共 52 条
On-line separation and preconcentration of hexavalent chromium on a novel mesoporous silica adsorbent with determination by solution-cathode glow discharge-atomic emission spectrometry
被引:1
作者:
Ma, Jiaxian
[1
,2
]
Wang, Zheng
[1
]
Li, Qing
[1
]
Gai, Rongyin
[1
]
Li, Xiaohong
[1
]
机构:
[1] Chinese Acad Sci, Shanghai Inst Ceram, Shanghai 200050, Peoples R China
[2] Shanghai Univ, Sch Mat Sci & Engn, Shanghai 200072, Peoples R China
基金:
中国国家自然科学基金;
关键词:
SOLID-PHASE EXTRACTION;
ATMOSPHERIC-PRESSURE;
ELEMENTAL ANALYSIS;
ABSORPTION-SPECTROMETRY;
TRACE-METALS;
CR(VI);
SPECIATION;
ADSORPTION;
SAMPLES;
WATER;
D O I:
10.1039/c4ja00273c
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
A flow injection (FI)-technique based on an on-line solid phase extraction (SPE) separation system coupled with solution-cathode glow discharge-atomic emission spectrometry (SCGD-AES) detection was developed for the determination of hexavalent chromium in aqueous samples. In this system, Cr(VI) was separated and preconcentrated on a novel lysine-modified mesoporous silica (Fmoc-SBA-15) at pH 5, and detected after elution. An enrichment factor of 91 was achieved under optimized experimental conditions, at a FI flow rate of 1.5 mL min(-1) and 0.1 mol L-1 NH3 center dot H2O for Cr(VI) elution. A FI-SCGD-AES detection limit of 0.75 mu g L-1 could be achieved, with a linear range of 10-10 000 mu g L-1. The precision of nine replicate Cr(VI) measurements was 4.2% at 100 mu g L-1. The accuracy was validated using a certified reference material of riverine water (GBW08607).
引用
收藏
页码:2315 / 2322
页数:8
相关论文