Comparison of Axial Length, Corneal Curvature, and Anterior Chamber Depth Measurements of 2 Recently Introduced Devices to a Known Biometer

被引:94
作者
Shajari, Mehdi [1 ]
Cremonese, Carla [1 ]
Petermann, Kerstin [1 ]
Singh, Pankaj [1 ]
Mueller, Michael [1 ]
Kohnen, Thomas [1 ]
机构
[1] Goethe Univ, Dept Ophthalmol, Theodor Stern Kai 7, D-60590 Frankfurt, Germany
关键词
OPTICAL COHERENCE TOMOGRAPHY; LENS POWER CALCULATION; CATARACT-SURGERY; IOL MASTER; REPEATABILITY; ASTIGMATISM; AGREEMENT; INTERFEROMETRY; REPRODUCIBILITY; REFLECTOMETRY;
D O I
10.1016/j.ajo.2017.02.027
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: Evaluate agreement and repeatability of axial length (AL), corneal curvature, and anterior chamber depth measurements of 2 new devices, 1 using Scheimpflug technology combined with partial coherence interferometry (Pentacam AXL) and 1 using swept-source optical coherence tomography (IOL Master 700) to a widely used optical biometer (IOL Master 500). DESIGN: Prospective, instrument reliability analysis. METHODS: This study was conducted in the Department of Ophthalmology at Goethe-University, Frankfurt, Germany. Seventy-nine eyes in 79 patients were measured with the IOL Master 500, the IOL Master 700, and with the Pentacam AXL. All eyes were measured twice with each device. The results were statistically compared using Bland-Altman plots. Outcome measures included AL, corneal curvature, and the measurement of anterior chamber depth. RESULTS: The Bland-Altman analysis of axial length measurements with the IOL Master 500 and IOL Master 700 showed a mean difference of 0.008 mm between the 2 devices, with a coefficient of repeatability (COR) of 0.044 (P = .05). For measurements with the IOL Master 500 and Pentacam AXL, a mean difference of 0.026 mm and COR of 0.055 (P = .75) was found. For measurements with Pentacam AXL and IOL Master 700, a mean difference of -0.019 mm and COR of 0.052 (P = .06) was found. In terms of repeatability, there was no significant difference in measurement of any variable with any device. Vector analysis showed no significant difference for corneal astigmatism measurements (P = .84). CONCLUSION: There was no statistically significant difference in compared biometric parameters performed with all 3 devices. (C) 2017 Elsevier Inc. All rights reserved.
引用
收藏
页码:58 / 64
页数:7
相关论文
共 33 条
[1]   Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500 [J].
Akman, Ahmet ;
Asena, Leyla ;
Gungor, Sirel Gur .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2016, 100 (09) :1201-1205
[2]  
[Anonymous], LANCET
[3]   Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry [J].
Aristodemou, Petros ;
Cartwright, Nathaniel E. Knox ;
Sparrow, John M. ;
Johnston, Robert L. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (01) :63-71
[4]   Foveal pit morphology evaluation during optical biometry measurements using a full-eye-length swept-source OCT scan biometer prototype [J].
Bertelmann, Thomas ;
Blum, Marcus ;
Kunert, Kathleen ;
Haigis, Wolfgang ;
Messerschmidt-Roth, Anke ;
Peter, Monika ;
Buehren, Tobias ;
Sekundo, Walter .
EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2015, 25 (06) :552-558
[5]   Measurement error [J].
Bland, JM ;
Altman, DG .
BRITISH MEDICAL JOURNAL, 1996, 312 (7047) :1654-1654
[6]   Reproducibility of a Long-Range Swept-Source Optical Coherence Tomography Ocular Biometry System and Comparison with Clinical Biometers [J].
Grulkowski, Ireneusz ;
Liu, Jonathan J. ;
Zhang, Jason Y. ;
Potsaid, Benjamin ;
Jayaraman, Vijaysekhar ;
Cable, Alex E. ;
Duker, Jay S. ;
Fujimoto, James G. .
OPHTHALMOLOGY, 2013, 120 (11) :2184-2190
[7]   Macular disease detection with a swept-source optical coherence tomography-based biometry device in patients scheduled for cataract surgery [J].
Hirnschall, Nino ;
Leisser, Christoph ;
Radda, Stephan ;
Maedel, Sophie ;
Findl, Oliver .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (04) :530-536
[8]   Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry [J].
Hoffer, Kenneth J. ;
Hoffmann, Peter C. ;
Savini, Giacomo .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (08) :1165-1172
[9]   Clinical results using the Holladay 2 intraocular lens power formula [J].
Hoffer, KJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (08) :1233-1237
[10]   Evaluating and reporting astigmatism for individual and aggregate data [J].
Holladay, JT ;
Dudeja, DR ;
Koch, DD .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (01) :57-65