was assessed using an ACR CT accreditation phantom, high-resolution performance using an anthropomorphic head phantom, and multi-energy and material decomposition performance using a multi-energy CT phantom containing various concentrations of iodine and hydroxyapatite. Finally, we demonstrate the feasibility of high-resolution, full FOV PCD-CT imaging for improved delineation of anatomical and pathological features in a patient with pulmonary nodules.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] Potential of employing a quantum iterative reconstruction algorithm for ultra-high-resolution photon-counting detector CT of the hip
    Huflage, H.
    Grunz, J. -P
    Kunz, A. S.
    Patzer, T. S.
    Sauer, S. T.
    Christner, S. A.
    Petritsch, B.
    Erguen, S.
    Bley, T. A.
    Luetkens, K. S.
    RADIOGRAPHY, 2023, 29 (01) : 44 - 49
  • [22] The Importance of Temporal Resolution for Ultra-High-Resolution Coronary Angiography: Evidence From Photon-Counting Detector CT
    Mergen, Victor
    Sartoretti, Thomas
    Cundari, Giulia
    Serifovic, Medina
    Higashigaito, Kai
    Allmendinger, Thomas
    Schmidt, Bernhard
    Flohr, Thomas
    Manka, Robert
    Eberhard, Matthias
    Alkadhi, Hatem
    INVESTIGATIVE RADIOLOGY, 2023, 58 (11) : 767 - 774
  • [23] Ultra-high-resolution spectral silicon-based photon-counting detector CT for coronary CT angiography: Initial results in a dynamic phantom
    Holmes, Thomas Wesley
    Yin, Zhye
    Stevens, Grant M.
    Slavic, Scott
    Okerlund, Darin R.
    Maltz, Jonathan S.
    Pourmorteza, Amir
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2023, 17 (05) : 341 - 344
  • [24] Potential of ultra-high-resolution photon-counting CT of bone metastases: initial experiences in breast cancer patients
    Wehrse, E.
    Sawall, S.
    Klein, L.
    Glemser, P.
    Delorme, S.
    Schlemmer, H. -P.
    Kachelriess, M.
    Uhrig, M.
    Ziener, C. H.
    Rotkopf, L. T.
    NPJ BREAST CANCER, 2021, 7 (01)
  • [25] Computed tomography with a full FOV photon-counting detector in a clinical setting, the first experience
    Ferda, Jiri
    Vendis, Tomas
    Flohr, Thomas
    Schmidt, Bernhard
    Henning, Andre
    Ulzheimer, Stefan
    Pecen, Ladislav
    Ferdova, Eva
    Baxa, Jan
    Mirka, Hynek
    EUROPEAN JOURNAL OF RADIOLOGY, 2021, 137
  • [26] Estimation of Signal and Noise for a Whole-Body Photon Counting Research CT System
    Li, Zhoubo
    Leng, Shuai
    Yu, Zhicong
    Kappler, Steffen
    McCollough, Cynthia H.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [27] Effect of temporal resolution on calcium scoring: insights from photon-counting detector CT
    Sartoretti, Thomas
    Mergen, Victor
    Dzaferi, Amina
    Allmendinger, Thomas
    Manka, Robert
    Alkadhi, Hatem
    Eberhard, Matthias
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2025, 41 (03) : 615 - 625
  • [28] Photon-counting Detector CT: System Design and Clinical Applications of an Emerging Technology
    Leng, Shuai
    Bruesewitz, Michael
    Tao, Shengzhen
    Rajendran, Kishore
    Halaweish, Ahmed F.
    Campeau, Norbert G.
    Fletcher, Joel G.
    McCollough, Cynthia H.
    RADIOGRAPHICS, 2019, 39 (03) : 729 - 743
  • [29] Interweaving Network: A Novel Monochromatic Image Synthesis Method for a Photon-Counting Detector CT System
    Zheng, Ao
    Yang, Hongkai
    Zhang, Li
    Xing, Yuxiang
    IEEE ACCESS, 2020, 8 : 217701 - 217710
  • [30] Spectral photon-counting CT imaging of colorectal peritoneal metastases: initial experience in rats
    Thivolet, Arnaud
    Si-Mohamed, Salim
    Bonnot, Pierre-Emmanuel
    Blanchet, Christophe
    Kepenekian, Vahan
    Boussel, Loic
    Douek, Philippe
    Rousset, Pascal
    SCIENTIFIC REPORTS, 2020, 10 (01)