Migration resistance of esophageal stents: The role of stent design

被引:26
作者
Mozafari, Hozhabr [1 ]
Dong, Pengfei [1 ]
Zhao, Shijia [1 ]
Bi, Yonghua [2 ]
Han, Xinwei [2 ]
Gu, Linxia [1 ,2 ,3 ]
机构
[1] Univ Nebraska, Dept Mech & Mat Engn, Lincoln, NE 68588 USA
[2] Zhengzhou Univ, Affiliated Hosp 1, Dept Intervent Radiol, Zhengzhou, Henan, Peoples R China
[3] Nebraska Ctr Mat & Nanosci, Lincoln, NE 68588 USA
基金
美国国家科学基金会;
关键词
Migration; Esophageal stent; Flared; Finite element method; Stent design; BIOMECHANICAL ANALYSIS; MECHANICAL-PROPERTIES; MODEL; SIMULATION; STRICTURES; TRANSPORT; SURVIVAL; SYSTEM;
D O I
10.1016/j.compbiomed.2018.06.031
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Objective: Stenting is one of the major treatments for malignant esophageal cancer. However, stent migration compromises clinical outcomes. A flared end design of the stent diminishes its migration. The goal of this work is to quantitatively characterize stent migration to develop new strategies for better clinical outcomes. Methods: An esophageal stent with flared ends and a straight counterpart were virtually deployed in an eso-phagus with asymmetric stricture using the finite element method. The resulted esophagus shape, wall stress, and migration resistance force of the stent were quantified and compared. Results: The lumen gain for both the flared stent and the straight one exhibited no significant difference. The flared stent induced a significantly larger contact force and thus a larger stress onto the esophagus wall. In addition, more migration resistance force was required to pull the flared stent through the esophagus. This force was inversely related to the occurrence rate of stent migration. A doubled strut diameter also increased the migration resistance force by approximately 56%. An increased friction coefficient from 0.1 to 0.3 also boosted the migration resistance force by approximately 39%. Summary: The mechanical advantage of the flared stent was unveiled by the significantly increased contact force, which provided the anchoring effect to resist stent migration. Both the strut diameter and friction coefficient positively correlated with the migration resistance force, and thus the occurrence of stent migration.
引用
收藏
页码:43 / 49
页数:7
相关论文
共 45 条
[1]   Realistic finite element-based stent design: The impact of balloon folding [J].
De Beule, Matthieu ;
Mortier, Peter ;
Carlier, Stphane G. ;
Verhegghe, Benedict ;
Van Impe, Rudy ;
Verdonck, Pascal .
JOURNAL OF BIOMECHANICS, 2008, 41 (02) :383-389
[2]   A two-layered mechanical model of the rat esophagus. Experiment and theory [J].
Fan, Yanhua ;
Gregersen, Hans ;
Kassab, Ghassan S. .
BIOMEDICAL ENGINEERING ONLINE, 2004, 3 (1)
[3]  
Franco KL, 2005, ADV THERAPY THORACIC
[4]   Esophageal stent migration: Testing few hypothesis with a simplified mathematical model [J].
Garbey, Marc ;
Salmon, Remi ;
Fikfak, Vid ;
Clerc, Claude O. .
COMPUTERS IN BIOLOGY AND MEDICINE, 2016, 79 :259-265
[5]   Influence of geometrical parameters on radial force during self-expanding stent deployment. Application for a variable radial stiffness stent [J].
Garcia, A. ;
Pena, E. ;
Martinez, M. A. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2012, 10 :166-175
[6]   The Relation Between the Arterial Stress and Restenosis Rate After Coronary Stenting [J].
Gu, Linxia ;
Zhao, Shijia ;
Muttyam, Aswini K. ;
Hammel, James M. .
JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2010, 4 (03)
[7]  
Hindy Pierre, 2012, Gastroenterol Hepatol (N Y), V8, P526
[8]   Delayed complications after placement of self-expanding stents in malignant esophageal obstruction: Treatment strategies and survival rate [J].
Homann, Nils ;
Noftz, Maria R. ;
Klingenberg-Noftz, Rolf D. ;
Ludwig, Diether .
DIGESTIVE DISEASES AND SCIENCES, 2008, 53 (02) :334-340
[9]   A STUDY OF THE GEOMETRICAL AND MECHANICAL-PROPERTIES OF A SELF-EXPANDING METALLIC STENT THEORY AND EXPERIMENT [J].
JEDWAB, MR ;
CLERC, CO .
JOURNAL OF APPLIED BIOMATERIALS, 1993, 4 (01) :77-85
[10]   A PTX/nitinol stent combination with temperature-responsive phase-change 1-hexadecanol for magnetocaloric drug delivery: Magnetocaloric drug release and esophagus tissue penetration [J].
Jin, Zhu ;
Wu, Keqin ;
Hou, Jingwen ;
Yu, Kunhong ;
Shen, Yuanyuan ;
Guo, Shengrong .
BIOMATERIALS, 2018, 153 :49-58