Calibration of a proton beam energy monitor

被引:26
|
作者
Moyers, M. F. [1 ]
Coutrakon, G. B. [1 ]
Ghebremedhin, A. [1 ]
Shahnazi, K. [1 ]
Koss, P. [1 ]
Sanders, E. [1 ]
机构
[1] Loma Linda Univ, Med Ctr, Dept Radiat Med, Loma Linda, CA 92354 USA
关键词
proton; energy; monitor; range; Monte Carlo;
D O I
10.1118/1.2717382
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Delivery of therapeutic proton beams requires an absolute energy accuracy of +/- 0.64 to 0.27 MeV for patch fields and a relative energy accuracy of +/- 0.10 to 0.25 MeV for tailoring the depth dose distribution using the energy stacking technique. Achromatic switchyard tunes, which lead to better stability of the beam incident onto the patient, unfortunately limit the ability of switchyard magnet tesla meters to verify the correct beam energy within the tolerances listed above. A new monitor to measure the proton energy before each pulse is transported through the switchyard has been installed into a proton synchrotron. The purpose of this monitor is to correct and/or inhibit beam delivery when the measured beam energy is outside of the tolerances for treatment. The monitor calculates the beam energy using data from two frequency and eight beam position monitors that measure the revolution frequency of the proton bunches and the effective offset of the orbit from the nominal radius of the synchrotron. The new energy monitor has been calibrated by measuring the range of the beam through water and comparing with published range-energy tables for various energies. A relationship between depth dose curves and range-energy tables was first determined using Monte Carlo simulations of particle transport and energy deposition. To reduce the uncertainties associated with typical scanning water phantoms, a new technique was devised in which the beam energy was scanned while fixed thickness water tanks were sandwiched between two fixed parallel plate ionization chambers. Using a multitude of tank sizes, several energies were tested to determine the nominal accelerator orbit radius. After calibration, the energy reported by the control system matched the energy derived by range measurements to better than 0.72 MeV for all nine energies tested between 40 and 255 MeV with an average difference of -0.33 MeV. A study of different combinations of revolution frequency and radial offsets to test the envelope of algorithm accuracy demonstrated a relative accuracy of +/- 0.11 MeV for small energy changes between 126 and 250 MeV. These new measurements may serve as a data set for benchmarking range-energy relationships. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1952 / 1966
页数:15
相关论文
共 50 条
  • [1] Calibration of a proton beam energy monitor
    Moyers, MF
    Ghebremedhin, A
    Shahnazi, K
    Arjomandy, B
    Coutrakon, GB
    MEDICAL PHYSICS, 2004, 31 (06) : 1867 - 1867
  • [2] Calibration of the Proton Beam Cherenkov Monitor
    S. V. Akulinichev
    Yu. K. Gavrilov
    R. M. Djilkibaev
    Instruments and Experimental Techniques, 2023, 66 : 365 - 368
  • [3] Proton beam monitor chamber calibration
    Goma, C.
    Lorentini, S.
    Meer, D.
    Safai, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (17): : 4961 - 4971
  • [4] Calibration of the Proton Beam Cherenkov Monitor
    Akulinichev, S. V.
    Gavrilov, Yu. K.
    Djilkibaev, R. M.
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2023, 66 (03) : 365 - 368
  • [5] Comment on 'Proton beam monitor chamber calibration'
    Palmans, Hugo
    Vatnitsky, Stanislav M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (17): : 6585 - 6593
  • [6] Proton beam monitor chamber calibration in clinical practice
    Goma, C.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S54 - S54
  • [7] Reply to comment on 'Proton beam monitor chamber calibration'
    Goma, Carles
    Lorentini, Stefano
    Meer, David
    Safai, Sairos
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (17): : 6594 - 6601
  • [8] Comprehensive Evaluation and Implementation of Two Independent Methods for Beam Monitor Calibration for Proton Scanning Beam
    Shen, J.
    Ding, X.
    Hu, Y.
    Kang, Y.
    Liu, W.
    Deng, W.
    Bues, M.
    MEDICAL PHYSICS, 2020, 47 (06) : E731 - E731
  • [9] Technical Note: Comprehensive evaluation and implementation of two independent methods for beam monitor calibration for proton scanning beam
    Shen, Jiajian
    Ding, Xiaoning
    Hu, Yanle
    Kang, Yixiu
    Liu, Wei
    Deng, Wei
    Bues, Martin
    MEDICAL PHYSICS, 2019, 46 (12) : 5867 - 5875
  • [10] RHIC beam position monitor calibration
    Gullotta, J.
    Dawson, W. C.
    Degen, C.
    Michnoff, R.
    Naase, S.
    Russo, T.
    Satogata, T.
    BEAM INSTRUMENTATION WORKSHOP 2006, 2006, 868 : 337 - +