First comparison of a new dynamic navigation system and surgical guides for implantology: an in vitro study

被引:0
作者
Dure, Michael [1 ]
Berlinghoff, Frank [1 ]
Kollmuss, Maximilian [1 ]
Hickel, Reinhard [1 ]
Huth, Karin Christine [1 ]
机构
[1] Ludwig Maximilians Univ Munchen, Dept Conservat Dent & Periodontol, Univ Hosp, Goethestr 70, D-80336 Munich, Germany
关键词
dynamic navigation; dental implantology; surgical guide; implant planning; ACCURACY; PLACEMENT; SURGERY;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim: The transfer of 3D implant position planning to the clinical site is challenging. The aim of this study was to compare in vitro this transfer accuracy by using dynamic real-time navigation or static surgical guides. Materials and methods: Deviations between planned and actual pilot drill positions were calculated (Denacam and NobelGuide; each n = 90), matching pre- and postoperative CBCT images: entry point, angle, tip (each 3D); depth, mesiodistal/bucco-oral entry points, and angles (each 1D). The influence of the maxilla and mandible, implant region (anterior/posterior), and marker position (ipsilateral/contralateral, Denacam only) was investigated (Mann-Whitney U test). Results: No significant differences occurred regarding entry point (3D), mesiodistal/bucco-oral entry points (1D) or mesiodistal angle (1D) (P > 0.05). The angular and tip deviations (3D) were significantly smaller using Denacam (2.16 +/- 0.59 degrees, 0.80 +/- 0.55 mm; NobelGuide 2.54 +/- 1.19 degrees, 1.09 +/- 0.56 mm; P = 0.024, P < 0.0001). The deviations in depth and bucco- oral angle (1D) were significantly smaller using NobelGuide (1.05 +/- 0.50 mm, 1.02 +/- 1.16 degrees; Denacam 1.50 +/- 0.64 mm, 1.51 +/- 0.82 degrees; P < 0.0001). Significantly smaller deviations occurred within the mandible (Denacam, 5/8 parameters). Region and marker position showed no influence. No distinct influences were found with the use of NobelGuide. Conclusion: Denacam might possibly be a promising alternative to static surgical guides.
引用
收藏
页码:9 / 17
页数:9
相关论文
共 50 条
[41]   Influence of experience on dental implant placement: an in vitro comparison of freehand, static guided and dynamic navigation approaches [J].
Xiaotong Wang ;
Eman Shaheen ;
Sohaib Shujaat ;
Jan Meeus ;
Paul Legrand ;
Pierre Lahoud ;
Maurício do Nascimento Gerhardt ;
Constantinus Politis ;
Reinhilde Jacobs .
International Journal of Implant Dentistry, 8
[42]   Effect of the Configurations of Fiducial Markers on the Accuracy of Surgical Navigation in Zygomatic Implant Placement: An In Vitro Study [J].
Fan, Shengchi ;
Hung, Kuofeng ;
Bornstein, Michael M. ;
Huang, Wei ;
Wang, Feng ;
Wu, Yiqun .
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2019, 34 (01) :85-90
[43]   The Workflow of a New Dynamic Navigation System for the Insertion of Dental Implants in the Rehabilitation of Edentulous Jaws: Report of Two Cases [J].
Lopes, Armando ;
de Araujo Nobre, Miguel ;
Santos, Diogo .
JOURNAL OF CLINICAL MEDICINE, 2020, 9 (02)
[44]   Verification of the usability of a navigation method in dental implant surgery: In vitro comparison with the stereolithographic surgical guide template method [J].
Kang, Sang-Hoon ;
Lee, Jae-Won ;
Lim, Se-Ho ;
Kim, Yeon-Ho ;
Kim, Moon-Key .
JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 2014, 42 (07) :1530-1535
[45]   Effect of Storage Conditions and Time on the Dimensional Stability of 3D Printed Surgical Guides: An In Vitro Study [J].
Ntovas, Panagiotis ;
Marchand, Laurent ;
Basir, Barmak ;
Kudara, Yukio ;
Revilla-Leon, Marta ;
Att, Wael .
CLINICAL ORAL IMPLANTS RESEARCH, 2025, 36 (01) :92-99
[46]   Accuracy of a Computer-Aided Dynamic Navigation System in the Placement of Zygomatic Dental Implants: An In Vitro Study [J].
Gonzalez Rueda, Juan Ramon ;
Garcia Avila, Irene ;
de Paz Hermoso, Victor Manuel ;
Riad Deglow, Elena ;
Zubizarreta-Macho, Alvaro ;
Pato Mourelo, Jesus ;
Montero Martin, Javier ;
Hernandez Montero, Sofia .
JOURNAL OF CLINICAL MEDICINE, 2022, 11 (05)
[47]   Navigation of Pedicle Screws in the Thoracic Spine with a New Electromagnetic Navigation System: A Human Cadaver Study [J].
Hahn, Patrick ;
Oezdemir, Semih ;
Komp, Martin ;
Giannakopoulos, Athanasios ;
Kasch, Richard ;
Merk, Harry ;
Liermann, Dieter ;
Godolias, Georgios ;
Ruetten, Sebastian .
BIOMED RESEARCH INTERNATIONAL, 2015, 2015
[48]   Accuracy of guided implant surgery obtained using 3D-printed surgical guides - An in vitro comparison of four evaluation methods [J].
Herstell, Hannah ;
Berndt, Saskia ;
Kuehne, Christina ;
Reich, Sven .
INTERNATIONAL JOURNAL OF COMPUTERIZED DENTISTRY, 2022, 25 (02) :161-172
[49]   Comparing accuracy in guided endodontics: dynamic real-time navigation, static guides, and manual approaches for access cavity preparation – an in vitro study using 3D printed teeth [J].
Karin Christine Huth ;
Lukas Borkowski ;
Anja Liebermann ;
Frank Berlinghoff ;
Reinhard Hickel ;
Falk Schwendicke ;
Marcel Reymus .
Clinical Oral Investigations, 28
[50]   Comparison of Intraosseous Temperature Rise in Different Surgical Guide Systems: An In Vitro Study [J].
Uckun, Gozde Gokce ;
Saygili, Sina ;
Sulun, Tonguc ;
Cankaya, Abdulkadir Burak .
JOURNAL OF ADVANCED ORAL RESEARCH, 2024, 15 (02) :219-227