CT Hounsfield Numbers of Soft Tissues on Unenhanced Abdominal CT Scans: Variability Between Two Different Manufacturers' MDCT Scanners

被引:116
作者
Lamba, Ramit [1 ]
McGahan, John P. [1 ]
Corwin, Michael T. [1 ]
Li, Chin-Shang [1 ]
Tran, Tien [1 ]
Seibert, J. Anthony [1 ]
Boone, John M. [1 ]
机构
[1] Univ Calif Davis Hlth Syst, Dept Radiol, Sacramento, CA 95817 USA
基金
美国国家卫生研究院;
关键词
Hounsfield numbers; MDCT; MACROVESICULAR HEPATIC STEATOSIS; RENAL-CELL CARCINOMA; LIVING LIVER-DONORS; ADRENAL ADENOMAS; MASSES; MANAGEMENT; NEOPLASMS; DISEASE; LESIONS; VALUES;
D O I
10.2214/AJR.12.10037
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study is to determine whether Hounsfield numbers of soft tissues on unenhanced abdominal CT of the same patient vary on repeat scans done on two different manufacturers' MDCT scanners. MATERIALS AND METHODS. A database search was performed to identify patients older than 18 years who underwent unenhanced CT of the abdomen and pelvis performed both on a Volume CT (GE Healthcare) and a Definition AS Plus (Siemens Healthcare) 64-MDCT scanner within 12 months of each other. After excluding those patients for whom Hounsfield unit measurements would be affected by mitigating factors, 48 patients (mean age, 58.8 years) were identified. Hounsfield unit measurements were obtained in nine different soft-tissue anatomic locations on each scan, and the location of these sites was kept identical on each scan pair. Data were analyzed to evaluate Hounsfield unit differences between these scanners. RESULTS. In general, there was a low consistency in the Hounsfield unit measurements for each of these sites on scans obtained by the two scanners, with the subcutaneous fat in the left posterolateral flank showing the lowest correlation (intraclass correlation coefficient, 0.198). There were differences in the Hounsfield unit measurements obtained in all anatomic sites on scans obtained by both scanners. Mean Hounsfield unit measurements obtained on the Definition AS Plus scanner were lower than those obtained on the Volume CT scanner, with the intriguing exception of the anterior midline subcutaneous fat Hounsfield unit measurements, which were higher on the Definition AS Plus scanner. All differences were statistically significant (p < 0.05). CONCLUSION. Hounsfield unit measurements for unenhanced abdominal soft tissues of the same patient vary between scanners of two common MDCT manufacturers.
引用
收藏
页码:1013 / 1020
页数:8
相关论文
共 29 条
[1]   Multi-detector row CT attenuation measurements: Assessment of intra- and interscanner variability with an anthropomorphic body CT phantom [J].
Birnbaum, Bernard A. ;
Hindman, Nicole ;
Lee, Julie ;
Babb, James S. .
RADIOLOGY, 2007, 242 (01) :109-119
[2]   Incidental Adrenal Lesions: Principles, Techniques, and Algorithms for Imaging Characterization [J].
Boland, Giles W. L. ;
Blake, Michael A. ;
Hahn, Peter F. ;
Mayo-Smith, William W. .
RADIOLOGY, 2008, 249 (03) :756-775
[3]   Characterization of adrenal masses using unenhanced CT: An analysis of the CT literature [J].
Boland, GWL ;
Lee, MJ ;
Gazelle, GS ;
Halpern, EF ;
McNicholas, MMJ ;
Mueller, PR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 171 (01) :201-204
[4]  
BOSNIAK MA, 1986, RADIOLOGY, V158, P1
[5]   Hepatic Steatosis (Fatty Liver Disease) in Asymptomatic Adults Identified by Unenhanced Low-Dose CT [J].
Boyce, Cody J. ;
Pickhardt, Perry J. ;
Kim, David H. ;
Taylor, Andrew J. ;
Winter, Thomas C. ;
Bruce, Richard J. ;
Lindstrom, Mary J. ;
Hinshaw, J. Louis .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (03) :623-628
[6]  
Bushberg J.T., 2012, ESSENTIAL PHYS MED I, V3
[7]  
Clark JM, 2006, J CLIN GASTROENTEROL, V40, pS5
[8]   Effect of scattered radiation on image noise in cone beam CT [J].
Endo, M ;
Tsunoo, T ;
Nakamori, N ;
Yoshida, K .
MEDICAL PHYSICS, 2001, 28 (04) :469-474
[9]   Adrenal lesions: Attenuation measurement differences between CT scanners [J].
Hahn, Peter F. ;
Blake, Michael A. ;
Boland, Giles W. L. .
RADIOLOGY, 2006, 240 (02) :458-463
[10]   Enhancement characteristics of papillary renal neoplasms revealed on triphasic helical CT of the kidneys [J].
Herts, BR ;
Coll, DM ;
Novick, AC ;
Obuchowski, N ;
Linnell, G ;
Wirth, SL ;
Baker, ME .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 178 (02) :367-372