Radiostereometric analysis (RSA) is the most accurate technique to measure implant migration. However, it requires special equipment, technical expertise and analysis software and has not gained wide acceptance. The current paper analyzes a novel method to measure implant migration utilizing widely available computer tomography (CT). Three uncemented total hip replacements were performed in three human cadavers and six tantalum beads were inserted into the femoral bone similar to RSA. Six different 28 mm heads (-3, 0, 2.5, 5.0, 7.5 and 10 mm) were added to simulate five reproducible translations (maximum total point migration) of the center of the head. Implant migration was measured in a 3-D analysis software (Geomagic Studio 7). Repeat manual reconstructions of the center of the head were performed by two investigators to determine repeatability and accuracy. The accuracy of measurements between the centers of two head sizes was 0.11 mm with a CI 95 % of 0.22 mm. The intra-observer repeatability was 0.13 mm (CI 95 % 0.25 mm). The interrater-reliability was 0.943. CT based measurement of head displacement in a cadaver model were highly accurate and reproducible.
引用
收藏
页码:857 / 863
页数:7
相关论文
共 33 条
[1]   Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement [J].
Biedermann, R ;
Krismer, M ;
Stöckl, B ;
Mayrhofer, P ;
Ornstein, E ;
Franzén, H .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (02) :266-272
[2]   Three-Dimensional Bone-Implant Movements in Trochanteric Hip Fractures: Precision and Accuracy of Radiostereometric Analysis in a Phantom Model [J].
Bojan, Alicja J. ;
Bragdon, Charles ;
Jonsson, Anders ;
Ekholm, Carl ;
Karrholm, Johan .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2015, 33 (05) :705-711
[3]   Different cup migration in rheumatoid arthritis and arthrosis - A radiographic analysis of 127 uncemented acetabular cups [J].
Dominkus, M ;
Wanivenhaus, AH ;
Morscher, M ;
Powell, G ;
Krismer, M ;
Wolfl, G .
ACTA ORTHOPAEDICA SCANDINAVICA, 1998, 69 (05) :455-462
[4]   EARLY MIGRATION AND LATE ASEPTIC FAILURE OF PROXIMAL FEMORAL PROSTHESES [J].
FREEMAN, MAR ;
PLANTEBORDENEUVE, P .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (03) :432-438
[5]   A New Technique for Measuring Wear in Total Hip Arthroplasty Using Computed Tomography [J].
Goldvasser, Dov ;
Noz, Marilyn E. ;
Maguire, G. Q., Jr. ;
Olivecrona, Henrik ;
Bragdon, Charles R. ;
Makhau, Henrik .
JOURNAL OF ARTHROPLASTY, 2012, 27 (09) :1636-1640
[6]   THE MEANING AND USE OF THE AREA UNDER A RECEIVER OPERATING CHARACTERISTIC (ROC) CURVE [J].
HANLEY, JA ;
MCNEIL, BJ .
RADIOLOGY, 1982, 143 (01) :29-36
[7]   The premature failure of the Charnley Elite-Plus stem - A confirmation of RSA predictions [J].
Hauptfleisch, J ;
Glyn-Jones, S ;
Beard, DJ ;
Gill, HS ;
Murray, DW .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (02) :179-183
[8]   Migration of the uncemented Harris-Galante acetabular cup - Results of the einbildroentgenanalyse (EBRA) method [J].
Hendrich, C ;
Bahlmann, J ;
Eulert, J .
JOURNAL OF ARTHROPLASTY, 1997, 12 (08) :889-895
[9]  
Iannotti J P, 1986, J Arthroplasty, V1, P99, DOI 10.1016/S0883-5403(86)80047-X
[10]   Migration and wear of long-term successful Charnley total hip replacements [J].
Ilchmann, T ;
Markovic, L ;
Joshi, A ;
Hardinge, K ;
Murphy, J ;
Wingstrand, H .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (03) :377-381